Denne side blev automatisk oversat, og nøjagtigheden af ​​oversættelsen er ikke garanteret. Der henvises til engelsk version for en kildetekst.

Accuracy of Noninvasive Pulse Oximeter Sensor (LNCS DBI)

28. juni 2017 opdateret af: Masimo Corporation
In this study, the level of oxygen within the blood will be reduced in a controlled manner by reducing the percentage of oxygen the study volunteer breathes. The accuracy of a noninvasive pulse oximeter sensor will be assessed by comparison to the oxygen saturation measurements from a laboratory blood gas analyzer.

Studieoversigt

Status

Afsluttet

Betingelser

Intervention / Behandling

Undersøgelsestype

Interventionel

Tilmelding (Faktiske)

13

Fase

  • Ikke anvendelig

Kontakter og lokationer

Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.

Studiesteder

    • California
      • Irvine, California, Forenede Stater, 92618
        • Masimo

Deltagelseskriterier

Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.

Berettigelseskriterier

Aldre berettiget til at studere

18 år til 40 år (Voksen)

Tager imod sunde frivillige

Ingen

Køn, der er berettiget til at studere

Alle

Beskrivelse

Inclusion Criteria:

  • Competent non-smoking adults between the ages of 18 and 40 for each series of tests.
  • Subjects must understand and consent to be in the study.
  • American Society of Anesthesiology Class 1 (Healthy subjects without any systemic disease at all).

Exclusion Criteria:

  • Subjects who have any systemic disease at all.
  • Subjects who do not understand the study and the risks.
  • Smokers.
  • Subjects who are pregnant.
  • Subjects having either signs or history of peripheral ischemia.
  • Others deemed ineligible by the clinical staff.

Studieplan

Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.

Hvordan er undersøgelsen tilrettelagt?

Design detaljer

  • Primært formål: Andet
  • Tildeling: N/A
  • Interventionel model: Enkelt gruppeopgave
  • Maskning: Ingen (Åben etiket)

Våben og indgreb

Deltagergruppe / Arm
Intervention / Behandling
Andet: Test Subject
All subjects are enrolled into the test group and all subjects received the LNCS DBI Sensor.
Noninvasive Measurement of SpO2

Hvad måler undersøgelsen?

Primære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Accuracy of Sensor by Arms Calculation
Tidsramme: 1-5 hours
Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value. In order to obtain the Arms value, the blood oxygen saturation measurement is subtracted from the pulse oximeter oxygen saturation measurement for a number of samples, the average of this difference is computed as the bias. The standard deviation of these differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.
1-5 hours

Samarbejdspartnere og efterforskere

Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.

Datoer for undersøgelser

Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.

Studer store datoer

Studiestart (Faktiske)

18. januar 2012

Primær færdiggørelse (Faktiske)

10. april 2012

Studieafslutning (Faktiske)

10. april 2012

Datoer for studieregistrering

Først indsendt

14. april 2017

Først indsendt, der opfyldte QC-kriterier

20. april 2017

Først opslået (Faktiske)

24. april 2017

Opdateringer af undersøgelsesjournaler

Sidste opdatering sendt (Faktiske)

28. juli 2017

Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier

28. juni 2017

Sidst verificeret

1. juni 2017

Mere information

Begreber relateret til denne undersøgelse

Andre undersøgelses-id-numre

  • TR22194-TP16113

Plan for individuelle deltagerdata (IPD)

Planlægger du at dele individuelle deltagerdata (IPD)?

Ingen

Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter

Studerer et amerikansk FDA-reguleret lægemiddelprodukt

Ingen

Studerer et amerikansk FDA-reguleret enhedsprodukt

Ja

produkt fremstillet i og eksporteret fra U.S.A.

Ingen

Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .

Kliniske forsøg med LNCS DBI Sensor

3
Abonner