Warming up to Sensors: Does Site Warming Improve Continuous Glucose Monitor Performance?
Descripción general del estudio
Estado
Estado
Condiciones
Condiciones
Intervención / Tratamiento
Intervención / Tratamiento
Descripción detallada
Hypotheses:
- Warming of the skin and subcutaneous tissue with the InsuPatch device will improve the accuracy of a CGM as compared to a CGM used alone when both are calibrated with a standard laboratory quality glucose measuring reference standard.
- Warming of the skin and subcutaneous tissue with the InsuPatch device will reduce the lag time of the CGM, as compared to a CGM used alone when both are calibrated with a standard laboratory quality glucose measuring reference standard.
- The combination of the InsuPatch device and a CGM will be well-tolerated and not cause skin irritation.
Tipo de estudio
Tipo de estudio
Inscripción (Actual)
Inscripción
Fase
Fase
- Fase temprana 1
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
Connecticut
-
New Haven, Connecticut, Estados Unidos, 06520
- Yale University School Of Medicine
-
-
Criterios de participación
Criterio de elegibilidad
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Descripción
Inclusion Criteria:
- Clinical diagnosis of type 1 diabetes and using daily insulin therapy for at least 12 months.
- Age 8 years to less than 45 years old.
- Subject understands the study protocol and agree to comply with it.
- Subject is on insulin pump or injection-based basal-bolus therapy
- Subject comprehends written English.
- For females, subject not intending to become pregnant during the study.
- Informed Consent Form signed by the subject
- No history of altered skin integrity, poor wound healing, adhesive allergy, or previous infection from sensor wear.
- Hemoglobin A1C ≤10%
Exclusion Criteria:
- The presence of a significant medical disorder that in the judgment of the investigator will affect the wearing of the sensors or the completion of any aspect of the protocol
The presence of any of the following diseases:
a. Current use of systemic corticosteroids (within the last week) b. Other major illness that in the judgment of the investigator might interfere with the completion of the protocol i. Adequately treated thyroid disease and celiac disease do not exclude subjects from enrollment
- Inpatient psychiatric treatment in the past 6 months for either the subject or the subject's primary care giver (i.e., parent or guardian)
- Current or past use of a continuous glucose sensor is not an exclusion criterion.
- Active skin condition that would affect sensor placement
- Use of acetaminophen or Vitamin C supplement within the last 7 days
- Subject is deemed by the Investigator to be unwilling or unable to follow the protocol;
Pregnancy A negative urine pregnancy test will be required for all premenopausal women who are not surgically sterile. Subjects who become pregnant will be discontinued from the study.
-
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Otro
- Asignación: No aleatorizado
- Modelo Intervencionista: Asignación paralela
- Enmascaramiento: Ninguno (etiqueta abierta)
Número de brazos
Armas e Intervenciones
Grupo de participantes/brazoGrupo de participantes/brazo |
Intervención / TratamientoIntervención / Tratamiento |
|---|---|
|
Comparador activo: Sensor alone
Glucose sensor site will not have Insupatch
|
Glucose sensor inserted without warming patch
|
|
Experimental: Sensor with Insupatch
Glucose sensor site with Insupatch
|
Device which applies local heating at 40 degrees Celsius to the vicinity of the subcutaneous site of glucose sensor.
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Point accuracy of the continuous glucose monitor plus Insupatch (CGM+IP) by mean absolute relative difference (MARD) calculation of sensors with and without IP use.
Periodo de tiempo: 4 hours after lunch, 8 hours after breakfast
|
Point accuracy of the continuous glucose monitor plus Insupatch (CGM+IP) by mean absolute relative difference (MARD) calculation of sensors with and without IP use, comparing sensor glucose readings and signal data to serum Yellow Springs Instrument (YSI) glucose measurements, with each subject acting as his or her own control.
|
4 hours after lunch, 8 hours after breakfast
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Colaboradores
Colaboradores
Investigadores
Investigadores
- Investigador principal: Camille Michaud, MD, Yale University
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio
Inicio del estudio
Finalización primaria (Actual)
Finalización primaria
Finalización del estudio (Actual)
Finalización del estudio
Fechas de registro del estudio
Enviado por primera vez
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Estimar)
Publicado por primera vez
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización publicada
Última actualización enviada que cumplió con los criterios de control de calidad
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
Otros números de identificación del estudio
- 1306012208
- R01DK085618-01 (Subvención/contrato del NIH de EE. UU.)
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .