Clinical Investigation of Visual Function After Bilateral Implantation of Two Presbyopia-Correcting Trifocal Intraocular Lenses (IOL)
Clinical Investigation of Visual Function After Bilateral Implantation of Two Presbyopia-Correcting Trifocal IOLs
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Diagnosis of bilateral cataracts with planned clear cornea cataract removal;
- Able to comprehend and willing to sign informed consent and complete all required postoperative follow-up procedures;
- Calculated lens power between 13.0 and 30.0 Diopters (D);
- Preoperative Best-corrected distance visual acuity (BCDVA) worse than 0.20 logMAR (ie, 0.22 logMAR or worse) in at least one eye;
- Potential postoperative BCDVA of 0.20 logMAR or better in both eyes. Note: Subjects with any pathology that could reduce visual potential should not be enrolled in this trial;
- Preoperative regular corneal astigmatism of < 1.00 D, in both eyes;
- Clear intraocular media other than cataract in both eyes;
- Other protocol-specified inclusion criteria may apply.
Exclusion Criteria:
- Reasonably expected to require an ocular surgical treatment at any time during the study (other than YAG capsulotomy);
- Previous refractive surgery or planned refractive surgery procedures throughout the entire duration of participation in the clinical study (including, but not limited to LASIK, astigmatic keratotomy and limbal relaxing incisions);
- Clinically significant corneal abnormalities including corneal dystrophy, inflammation or edema;
- Amblyopia;
- Previous corneal transplant;
- Any recurrent severe anterior or posterior segment inflammation of any etiology, and/or history of any disease producing an intraocular inflammatory reaction;
- Rubella, congenital, traumatic, or complicated cataracts;
- Glaucoma (uncontrolled or controlled with medication);
- Degenerative eye disorders;
- History of or current retinal conditions or predisposition to retinal conditions, previous history of, or a predisposition to, retinal detachment or presence of diabetic retinopathy;
- Optic nerve atrophy;
- Expected to require retinal laser treatment;
- Color vision deficiencies;
- Pregnant or lactating (current or planned during the course of the study);
- Other protocol-specified exclusion criteria may apply.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Randomized
- Interventional Model: Parallel Assignment
- Masking: Double
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: TFNT00
AcrySof® IQ PanOptix™ Presbyopia-Correcting IOL, bilateral implantation
|
Trifocal IOL(near, intermediate, distance) implanted for long-term use over the lifetime of the cataract patient
Other Names:
|
|
Active Comparator: 839MP
AT LISA® tri IOL, bilateral implantation
|
Trifocal IOL(near, intermediate, distance) implanted for long-term use over the lifetime of the cataract patient
Other Names:
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Least Squares Mean Binocular Uncorrected Intermediate Visual Acuity (UCIVA) (60cm) at Day 120-180
Time Frame: Day 120-180 from second eye implantation
|
VA was tested binocularly (both eyes together) with no refractive correction in place using an early treatment diabetic retinopathy study (ETDRS) chart set at 60 cm.
VA was measured in "logarithm of the minimum angle of resolution" (logMAR), with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart.
A lower numeric value represents better visual acuity.
|
Day 120-180 from second eye implantation
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Least Squares Mean Binocular UCIVA (60cm) at Day 120-180
Time Frame: Day 120-180 from second eye implantation
|
VA was tested binocularly with no refractive correction in place using an ETDRS chart set at 60 cm.
VA was measured in logMAR, with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart.
A lower numeric value represents better visual acuity.
|
Day 120-180 from second eye implantation
|
|
Least Squares Mean Binocular Uncorrected Distance VA (4m) at Day 120-180
Time Frame: Day 120-180 from second eye implantation
|
VA was tested binocularly with no refractive correction in place using an ETDRS chart set at 4 meters.
VA was measured in logMAR, with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart.
A lower numeric value represents better visual acuity.
|
Day 120-180 from second eye implantation
|
|
Least Squares Mean Binocular Uncorrected Near VA (40cm) at Day 120-180
Time Frame: Day 120-180 from second eye implantation
|
VA was tested binocularly with no refractive correction in place using an ETDRS chart set at 40 cm.
VA was measured in logMAR, with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart.
A lower numeric value represents better visual acuity.
|
Day 120-180 from second eye implantation
|
|
Mean Photopic Binocular Defocus Curve at Day 120-180
Time Frame: Day 120-180 from second eye implantation
|
A defocus curve is created by multiple measurements of one's visual acuity (VA) at different spherical powers.
VA was measured in logMAR.
A lower logMAR value indicates better visual acuity.
No statistical test was performed.
|
Day 120-180 from second eye implantation
|
|
Mean Photopic Without Glare Binocular Distance Contrast Sensitivity at Day 120-180
Time Frame: Day 120-180 from second eye implantation
|
Contrast sensitivity (ie, the ability to detect slight changes in luminance before they become indistinguishable) was assessed binocularly in lighted conditions with distance manifest correction in place and uncorrected, with no glare source illumination.
Contrast sensitivity was assessed at spatial frequencies of 3, 6, 12, and 18 cycles per degree (CPD) and reported in log units.
A higher numeric value represents better contrast sensitivity.
No statistical test was performed.
|
Day 120-180 from second eye implantation
|
|
Mean Photopic With Glare Binocular Distance Contrast Sensitivity at Day 120-180
Time Frame: Day 120-180 from second eye implantation
|
Contrast sensitivity (ie, the ability to detect slight changes in luminance before they become indistinguishable) was assessed binocularly in lighted conditions with distance manifest correction in place and uncorrected, with glare source illumination.
Contrast sensitivity was assessed at spatial frequencies of 3, 6, 12, and 18 CPDs and reported in log units.
A higher numeric value represents better contrast sensitivity.
No statistical test was performed.
|
Day 120-180 from second eye implantation
|
|
Mean Mesopic Without Glare Binocular Distance Contrast Sensitivity at Day 120-180
Time Frame: Day 120-180 from second eye implantation
|
Contrast sensitivity (ie, the ability to detect slight changes in luminance before they become indistinguishable) was assessed binocularly in dim to dark conditions with distance manifest correction in place and uncorrected, with no glare source illumination.
Contrast sensitivity was assessed at spatial frequencies of 1.5, 3, 6, and 12 CPDs and reported in log units.
A higher numeric value represents better contrast sensitivity.
No statistical test was performed.
|
Day 120-180 from second eye implantation
|
|
Mean Mesopic With Glare Binocular Distance Contrast Sensitivity at Day 120-180
Time Frame: Day 120-180 from second eye implantation
|
Contrast sensitivity (ie, the ability to detect slight changes in luminance before they become indistinguishable) was assessed binocularly with distance manifest correction in place and uncorrected, with glare source illumination.
Contrast sensitivity was assessed at spatial frequencies of 1.5, 3, 6, and 12 CPDs and reported in log units.
A higher numeric value represents better contrast sensitivity.
No statistical test was performed.
|
Day 120-180 from second eye implantation
|
|
Subject Satisfaction Recorded at Day 120-180
Time Frame: Day 120-180 from second eye implantation
|
At the Month 6 Visit, subjects responded to the question: "Given your current postoperative vision, if you had to do it all over would you have the same lens implanted again?".
Responses were reported as a percentage of subjects.
No statistical test was performed.
|
Day 120-180 from second eye implantation
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Study Director: Sr. Clinical Manager, GCRA, Alcon, a Novartis Company
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Estimate)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- ILH297-P003
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.