Mini Incision Dynamic Hip Screw Technique (MIDHST)
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
This technique is the same of AO Dynamic Hip Screw (DHS) technique using 4 holes plate or smaller, in this case it is enough to make just a 3 cm skin incision over the guide wire that is inserted parallel to the anteversion wire by using the 135° guide fixed on the T-handle; (step 2-2), and continuing AO technique steps, until step 3-4, for the plate assembling we remove the guide wire, and insert the plate sliding it through the skin cut over the femur shaft under the muscles using the plate for dissection the soft tissue above the bone surface non using any tool as periosteal elevator but the plate itself by holding it from its DHS screw canal and in 180 degree rotation position in axial aspect (transvers aspect) using the plate end for soft tissue gentle dissection over the bone, then turning the plate 180 degree over the bone shaft to the right position and sliding the DHS screw plat canal over the DHS screw using the measuring bar. Two skin stitches are enough.
Technique
Step 1-reduction
In many cases the traction table is used. The reduction is done on this table and before the patient is draped. Important also is to guarantee smooth access of the image intensifier in both planes; AP and lateral.
Step 2-guide wire insertion
- The anteversion of the femoral neck is determined with a long K-wire inserted with the blunt end first. An alternative is to use a long, non-threaded K-wire.
- Under X-ray control, the guide wire is inserted parallel to the anteversion wire by using the 135° guide fixed on the T-handle.
Step 3-screw insertion
- The length of the screw must be measured/determined with the guide wire. Note that the screw must be 10 mm shorter than the length of the guide wire. The surgeon will deduct 10 mm of the measured length to determine the screw length.
- Set the triple reamer to 10 mm shorter than the measured length. The hole is drilled over the guide wire.
- Tapping is only required in young patients with dense bone.
For the plate assembling we remove the guide wire, and insert the plate sliding it through the skin cut over the femur shaft under the muscles using the plate for dissection the soft tissue above the bone surface non using any tool as periosteal elevator but the plate itself by holding it from its DHS screw canal and in 180 degree rotation position in axial aspect (transvers aspect) using the plate end for soft tissue gentle dissection over the bone, then turning the plate 180 degree over the bone shaft to the right position.
- The plate is slid over the wrench.
- The DHS screw is attached to the coupling wrench.
- The sleeve is assembled over the wrench.
Step 4-plate fixation
Impact the plate in order to have best contact with the femur. The plate is fixed with conventional 4.5 mm screws:
- Drill bit 3.2 mm with sleeve
- Measuring
- Tap (when no self-tapping screws are used)
- Insertion of screw
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Aleppo, Syrian Arab Republic
- Dr. Sayed Issa's Clinic
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Patients presented with intertrochanteric hip fractures
Exclusion Criteria:
- not operative patients
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Abdulhai
One show man technique, it is the same of AO Dynamic Hip Screw (DHS) technique using 4 holes plate or smaller, in this case it is enough to make just a 3 cm skin incision.
|
This technique is the same of AO Dynamic Hip Screw (DHS) technique using 4 holes plate or smaller, in this case it is enough to make just a 3 cm skin incision over the guide wire that is inserted parallel to the anteversion wire by using the 135° guide fixed on the T-handle; (step 2-2), and continuing AO technique steps, until step 3-4, for the plate assembling we remove the guide wire, and insert the plate sliding it through the skin cut over the femur shaft under the muscles using the plate for dissection the soft tissue above the bone surface non using any tool as periosteal elevator but the plate itself by holding it from its DHS screw canal and in 180 degree rotation position in axial aspect (transvers aspect) using the plate end for soft tissue gentle dissection over the bone, then turning the plate 180 degree over the bone shaft to the right position and sliding the DHS screw plat canal over the DHS screw using the measuring bar.
Two skin stitches are enough.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Proportion of patients with post surgical trauma
Time Frame: 12 -14 days for soft tissue healing
|
mini incision
|
12 -14 days for soft tissue healing
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Time Frame |
|---|---|
|
Proportion of patients with cosmetic scar
Time Frame: Up to 6 weeks
|
Up to 6 weeks
|
Collaborators and Investigators
Sponsor
Sponsor
Study record dates
Study Major Dates
Study Start
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Other Study ID Numbers
Other Study ID Numbers
- Abdulhai's Technique
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Intratrochontric Fracture
-
NCT06658379RecruitingFemoral Neck Fractures | Femoral Shaft Fracture | Tibial Shaft Fracture | Forearm Fracture | Fracture Non Union | Distal Tibia Fracture | Fracture Healing | Femur Distal Fracture | Proximal Tibia Fracture | Humerus Shaft Fracture
-
NCT01583556CompletedDistal Radius Fracture | Metacarpal Fracture | Mallet Fracture | Radial Head Fracture | Metacarpal Neck Fracture
-
NCT02609347CompletedPilon Fracture | Calcaneus Fracture | Ankle Fracture | Talus Fracture
-
NCT02032966UnknownAnkle Fracture | Trimalleolar Fracture | Medial Malleolus Fracture | Bimalleolar Fracture | Malleolus Fracture
-
NCT06024798RecruitingStress Fracture Metatarsal | Lower Limb Fracture | Stress Fracture Foot | Stress Fracture Ankle | Stress Fracture of Tibia
-
NCT03952182TerminatedCompression Fracture of Thoracic Vertebral Body | Thoracic Fracture | Lumbar Fracture | Compression Fracture of Lumbar Spine | Burst Fracture of Thoracic Vertebra | Burst Fracture of Lumbar Vertebra
-
NCT02826538TerminatedFracture of Clavicle | Fracture of Femur | Fracture of Humerus | Fracture of Ulna Radius | Fracture of Hand | Fracture of Pelvis | Fracture of Tibia Fibula | Fracture of Skull
-
NCT07221019RecruitingFracture | Fracture Dislocation of Ankle Joint | Fracture Leg | Fracture Femur | Fracture Lower Leg
-
NCT07552766Not yet recruitingFracture Forearm | Fracture Fixation, Internal | Fracture Elbow | Fracture Arm | Fracture Distal Radius | Fracture Fixation | Fracture Closed of Lower End of Forearm, Unspecified
-
NCT05445960RecruitingAnkle Fractures | Ankle Fracture - Lateral Malleolus | Syndesmotic Injuries | Ankle Fracture, Trimalleolar | Ankle Fracture, Bimalleolar | Fibula Fracture | Ankle Fracture - Medial Malleolus | Maisonneuve's Fracture
Clinical Trials on Mini Incision Dynamic Hip Screw Technique
-
NCT01427036CompletedOther Reconstructive Surgery
-
NCT05781815Recruiting
-
NCT05677061CompletedIntertrochanteric Fractures
-
NCT03849014Completed
-
NCT06360835Completed
-
NCT02788994UnknownUnilateral Unstable Hip
-
NCT03666637CompletedAvascular Necrosis | Femoral Neck Fractures
-
NCT00555945CompletedFemoral Neck Fractures | Intertrochanteric Fracture
-
NCT00597779Completed
-
NCT04959370Completed