Enhancing SUV Accuracy of PET/MR for Clinical Trial Qualification
Standardized Uptake Values (SUVs), normalized activity concentration, measured using PET/MR have inaccuracies ≥ 20% which exceeds National Cancer Institute / American College of Radiology Imaging Network (NCI/ACRIN), Radiological Society of North America / Quantitative Imaging Biomarkers Alliance (RSNA/QIBA) specifications and disqualifies PET/MR from multicenter or cooperative group clinical trials. High inaccuracy is primarily due to poor attenuation correction (AC) owing to lack of computed tomography (CT) data. This study will develop acquisition and analyses methods to synthesize CT images from MR data that can be used to achieve SUVs that are within 5% of those obtained using PET/CT (reference standard), thus meeting accuracy requirements needed to qualify for multicenter trials.
The overall goal of this research project is to validate clinically practical methods for producing MR-based attenuation correction information which is needed to produce quantitatively accurate PET images from a PET/MR scanner. Existing commercial PET/MR systems use methods that are inaccurate.
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Objective The primary objective is to demonstrate that, using the new acquisition and analysis methods for MR-AC, PET SUVs in lesions and normal tissues can be measured using PET/MR and be within 5% agreement of those measured using PET/CT.
The secondary objective is to demonstrate visual and quantitative agreement between synthesized CT images generated from MR data and the reference, measured CT images.
Study Design This study would like to enroll patients receiving a clinically indicated PET/CT scan. The patients will be asked to agree to a receive research PET/MR scan within the study which requires additional time and potential MR risks for the patient. It does not entail extra injections or radiation exposure. Research acquisition and processing will be performed on the PET/MR data to create PET images that are expected to have quantitatively accurate SUVs. These will be compared to SUVs from the clinical PET/CT which will serve as the reference standard.
Outcome By bringing together cutting-edge advances in both MR acquisition and image analyses, the successful completion of these aims will achieve SUVs that are within 5% of those obtained with PET/CT (reference standard) with clinically appropriate acquisition time, image quality, and diagnostic accuracy, so that PET/MR systems meet SUV accuracy requirements needed to qualify for cooperative group clinical trials.
Study Type
Study Type
Enrollment (Actual)
Enrollment
Contacts and Locations
Study Locations
-
-
Ohio
-
Cleveland, Ohio, United States, 44106
- University Hospitals Seidman Cancer Center, Case Comprehensive Cancer Center
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- Receiving PET/CT at University Hospitals Cleveland Medical Center (UHCMC) for a clinical indication
- No Contraindications to undergo MR as assessed using University Hospitals (UH) Radiology standard MR assessment form
- Has the ability to understand and willingness to sign a written informed consent
- The circumference of the volunteer in the body section, as determined using a measuring tape, to be scanned must be less than or equal to 110 cm to avoid field of view limitations on the PET/MR
Exclusion Criteria:
- Patient size or circumference greater than the MR gantry of the PET/MR.
- Pregnancy or lactation.
- Contraindications to undergo MR as assessed using UH Radiology standard MR assessment form.
- Claustrophobia or inability to tolerate MR examination (lay still for approximately 1 hour and hold breath intermittently).
- Individuals who are not willing or capable of giving informed consent or assent (with legal guardian consent).
Study Plan
How is the study designed?
Design Details
Number of groups / cohorts
Cohorts and Interventions
Group / CohortGroup / Cohort |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
PET/MR + PET/CT
The interventions for participating in this research are centered around steps needed to safely and ethically collect a research PET/MR scan following a standard-of-care PET/CT scan.
The patient will be imaged in at least one of several standard anatomic areas: head/neck, thorax, abdomen, pelvis or whole-body.
|
The Philips Ingenuity PET/MR is used for research scanning.
Patients receiving a clinically indicated PET/CT, will be approached to be included in this study.
There is no special preparation other than that needed for the prerequisite PET/CT scanning.
At the completion of the PET/CT scanning, the subject will be taken to the PET/MR scanner, which is in close in proximity, for research scanning.
The research scanning would take up to one hour or as tolerated.
Patients will receive PET / CT imaging for the detection of cancer or other clinically indicated anomalies.
The completed study visit is expected to take approximately 2 hours and not longer than 3 hours.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Quantitative agreement between measured CT and CT synthesized from MR data (Hounsfield Units)
Time Frame: Up to 3 hours after beginning scan
|
Demonstrate quantitative agreement between measured CT and CT synthesized from MR data
|
Up to 3 hours after beginning scan
|
|
Quantitative agreement in the derived linear attenuation coefficients at 511 kilo electro volts (keV).
Time Frame: Up to 3 hours after beginning scan
|
Demonstrate quantitative agreement in the derived linear attenuation coefficients at 511 keV.
|
Up to 3 hours after beginning scan
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Percent difference in Standardized Uptake Values (SUVs) in PET/CT scans compared to PET/MR scans
Time Frame: Up to 3 hours after beginning scan
|
Demonstrate that SUVs in lesions and normal tissues measured using MR-based attenuation correction (MR-AC) are within 5% of those measured using PET/CT.
|
Up to 3 hours after beginning scan
|
Collaborators and Investigators
Sponsor
Sponsor
Collaborators
Collaborators
Investigators
Investigators
- Principal Investigator: Raymond F. Muzic, PhD, University Hospitals Seidman Cancer Center, Case Comprehensive Cancer Center
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
Other Study ID Numbers
- CASE3Y18
- R01CA196687 (U.S. NIH Grant/Contract)
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Positron Emission Tomography-Magnetic Resonance (PET-MR)
-
NCT05989997CompletedSolid Tumor | Positron-Emission Tomography(PET)
-
NCT05490264RecruitingSolid Tumor | Positron-Emission Tomography(PET)
-
NCT04605939RecruitingMagnetic Resonance Imaging | Liver Fibrosis | Positron Emission Tomography
-
NCT02988531CompletedCoronary Artery Disease | MRI | Cardiovascular Disease | Magnetic Resonance Imaging | Positron-Emission Tomography | PET Scan
-
NCT04660279CompletedPositron-Emission Tomography
-
NCT00228514CompletedHeart Failure | Rosuvastatin | Positron Emission Tomography (PET)
-
NCT01956409UnknownMagnetic Resonance Imaging | Positron Emission Tomography | Breast Cancer Diagnosis
-
NCT05506566RecruitingTumor | Positron-Emission Tomography
-
NCT05140083RecruitingPositron-Emission Tomography
-
NCT06387381RecruitingPositron-Emission Tomography | FAP | PSMA
Clinical Trials on Positron Emission Test / Magnetic Resonance (PET/MR)
-
NCT03423407Terminated
-
NCT05662540Recruiting
-
NCT01759355CompletedCervical Adenosquamous Carcinoma | Cervical Squamous Cell Carcinoma | Cervix Carcinoma
-
NCT05003830RecruitingAlzheimer Disease | PET/MR
-
NCT04605939RecruitingMagnetic Resonance Imaging | Liver Fibrosis | Positron Emission Tomography
-
NCT01557881TerminatedMelanoma | Lymphoma | Unspecified Childhood Solid Tumor, Protocol Specific | Head and Neck Cancer | Lung Cancer | Unspecified Adult Solid Tumor, Protocol Specific | Colon Cancer | Malignant Neoplasm
-
NCT05761535Completed
-
NCT00741559Completed
-
NCT07213219Not yet recruitingParkinson's Disease (PD)