Development of an Algorithm to Differentiate Heart Murmurs Using Electronic Stethoscopes
Development of an Algorithm to Differentiate Heart Murmurs by Timing, Disease Type, and Severity in Adult Patients Using Electronic Stethoscopes
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Valvular heart disease (VHD), a subclass of structural heart disease (SHD), is a major cause of mortality and reduced quality of life for tens of millions of patients worldwide. Detecting VHD early allows the disease to be carefully monitored so that surgery can be considered as soon as the stenosis progresses to severe. Unfortunately, it is difficult for many clinicians to accurately identify VHD in its early stages and assess its severity through physical examination alone. This difficulty stems from both the subtlety of symptoms associated with VHD, and from clinicians' inability to reliably detect the heart murmurs indicative of VHD using a traditional stethoscope, particularly in the noisy and rushed environments characteristic of many healthcare settings.
AS, mitral regurgitation (MR), and tricuspid regurgitation (TR) all result in reliably reproducible auscultatory findings. Yet despite this, auscultation as a diagnostic procedure is notoriously poor: its accuracy to detect AS, MR, and TR ranges only from 5-40%. In addition, diastolic and continuous murmurs, which are rarer than the systolic murmurs caused by AS, MR and TR, and almost always pathologic, are usually very difficult to detect by frontline clinicians. Diastolic murmurs are typically caused by VHD such as aortic regurgitation (AR), pulmonic regurgitation (PR), mitral stenosis (MS), or tricuspid stenosis (TS), and continuous murmurs are typically caused by non-valvular forms of SHD, such as congenital aortopulmonary window, shunts, and patent ductus arteriosus (PDA). Missing one of these murmurs could lead to significantly delayed diagnosis and care.
Because of how difficult it is for clinicians to reliably screen for SHD through history taking and physical examination, the diagnosis of SHD relies heavily on echocardiography (echos), which is accepted as the gold standard for diagnosis of SHD severity by the cardiology community. The consequence of this is several-fold. First, many patients receive echos that would not be necessary if better screening and assessment tools existed. Second, echocardiography makes up a disproportionately large segment of healthcare expenditure. Third, patients are burdened with the responsibility of getting repeated echos due to the lack of better screening tools. The emotional, physical, and financial burden that accompanies such excessive testing, especially for patients living in underserved communities, cannot be overstated.
Overall, there is a growing need for better preliminary screening tools that would identify these conditions, even when asymptomatic, early on and more consistently, to a degree of accuracy comparable to gold standard echocardiography. Such a tool would improve clinicians' confidence in their assessment during physical exams, which would enable them to help their patients make informed care decisions, offer disease-altering therapies, and at the same time reduce the reliance on, and frequency of, unnecessary and expensive auxiliary workups. Lastly, by facilitating early detection of VHD, morbidity and mortality rates would be expected to decrease due to earlier intervention.
The FDA-approved Eko CORE and Eko DUO electronic stethoscopes offer clinicians a familiar and inexpensive tool that is widely accepted by patients, while at the same time offer sensors and technology that can improve screening and detection of medical conditions such as valve disease, as compared to auscultation with the traditional stethoscope. Both CORE and DUO feature sound amplification during auscultation, which improves the ability for the clinician to detect nuanced changes in heart sounds and patterns. The CORE allows recording of the audio to produce a phonocardiogram (PCG), while the DUO is able to record audio to produce a PCG as well as simultaneous recording of a single-lead electrocardiogram (ECG). Both devices use Bluetooth to wirelessly transmit PCG and ECG data to the Eko mobile application, which allows clinicians to visualize PCG and ECG as they auscultate, as well as to playback heart sound recordings, annotate notes on recorded audio, and save and share recordings. These features uniquely situate Eko electronic stethoscopes to be highly effective as preliminary screening tools in clinical practice.
Eko Devices has developed and received 510(k) FDA approval for a machine-learning (ML) algorithm that can detect the presence or absence of heart murmur in adult patients with VHD, using either the Eko CORE or DUO electronic stethoscopes to record heart sounds. This algorithm was built using a deep neural network model trained on 5,878 PCG recordings collected on Eko CORE and Eko DUO from over 5,318 unique patients to classify heart sound recordings as containing a heart murmur or containing no audible murmur, and to discriminate good quality heart sound recordings from poor quality heart sound recordings. The trained model has been independently validated in a multi-site clinical study that enrolled 681 VHD patients with matched gold-standard echocardiograms. FDA approval was received in January 2020.
The present study seeks to expand murmur detection to include VHD classification through the development of novel ML algorithms that are able to distinguish between systolic vs. diastolic vs. continuous murmurs, as well as classify VHD type and severity, using 4-point auscultation with Eko CORE and DUO electronic stethoscopes to record heart sounds. These new algorithms will integrate the time-synchronized ECG streams from DUO to improve distinction between S1 and S2 heart sounds and improve performance of identifying systolic vs. diastolic time intervals. When combined with the FDA-approved algorithm for signal quality and murmur detection, these new algorithms will enable clinicians to have a complete VHD decision support system, as based on ACC/AHA Patient Management Guidelines, when using Eko stethoscopes in their clinical practice.20 This will arm clinicians with information to help them detect SHD early, make clinical care decisions with confidence and consistency, and improve overall patient outcomes while decreasing cost and increasing efficiency.
Study Type
Study Type
Enrollment (Actual)
Enrollment
Contacts and Locations
Study Locations
-
-
Missouri
-
Springfield, Missouri, United States, 65807
- CoxHealth
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- Adults aged 18 years and older
- Patient or patient's legal healthcare proxy consents to participation
- Have an audible heart murmur in at least one auscultation position, as heard by a clinician
- Undergoing or have undergone a complete (i.e., not limited) echocardiogram
- Willing to have heart recordings done with two different electronic stethoscopes
Exclusion Criteria:
- Patient or proxy is unwilling or unable to give written informed consent
- Unable to complete a full clinical echocardiogram
- No audible heart murmur in at least one auscultation position, as heard by a clinician
- Experiencing a known or suspected acute cardiac event
- Mechanical ventricular support (such as ECMO, LVAD, RVAD, BiVAD, Impella, intra-aortic balloon pumps, TAH, VentrAssist, DuraHeart, HVAD, EVAHEART LVAS, HeartMate, Jarvik 2000)
- Unwilling or unable to follow or complete study procedures
Study Plan
How is the study designed?
Design Details
- Observational Models: Other
- Time Perspectives: Cross-Sectional
Number of groups / cohorts
Cohorts and Interventions
Group / CohortGroup / Cohort |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Aortic Stenosis
Subjects with echo-confirmed AS graded moderate-to-severe or worse
|
Auscultation of heart sounds using electronic stethoscopes
|
|
Mitral Regurgitation
Subjects with echo-confirmed MR graded moderate-to-severe or worse
|
Auscultation of heart sounds using electronic stethoscopes
|
|
Tricuspid Regurgitation
Subjects with echo-confirmed TR graded moderate-to-severe or worse
|
Auscultation of heart sounds using electronic stethoscopes
|
|
Innocent Murmur
Subjects with echo-confirmed trace/trivial valve disease severity
|
Auscultation of heart sounds using electronic stethoscopes
|
|
Diastolic Murmur
Subjects with pathology associated with diastolic murmur (e.g.
AR, PR, MS, TS)
|
Auscultation of heart sounds using electronic stethoscopes
|
|
Continuous Murmur
Subjects with pathology associated with continuous murmur (e.g.
congenital shunts, PDA)
|
Auscultation of heart sounds using electronic stethoscopes
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Murmur Timing
Time Frame: Within two minutes of device use
|
Develop a model that can classify murmurs based on the timing of the murmur in the cardiac cycle, allowing the algorithm to distinguish between systolic, diastolic, and continuous murmurs using time-synchronized ECG and PCG data.
|
Within two minutes of device use
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Systolic Murmur Type Classification
Time Frame: Within two minutes of device use
|
Develop a model that can classify systolic murmurs into disease categories including aortic stenosis (AS), mitral regurgitation (MR), tricuspid regurgitation (TR), other (O), and innocent murmurs (I).
|
Within two minutes of device use
|
|
Systolic Murmur Severity Classification
Time Frame: Within two minutes of device use
|
Develop a model that can grade the severity of AS, MR, and TR following the American Society of Echocardiography (ASE) guidelines as mild, moderate, or severe.
|
Within two minutes of device use
|
Collaborators and Investigators
Sponsor
Sponsor
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- 2020.3
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Aortic Valve Stenosis
-
NCT07620925RecruitingAortic Stenosis | Calcific Aortic Valve Disease | Chronic Coronary Syndrome | Calcific Aortic Stenosis
-
NCT07144189RecruitingAortic Stenosis | Low-gradient Aortic Stenosis
-
NCT07450196Not yet recruiting
-
NCT07278310RecruitingAortic Stenosis | Symptomatic Severe Native Aortic Stenosis
-
NCT05712161Active, not recruitingAortic Stenosis | Aortic Valve Calcification | Severe Aortic Valve Stenosis | Symptomatic Aortic Stenosis
-
NCT07456280Not yet recruitingCardiovascular Diseases | Severe Aortic Valve Stenosis
-
NCT07335900RecruitingSevere Aortic Valve Stenosis
-
NCT07304427CompletedSevere Aortic Stenosis | Transcatheter Aortic Valve Replacemen
-
NCT07479992Recruiting
-
NCT07205068Recruiting
Clinical Trials on Use of Eko CORE and Eko DUO electronic stethoscopes
-
NCT04404452TerminatedHeart Failure | Heart Failure With Preserved Ejection Fraction | Crackles
-
NCT04317521Completed
-
NCT04400435CompletedLeft Ventricular Dysfunction
-
NCT05010642CompletedHeart Diseases | Cardiovascular Diseases | Ventricular Dysfunction | Left Ventricular Dysfunction
-
NCT05459545CompletedAtrial Fibrillation | Heart Murmurs | Murmur, Heart | Innocent Murmurs | Pathologic Murmur
-
NCT04933890CompletedHeart Failure | Left Ventricular Dysfunction
-
NCT05010655CompletedHeart Diseases | Cardiovascular Diseases | Ventricular Dysfunction | Left Ventricular Dysfunction
-
NCT06382207Recruiting
-
NCT05176899RecruitingHeart Murmurs | Murmur, Heart | Innocent Murmurs | Pathologic Murmur
-
NCT05987670Active, not recruitingHeart Failure | Atrial Fibrillation | Heart Valve Diseases | Heart Murmurs | Congestive Heart Failure | Heart Failure With Reduced Ejection Fraction