Post-Market Clinical Investigation of the Clareon® IOL
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Sydney, Australia, 2067
- Alcon Investigative Site
-
-
Victoria
-
Footscray, Victoria, Australia, 3011
- Alcon Investigative Site
-
Mornington, Victoria, Australia, 3931
- Alcon Investigative Site
-
-
-
-
-
Lyon, France, 69003
- Alcon Investigative Site
-
Paris, France, 75014
- Alcon Investigative Site
-
-
Cedex
-
Brest, Cedex, France, 29609
- Alcon Investigative Site
-
-
-
-
-
Frankfurt, Germany, 60590
- Alcon Investigative Site
-
Heidelberg, Germany, 69120
- Alcon Investigative Site
-
-
-
-
-
Perugia, Italy, 06129
- Alcon Investigative Site
-
Pisa, Italy, 56124
- Alcon Investigative Site
-
-
-
-
-
Amsterdam, Netherlands, 1105AZ
- Alcon Investigative Site
-
-
-
-
-
Alcalá De Henares, Spain, 28805
- Alcon Investigative Site
-
Oviedo, Spain, 33012
- Alcon Investigative Site
-
San Sebastián, Spain, 20014
- Alcon Investigative Site
-
Valencia, Spain, 46015
- Alcon Investigative Site
-
-
Cadiz
-
Jerez De La Frontera, Cadiz, Spain, 11408
- Alcon Investigative Site
-
-
-
-
-
Brighton, United Kingdom, BN2 5BF
- Alcon Investigative Site
-
London, United Kingdom, EC1V2PD
- Alcon Investigative Site
-
-
Carshalton
-
Sutton, Carshalton, United Kingdom, SM5 1AA
- Alcon Investigative Site
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Description
Key Inclusion Criteria:
- Diagnosed with bilateral cataracts
- Planned small incision cataract removal surgery
- Able to comprehend and willing to sign a statement of informed consent and complete all required post-implantation visits
Key Exclusion Criteria:
- Subjects who may reasonably be expected to require ocular surgical treatment or refractive surgical procedures at any time during the study
- Certain eye conditions, as specified in the protocol, including but not limited to glaucoma, diabetic retinopathy, diabetic macular edema, and macular degeneration
- Pregnant or lactating, current or planned during the course of the study
Other protocol-defined inclusion/exclusion criteria may apply.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Clareon IOL
Clareon® aspheric hydrophobic acrylic intraocular lens implanted with the Alcon Monarch III-D delivery system as a replacement of the human crystalline lens during cataract surgery
|
Foldable, single-piece posterior chamber IOL intended for long-term use over the lifetime of the cataract patient.
This device is CE-marked (European Conformity) in the countries with participating Investigators.
Other Names:
Routine small incision cataract surgery with unilateral IOL implantation
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Percentage of First Implanted Eyes Achieving Best-Corrected Distance Visual Acuity (BCDVA) of 0.3 Logarithm Minimum Angle of Resolution (logMAR) or Better at 1 Year Post-Implantation
Time Frame: Year 1 post-implantation from second eye surgery
|
Visual acuity (VA) was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic Early Treatment Diabetic Retinopathy Study (ETDRS) charts at a distance of 4 meters (m) from the eye.
VA was measured in logarithm of the minimum angle of resolution (logMAR), with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity.
VA was compared to the historical safety and performance endpoint (SPE) of 92.5% [as reported in the European Standard International Organization for Standardization (EN ISO) 11979-7:2014].
|
Year 1 post-implantation from second eye surgery
|
|
Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 1 Year Post-Implantation
Time Frame: Year 1 post-implantation from second eye surgery
|
VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity.
VA was compared to the historical SPE of 92.5% (EN ISO 11979-7:2014).
|
Year 1 post-implantation from second eye surgery
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-Implantation
Time Frame: Year 2 post-implantation from second eye surgery
|
VA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity.
No hypothesis testing was prespecified for this endpoint.
|
Year 2 post-implantation from second eye surgery
|
|
Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-Implantation
Time Frame: Year 2 post-implantation from second eye surgery
|
VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity.
No hypothesis testing was prespecified for this endpoint.
|
Year 2 post-implantation from second eye surgery
|
|
Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-Implantation
Time Frame: Year 3 post-implantation from second eye surgery
|
VA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity.
No hypothesis testing was prespecified for this endpoint.
|
Year 3 post-implantation from second eye surgery
|
|
Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-Implantation
Time Frame: Year 3 post-implantation from second eye surgery
|
VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity.
No hypothesis testing was prespecified for this endpoint.
|
Year 3 post-implantation from second eye surgery
|
|
Uncorrected Distance Visual Acuity (UCDVA) at 1 Year Post-Implantation - First Implanted Eye
Time Frame: Year 1 post-implantation from second eye surgery
|
VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
+0.25 diopter (D) spherical power was applied to correct for optical infinity.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity.
No hypothesis testing was prespecified for this endpoint.
|
Year 1 post-implantation from second eye surgery
|
|
UCDVA at 1 Year Post-Implantation - Second Implanted Eye
Time Frame: Year 1 post-implantation from second eye surgery
|
VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
+0.25 D spherical power was applied to correct for optical infinity.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity.
No hypothesis testing was prespecified for this endpoint.
|
Year 1 post-implantation from second eye surgery
|
|
UCDVA at 2 Years Post-Implantation - First Implanted Eye
Time Frame: Year 2 post-implantation from second eye surgery
|
VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
+0.25 D spherical power was applied to correct for optical infinity.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity.
No hypothesis testing was prespecified for this endpoint.
|
Year 2 post-implantation from second eye surgery
|
|
UCDVA at 2 Years Post-Implantation - Second Implanted Eye
Time Frame: Year 2 post-implantation from second eye surgery
|
VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
+0.25 D spherical power was applied to correct for optical infinity.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity.
No hypothesis testing was prespecified for this endpoint.
|
Year 2 post-implantation from second eye surgery
|
|
UCDVA at 3 Years Post-Implantation - First Implanted Eye
Time Frame: Year 3 post-implantation from second eye surgery
|
VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
+0.25 D spherical power was applied to correct for optical infinity.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity.
No hypothesis testing was prespecified for this endpoint.
|
Year 3 post-implantation from second eye surgery
|
|
UCDVA at 3 Years Post-Implantation - Second Implanted Eye
Time Frame: Year 3 post-implantation from second eye surgery
|
VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye.
+0.25 D spherical power was applied to correct for optical infinity.
VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart.
A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity.
No hypothesis testing was prespecified for this endpoint.
|
Year 3 post-implantation from second eye surgery
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Study Director: Sr. Clinical Trial Lead, CDMA, Surgical, Alcon Research, LLC
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- ILJ466-P003
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Cataract
-
NCT07589894Enrolling by invitationCataract | Cataract Bilateral | Cataract Surgery | Cataract (Post-operative Cataract Surgery Follow-up)
-
NCT07223866Not yet recruitingNuclear Cataract | Cataract Surgery Anesthesia | Cataract and IOL Surgery | Cataract Surgery Experience
-
NCT07262866Active, not recruitingCataract Surgery | Cataract Patients | Cataract After Surgery
-
NCT07317661Enrolling by invitationIntraocular Lens | Cataract Extraction | Cataract Surgery | Eye Disorders | Cataract and IOL Surgery | Cataract Surgery Experience
-
NCT05832749RecruitingBilateral Cataract | Unilateral Cataract
-
NCT07513402Not yet recruitingCataract (Post-operative Cataract Surgery Follow-up)
-
NCT03931629CompletedCataract | Cataract Mature | Cataract, Nuclear
-
NCT07468877RecruitingCataract Bilateral | Nuclear Sclerosis of the Lens | Phacoemulsification Cataract Surgery
-
NCT05779306CompletedCataract Complicated | Cataract Complications Operations
-
NCT06611670Not yet recruitingCataract | Cataract Bilateral | Cataract Surgery
Clinical Trials on Clareon® IOL
-
NCT03824028Completed
-
NCT06717607Completed
-
NCT07589894Enrolling by invitationCataract | Cataract Bilateral | Cataract Surgery | Cataract (Post-operative Cataract Surgery Follow-up)
-
NCT05129566CompletedCorneal Astigmatism | Cataract Senile | Myopia, High-Grade
-
NCT06555289Recruiting
-
NCT06784063Completed
-
NCT05796453Active, not recruitingPresbyopia | Astigmatism | Aphakia