Safety and Clinical Outcomes of Magnetic Resonance Imaging in Patients With Cardiac Implantable Electrical Devices (BIDMC-MRI)

July 19, 2022 updated by: Dan B. Kramer, Beth Israel Deaconess Medical Center
This protocol outlines a prospective study evaluating safety and clinical outcomes of magnetic resonance imaging (MRI) examinations performed on patients with cardiac implantable electrical devices (CIEDs) including pacemakers (PMs) and implantable cardioverter-defibrillators (ICDs). This prospective study aims to improve the evidence base regarding this common clinical scenario. Specifically, the investigators aim to address whether results of MRI in PM/ICD patients affect physician decision making related to clinical management strategy and planning treatment interventions. Further, this study will evaluate whether the results of MRI in these patients affects patient outcomes related to survival and adverse events during or after MRI scanning.

Study Overview

Status

Completed

Conditions

Detailed Description

MRI is the preferred imaging modality for both acute and elective evaluation of many conditions. For example, for many diseases of the central nervous system, no other imaging modality provides the requisite soft tissue resolution to support critical care elements such as neurosurgical planning or assessment of intracranial malignancies.

Historically, MRIs were considered contraindicated for patients with CIEDs including PMs and ICDs. However, studies have demonstrated very high utilization of MRIs among patients eligible for CIEDs. In many cases, alternative imaging is either insufficient or, as with CT-myelography, poses substantial risks or procedural hurdles that may be less favorable than the potential risks of an MRI.

In recent years, a growing literature has suggested that MRIs can be performed safely in the setting of CIEDs if certain precautions are taken. Additionally, guidance documents have been released by professional societies including the American Heart Association (AHA) and the American College of Radiology (ACR) outlining recommended safety measures that can be taken to minimize risk and maximize benefit for patients in need of imaging. However, there are fewer data on performing MRIs in pacemaker-dependent patients with ICDs, who were excluded from the largest published study as well as the MagnaSafe Registry. Thus, demonstrating the safety and clinical utility of performing MRIs in this context is of high importance.

In response to urgent clinical considerations, Beth Israel Deaconess Medical Center has established a clinical protocol to provide safe MRIs in cases of compelling clinical indications, with close supervision and oversight from Radiology and Cardiac Electrophysiology staff. Thus, the investigators now propose a prospective study to evaluate safety and clinical outcomes for patients treated under this clinical protocol.

Specifically, this study proposes to perform a prospective clinical trial in patients with CIEDs that have a clinical need for MRI as outpatients to determine:

  1. Safety according to the incidence rates of acute major and minor adverse events according to pre-specified definitions.
  2. Clinical outcomes and scan utility affecting patients' survival and quality of life and treatment, including acute and longitudinal changes in device parameters, the need for system revision, and proportion of MRI findings influencing clinical management including medical or surgical interventions, and diagnostic or prognostic changes.

Study Type

Observational

Enrollment (Actual)

1169

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Massachusetts
      • Boston, Massachusetts, United States, 02215
        • Beth Israel Deaconess Medical Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

This is a prospective observational study of patients with a cardiac implantable electronic device (CIED) referred for a clinicially indicated MRI.

Description

Inclusion Criteria:

  • Patients have a medically/clinically indicated need for an MRI
  • Patients with a non-MRI conditional pacemaker and/or ICD implanted after the year 2000
  • Patients are willing and able to sign consent and HIPAA authorization or an authorized representative of the patient is willing to sign consent for the patient

    *The following factors are considered relative contraindications to the clinical protocol at BIDMC, and will similarly be considered exclusion criteria for the study. Exceptions on a case-by-case basis for circumstances of dire, life-threatening need may be considered.

  • Leads implanted <6 weeks prior to the MRI
  • Presence of any capped/abandoned leads
  • Presence of nontransvenous epicardial leads

Exclusion Criteria:

  • Other contraindications to MRI (e.g. other non-MRI-conditional implants)
  • Pacemaker or ICD implanted prior to the year 2000
  • Leads and/or generator implanted within 6 weeks of the proposed MRI
  • CIED that is FDA approved as MRI conditional
  • Pregnant women

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: Cohort
  • Time Perspectives: Prospective

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of Patients with Any of the Following Major Adverse Events
Time Frame: Change from baseline at 1 week and 6 months
  1. Acute loss of pacing peri-MRI (defined in scanner, or after scan prior to discharge)
  2. Inappropriate shock or ATP therapy peri-MRI
  3. Any death peri-MRI adjudicated to be related to the scan
Change from baseline at 1 week and 6 months
Number of Patients with Any of the Following Minor Adverse Events
Time Frame: Change from baseline at 1 week and 6 months
  1. Clinically significant (symptoms or hemodynamic compromise) inappropriate pacing (e.g. undersensing or inappropriate asynchronous pacing)
  2. Any clinically significant arrhythmias peri-MRI
  3. Power-on-reset events
  4. Acute variation in pre/post MRI capture thresholds ≥ 50%
  5. Acute variation in pre/post MRI lead impedance ≥ 30%
  6. Acute variation in pre/post MRI P/R wave amplitude ≥ 50%
Change from baseline at 1 week and 6 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Principal Investigator: Daniel Kramer, MD, Beth Israel Deaconess Medical Center

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

December 1, 2016

Primary Completion (Actual)

March 5, 2021

Study Completion (Actual)

March 5, 2021

Study Registration Dates

First Submitted

January 6, 2017

First Submitted That Met QC Criteria

January 9, 2017

First Posted (Estimate)

January 10, 2017

Study Record Updates

Last Update Posted (Actual)

July 20, 2022

Last Update Submitted That Met QC Criteria

July 19, 2022

Last Verified

July 1, 2022

More Information

Terms related to this study

Additional Relevant MeSH Terms

Other Study ID Numbers

  • 2016P000204

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.