Post-Market Clinical Investigation of the Clareon® IOL

August 17, 2023 updated by: Alcon Research
The primary purpose of this study is to evaluate the long-term (3 years) visual acuity and adverse event outcomes for the Clareon® Intraocular Lens (IOL). A comparison to historical safety and performance endpoint (SPE) rates as reported in the European Standard International Organization for Standardization (EN ISO) 11979-7:2014 will be conducted at one year. The secondary purpose of this study is to evaluate the visual acuity outcomes with the Clareon IOL at Years 2 and 3.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Detailed Description

Subjects will attend a total of 12 study visits over a period of approximately 36 months: One preoperative screening visit, two operative visits, and nine post-implantation visits. Both eyes will be implanted, with the second eye implantation to occur 2-21 days after the first. Primary endpoint data will be collected at the Year 1 visit (Day 330-420 post-implantation from second eye surgery).

Study Type

Interventional

Enrollment (Actual)

245

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Sydney, Australia, 2067
        • Alcon Investigative Site
    • Victoria
      • Footscray, Victoria, Australia, 3011
        • Alcon Investigative Site
      • Mornington, Victoria, Australia, 3931
        • Alcon Investigative Site
      • Lyon, France, 69003
        • Alcon Investigative Site
      • Paris, France, 75014
        • Alcon Investigative Site
    • Cedex
      • Brest, Cedex, France, 29609
        • Alcon Investigative Site
      • Frankfurt, Germany, 60590
        • Alcon Investigative Site
      • Heidelberg, Germany, 69120
        • Alcon Investigative Site
      • Perugia, Italy, 06129
        • Alcon Investigative Site
      • Pisa, Italy, 56124
        • Alcon Investigative Site
      • Amsterdam, Netherlands, 1105AZ
        • Alcon Investigative Site
      • Alcalá De Henares, Spain, 28805
        • Alcon Investigative Site
      • Oviedo, Spain, 33012
        • Alcon Investigative Site
      • San Sebastián, Spain, 20014
        • Alcon Investigative Site
      • Valencia, Spain, 46015
        • Alcon Investigative Site
    • Cadiz
      • Jerez De La Frontera, Cadiz, Spain, 11408
        • Alcon Investigative Site
      • Brighton, United Kingdom, BN2 5BF
        • Alcon Investigative Site
      • London, United Kingdom, EC1V2PD
        • Alcon Investigative Site
    • Carshalton
      • Sutton, Carshalton, United Kingdom, SM5 1AA
        • Alcon Investigative Site

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

22 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Description

Key Inclusion Criteria:

  • Diagnosed with bilateral cataracts
  • Planned small incision cataract removal surgery
  • Able to comprehend and willing to sign a statement of informed consent and complete all required post-implantation visits

Key Exclusion Criteria:

  • Subjects who may reasonably be expected to require ocular surgical treatment or refractive surgical procedures at any time during the study
  • Certain eye conditions, as specified in the protocol, including but not limited to glaucoma, diabetic retinopathy, diabetic macular edema, and macular degeneration
  • Pregnant or lactating, current or planned during the course of the study

Other protocol-defined inclusion/exclusion criteria may apply.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Clareon IOL
Clareon® aspheric hydrophobic acrylic intraocular lens implanted with the Alcon Monarch III-D delivery system as a replacement of the human crystalline lens during cataract surgery
Foldable, single-piece posterior chamber IOL intended for long-term use over the lifetime of the cataract patient. This device is CE-marked (European Conformity) in the countries with participating Investigators.
Other Names:
  • Model SY60WF
Routine small incision cataract surgery with unilateral IOL implantation

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percentage of First Implanted Eyes Achieving Best-Corrected Distance Visual Acuity (BCDVA) of 0.3 Logarithm Minimum Angle of Resolution (logMAR) or Better at 1 Year Post-Implantation
Time Frame: Year 1 post-implantation from second eye surgery
Visual acuity (VA) was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic Early Treatment Diabetic Retinopathy Study (ETDRS) charts at a distance of 4 meters (m) from the eye. VA was measured in logarithm of the minimum angle of resolution (logMAR), with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. VA was compared to the historical safety and performance endpoint (SPE) of 92.5% [as reported in the European Standard International Organization for Standardization (EN ISO) 11979-7:2014].
Year 1 post-implantation from second eye surgery
Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 1 Year Post-Implantation
Time Frame: Year 1 post-implantation from second eye surgery
VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. VA was compared to the historical SPE of 92.5% (EN ISO 11979-7:2014).
Year 1 post-implantation from second eye surgery

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-Implantation
Time Frame: Year 2 post-implantation from second eye surgery
VA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.
Year 2 post-implantation from second eye surgery
Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-Implantation
Time Frame: Year 2 post-implantation from second eye surgery
VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.
Year 2 post-implantation from second eye surgery
Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-Implantation
Time Frame: Year 3 post-implantation from second eye surgery
VA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.
Year 3 post-implantation from second eye surgery
Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-Implantation
Time Frame: Year 3 post-implantation from second eye surgery
VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.
Year 3 post-implantation from second eye surgery
Uncorrected Distance Visual Acuity (UCDVA) at 1 Year Post-Implantation - First Implanted Eye
Time Frame: Year 1 post-implantation from second eye surgery
VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 diopter (D) spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
Year 1 post-implantation from second eye surgery
UCDVA at 1 Year Post-Implantation - Second Implanted Eye
Time Frame: Year 1 post-implantation from second eye surgery
VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
Year 1 post-implantation from second eye surgery
UCDVA at 2 Years Post-Implantation - First Implanted Eye
Time Frame: Year 2 post-implantation from second eye surgery
VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
Year 2 post-implantation from second eye surgery
UCDVA at 2 Years Post-Implantation - Second Implanted Eye
Time Frame: Year 2 post-implantation from second eye surgery
VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
Year 2 post-implantation from second eye surgery
UCDVA at 3 Years Post-Implantation - First Implanted Eye
Time Frame: Year 3 post-implantation from second eye surgery
VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
Year 3 post-implantation from second eye surgery
UCDVA at 3 Years Post-Implantation - Second Implanted Eye
Time Frame: Year 3 post-implantation from second eye surgery
VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
Year 3 post-implantation from second eye surgery

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Study Director: Sr. Clinical Trial Lead, CDMA, Surgical, Alcon Research, LLC

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

March 14, 2018

Primary Completion (Actual)

October 18, 2019

Study Completion (Actual)

September 23, 2021

Study Registration Dates

First Submitted

October 18, 2017

First Submitted That Met QC Criteria

October 18, 2017

First Posted (Actual)

October 20, 2017

Study Record Updates

Last Update Posted (Actual)

August 21, 2023

Last Update Submitted That Met QC Criteria

August 17, 2023

Last Verified

August 1, 2023

More Information

Terms related to this study

Additional Relevant MeSH Terms

Other Study ID Numbers

  • ILJ466-P003

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Cataract

Clinical Trials on Clareon® IOL