Point-of-Care Bedside Lung Ultrasound Examination Advanced Trial Protocol (PoCBLUEPlus)

25. března 2015 aktualizováno: Hong Liu, Nanjing Medical University

A Prospective, Multi-center, Open-label and Double-blind, Standard-controlled, Non-inferiority, Diagnostic Study of Lung Ultrasound for Management of Mechanical Ventilation in Neonatal Acute Respiratory Distress Syndrome

The purpose of this study was to evaluate the availability and diagnostic accuracy of point-of-care bedside lung ultrasound examination in management of mechanical ventilation in neonatal acute respiratory distress syndrome.

Přehled studie

Detailní popis

Neonatal acute respiratory distress syndrome (ARDS) is a critical condition requiring dynamic evaluation and interventions. Point-of-care bedside lung ultrasound examination (PoC-BLUE) is a noninvasive, readily available imaging modality that can complement physical and clinical evaluation.

At any time, most neonates and infants with ARDS in medical intensive care units (ICUs) require mechanical ventilation, making it one of the most frequently used critical care technologies. However, difficulties with regard to the accurate diagnosis of ARDS before administration of mechanical ventilation, dynamic monitoring of treatment effects during administration of mechanical ventilation, and decision-making of timing in weaning from mechanical ventilation, are often encountered in the majority neonates and infants who require mechanical ventilation.

Hence, techniques that expedite and advance the knowledge of the administration of mechanical ventilation should have an important clinical significance in the diagnosis, treatment and prognosis of ARDS.

Preliminary researches have suggested that BLUE has a high diagnostic accuracy in patients with acute respiratory failure, and has the potential to quantify the rate and degree of diaphragm thinning during mechanical ventilation, which may be useful to predict extubation success or failure during either spontaneous breathing (SB) or pressure support (PS) trials.

Unfortunately, few studies focus on the availability of PoC-BLUE in management of mechanical ventilation in neonatal acute respiratory distress syndrome. Also, they call into question the possible association of the demographic and clinical confounders with the diagnostic accuracy of PoC-BLUE for diagnosis, monitoring and prognosis of ARDS in the whole process of mechanical ventilation.

Given that a considerable need for a timely and dynamic diagnosis of severe condition and therapy evaluation during mechanical ventilation has been triggered to integrate the currently available bulk of knowledge and information, the objective of this study is to investigate the availability and diagnostic accuracy of PoC-BLUE Plus protocol in management of mechanical ventilation in ARDS.

Typ studie

Intervenční

Zápis (Aktuální)

1000

Fáze

  • Nelze použít

Kritéria účasti

Výzkumníci hledají lidi, kteří odpovídají určitému popisu, kterému se říká kritéria způsobilosti. Některé příklady těchto kritérií jsou celkový zdravotní stav osoby nebo předchozí léčba.

Kritéria způsobilosti

Věk způsobilý ke studiu

1 den až 1 měsíc (Dítě)

Přijímá zdravé dobrovolníky

Ne

Pohlaví způsobilá ke studiu

Všechno

Popis

Inclusion Criteria:

  • All infants who present to the ICU with respiratory symptoms suspicious for ARDS, according to Pediatric Acute Respiratory Distress Syndrome: Consensus Recommendations From the Pediatric Acute Lung Injury Consensus Conference, which is proposed by The Pediatric Acute Lung Injury Consensus Conference Group.
  • In whom the treating Collaborative Review Groups of Poc-BLUE-Plus protocol believe would benefit from diagnostic imaging.

Exclusion Criteria:

  • Infants who arrive at the ICU with a previously performed chest radiography
  • Unstable infants with life-threatening injuries who require ongoing resuscitation

Studijní plán

Tato část poskytuje podrobnosti o studijním plánu, včetně toho, jak je studie navržena a co studie měří.

Jak je studie koncipována?

Detaily designu

  • Primární účel: Diagnostický
  • Přidělení: Randomizované
  • Intervenční model: Paralelní přiřazení
  • Maskování: Dvojnásobek

Zbraně a zásahy

Skupina účastníků / Arm
Intervence / Léčba
Experimentální: Lung Ultrasound
In infants allocated to this arm Lung ultrasound for detection of ARDS will be performed before chest radiography.
Lung ultrasound evaluation is performed after clinical assessment and before chest radiography
Ostatní jména:
  • Bedside Ultrasonography System
Aktivní komparátor: Chest Radiography
In infants allocated to this arm chest radiography will be performed for the detection of indirect signs of ARDS without ultrasound evaluation.
Chest Radiography is performed after clinical evaluation without using ultrasound assessment.
Ostatní jména:
  • Chest X-ray System

Co je měření studie?

Primární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Accuracy of Lung Ultrasound in Neonatal ARDS
Časové okno: Accuracy will be measured at the end of clinical evaluation in the ICU, an expected average of 4 days
Accuracy of lung ultrasound and chest radiography will be measured using as gold standard the independent evaluation of the entire medical records by two expert emergency physicians blinded to the lung ultrasound results and radiographic reports (digitalized chest radiography images will be available).
Accuracy will be measured at the end of clinical evaluation in the ICU, an expected average of 4 days

Sekundární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Accuracy of Lung Ultrasound in the Ventilator-Associated Pneumonia
Časové okno: Accuracy will be measured at the end of clinical evaluation in the ICU, an expected average of 4 days
Accuracy of lung ultrasound and chest radiography will be measured using as gold standard the independent evaluation of the entire medical records by two expert emergency physicians blinded to the lung ultrasound results and radiographic reports (digitalized chest radiography images will be available).
Accuracy will be measured at the end of clinical evaluation in the ICU, an expected average of 4 days
Accuracy of Lung Ultrasound in the Weaning of Ventilatory Support
Časové okno: Accuracy will be measured at the end of clinical evaluation in the ICU, an expected average of 4 days
Accuracy of lung ultrasound and chest radiography will be measured using as gold standard the independent evaluation of the entire medical records by two expert emergency physicians blinded to the lung ultrasound results and radiographic reports (digitalized chest radiography images will be available).
Accuracy will be measured at the end of clinical evaluation in the ICU, an expected average of 4 days

Spolupracovníci a vyšetřovatelé

Zde najdete lidi a organizace zapojené do této studie.

Publikace a užitečné odkazy

Osoba odpovědná za zadávání informací o studiu tyto publikace poskytuje dobrovolně. Mohou se týkat čehokoli, co souvisí se studiem.

Termíny studijních záznamů

Tato data sledují průběh záznamů studie a předkládání souhrnných výsledků na ClinicalTrials.gov. Záznamy ze studií a hlášené výsledky jsou před zveřejněním na veřejné webové stránce přezkoumány Národní lékařskou knihovnou (NLM), aby se ujistily, že splňují specifické standardy kontroly kvality.

Hlavní termíny studia

Začátek studia

1. března 2014

Primární dokončení (Aktuální)

1. března 2015

Dokončení studie (Aktuální)

1. března 2015

Termíny zápisu do studia

První předloženo

22. března 2015

První předloženo, které splnilo kritéria kontroly kvality

25. března 2015

První zveřejněno (Odhad)

31. března 2015

Aktualizace studijních záznamů

Poslední zveřejněná aktualizace (Odhad)

31. března 2015

Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality

25. března 2015

Naposledy ověřeno

1. března 2015

Více informací

Termíny související s touto studií

Informace o lécích a zařízeních, studijní dokumenty

Studuje lékový produkt regulovaný americkým FDA

Ne

Studuje produkt zařízení regulovaný americkým úřadem FDA

Ne

produkt vyrobený a vyvážený z USA

Ne

Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .