Esta página se tradujo automáticamente y no se garantiza la precisión de la traducción. por favor refiérase a versión inglesa para un texto fuente.

WEDGES: Coronary Lesion Preparation With Enhanced Dilatation for Good Expansion of Stent: Assessment of a New Scoring Balloon by Quantitative Stent Enhanced Visualization (WEDGES)

18 de agosto de 2021 actualizado por: Stephane Carlier

WEDGES: a Prospective, Multi-center Randomized Trial for the Evaluation of Coronary Lesion Preparation With Enhanced Dilatation for Good Expansion of Stent: Assessment of a New Scoring Balloon by Quantitative Stent Enhanced Visualization

Stent under-expansion during percutaneous coronary intervention (PCI) is one of the suboptimal results associated with a higher risk of restenosis and thrombosis. In order to obtain an optimal stent expansion, especially in lesions with calcific or fibrotic plaque, the appropriate lesion preparation before stent implantation is a crucial phase.

For this, different devices are used in everyday practices, including non-compliant balloon (NC), scoring / cutting balloons or rotational atherectomy; however, their effectiveness depends on the experience and skill of the operator. The purpose of this study is to assess the feasibility, efficacy and safety of a novel scoring balloon intended to treat calcific or fibrotic plaque of a coronary lesion

Descripción general del estudio

Tipo de estudio

Intervencionista

Inscripción (Anticipado)

100

Fase

  • No aplica

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Estudio Contacto

Copia de seguridad de contactos de estudio

Ubicaciones de estudio

      • Genk, Bélgica, 3600
        • Reclutamiento
        • UZ Genk
        • Contacto:
          • Jo Dens
      • Ghent, Bélgica, 9000
        • Reclutamiento
        • UZ Ghent
        • Contacto:
          • Peter Kayaert
      • La Louvière, Bélgica, 7100
        • Reclutamiento
        • CHU Jolimont
        • Contacto:
          • Claudiu Ungureanu
      • Liège, Bélgica, 4000
        • Reclutamiento
        • CHU Liège
        • Contacto:
          • Lancelotti
      • Mons, Bélgica, 7000

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

18 años y mayores (Adulto, Adulto Mayor)

Acepta Voluntarios Saludables

Géneros elegibles para el estudio

Todos

Descripción

Inclusion Criteria:

  • Patient written consent is given A de novo lesion
  • Mild to moderate burden of calcifications detected on two orthogonal views using normal radiation power

Exclusion Criteria:

  • Exclusion criteria patient-related:

    1. Patient <18 years old
    2. Pregnant female
    3. Contraindication to dual antiplatelet therapy
    4. Thrombocytopenia <100 000
    5. Major surgical intervention planned

Exclusion criteria lesion-related:

  1. Significant left main lesion
  2. Bifurcating lesion with a significant side branch to be treated
  3. Lesion length > 25 mm
  4. Chronic total occlusion
  5. Lesion with heavy calcifications where a rotational atherectomy or lithotripsy should be planned
  6. Lesion in a graft
  7. In-stent restenosis lesion
  8. Lesion responsible of a STEMI

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Propósito principal: Tratamiento
  • Asignación: Aleatorizado
  • Modelo Intervencionista: Asignación paralela
  • Enmascaramiento: Doble

Armas e Intervenciones

Grupo de participantes/brazo
Intervención / Tratamiento
Comparador activo: Conventional balloon predilation
the target lesion can be prepared according to standard clinical practice using SC balloons sized 0.8-1:1. Stenting using a drug-eluting stent (DES) is performed and the inflation pressure for delivery is noted, as well as the expected balloon diameter at that pressure. 2 enhanced stent imaging acquisitions (StentBoost or equivalent depending on the X-ray system used), are acquired in 2 orthogonal views. Post dilatation can be performed at operator discretion followed by a final acquisition of 2 similar orthogonal views with enhanced stent imaging. Intravascular imaging is left to the operator's discretion and is encouraged
Conventional balloon
Experimental: Scoring balloon predilation
the target lesion must be prepared with a scoring balloon (SC) inflated 3 times at nominal pressure. To allow the scoring balloon to cross the lesion, it might be necessary to predilate with a small compliant balloon. The size of the scoring balloon to be used for is based on the size of the vessel. Stenting using a DES is performed and the inflation pressure for delivery is noted, as well as the expected balloon diameter at that pressure. 2 enhanced stent imaging acquisitions (StentBoost or equivalent depending on the X-ray system used), are acquired in 2 orthogonal views. Post dilatation can be performed at operator discretion followed by a final acquisition of 2 similar orthogonal views with enhanced stent imaging. Intravascular imaging is left to the operator's discretion and is encouraged
Scoring balloon

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Stent expansion in calcified lesions assessed by stent diameter calculation using quantitative and intravascular analysis after lesion preparation using a scoring balloon is superior to conventional lesion preparation
Periodo de tiempo: Intravascular ultrasound (IVUS) or quantitative analysis is directly done after procedure
The measure of stent expansion in calcified coronary lesions as assessed by quantitative coronary angiography, enhanced stent imaging and intravascular ultrasound after lesion preparation using the novel scoring semi-compliant and noncompliant balloons are superior (p<0.05) to semi compliant and non-compliant balloons.
Intravascular ultrasound (IVUS) or quantitative analysis is directly done after procedure

Medidas de resultado secundarias

Medida de resultado
Periodo de tiempo
Rate of successful predilatation of calcified lesions of a scoring semi-compliant/non-compliant balloon and semi-compliant (SC)/ non-compliant (NC) balloon
Periodo de tiempo: Directly after procedure
Directly after procedure
Rate of residual under-expansion of stent after post-dilatation with a non-compliant balloon
Periodo de tiempo: Directly after procedure
Directly after procedure
Rate of procedural complication (composite of cardiac death, myocardial infarction, coronary artery perforation and cardiac tamponade) of a scoring balloon and a SC/NC balloon followed by stenting.
Periodo de tiempo: 1 month follow up
1 month follow up

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Patrocinador

Investigadores

  • Director de estudio: Stéphane Carlier, University of Mons

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

1 de enero de 2020

Finalización primaria (Anticipado)

30 de julio de 2022

Finalización del estudio (Anticipado)

30 de octubre de 2022

Fechas de registro del estudio

Enviado por primera vez

13 de agosto de 2021

Primero enviado que cumplió con los criterios de control de calidad

18 de agosto de 2021

Publicado por primera vez (Actual)

24 de agosto de 2021

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

24 de agosto de 2021

Última actualización enviada que cumplió con los criterios de control de calidad

18 de agosto de 2021

Última verificación

1 de agosto de 2021

Más información

Términos relacionados con este estudio

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

NO

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

producto fabricado y exportado desde los EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .

Suscribir