Plating of Distal Femur Fractures by Far Cortical Locking Using MotionLoc Screws
MotionLoc Study: Healing of Distal Femur Fractures Stabilized With a Flexible Plating Construct Using MotionLoc Screws
Aperçu de l'étude
Statut
Les conditions
Description détaillée
Rigid locked plating constructs can suppress fracture healing, particularly at the near cortex adjacent to the plate where interfragmentary motion is minimal. Dynamic fixation with Far Cortical Locking (FCL) screws reduces construct stiffness and induces axial interfragmentary motion to stimulate symmetric callus formation and healing. Two versions of FCL screws are commercially available, but the clinical durability of this novel concept has not been documented to date. This prospective observational study documented our early clinical experience with MotionLoc® FCL screws for stabilization of distal femur fractures to assess their durability and potential complications.
Thirty-two consecutive patients with 33 distal femur fractures (AO/OTA types 33-A and 33-C) were prospectively enrolled at three trauma centers. Fractures were stabilized by plate osteosynthesis with MotionLoc® FCL screws without supplemental bone graft or bone morphogenic proteins. Thirty patients with 31 fractures were available for follow-up until union or revision. Follow-up visits at 6, 12, and 24 weeks comprised functional and radiographic assessment of implant fixation and fracture healing, including computed tomography scans at week 12. The primary endpoint was fracture healing in absence of complications and revision.
There was no incidence of implant breakage or diaphyseal fixation failure. Thirty of 31 fractures healed within 15.6 ± 6.2 weeks, as evident by bridging callus and pain-free load bearing. There were two revisions, one at 5 days post surgery to correct a mal-rotation, and one at 6 months post surgery to revise a non-union. Periosteal callus distribution at week 6 was symmetrical, with similar amounts of callus at the medial cortex (35%) anterior cortex (30%) and posterior cortex (35%). In 23 fractures (74%), callus formation extended to the lateral cortex under the plate.
Absence of hardware and fixation failure suggests that dynamic plating of distal femur fractures with FCL screws provides safe and effective fixation. Moreover, the amount and symmetric distribution of periosteal callus suggests that dynamic fixation with FCL screws may promote increased fracture healing over standard locked plating. However, this hypothesis on the stimulatory effect of dynamic fixation on fracture healing requires investigation in a future randomized control trial.
Type d'étude
Inscription (Anticipé)
Contacts et emplacements
Lieux d'étude
-
-
Oregon
-
Portland, Oregon, États-Unis, 97232
- Legacy Health System
-
-
Critères de participation
Critère d'éligibilité
Âges éligibles pour étudier
Accepte les volontaires sains
Sexes éligibles pour l'étude
Méthode d'échantillonnage
Population étudiée
La description
Inclusion Criteria:
- Patients with distal femur fracture (AO/OTA Type 33A and 33C)
- Patients 17 years of age and over.
- Patients able to be operated on by selected surgeons at the participating centers.
Exclusion Criteria:
- Pregnancy
- Patients who are enrolled in an investigational treatment trial.
- Patients who are not expected to survive the follow-up period.
- Considered an inappropriate participant by the study physician.
- Revision surgery
- Patients currently incarcerated or awaiting incarceration.
- Severe spinal injury with neurological deficit resulting in paralysis.
- Fracture fixed more then 28 days after injury.
- Acute or chronic local or systemic infections
- Periprosthetic fractures
Plan d'étude
Comment l'étude est-elle conçue ?
Détails de conception
Cohortes et interventions
Groupe / Cohorte |
|---|
|
Femur Fractures
Patients with Distal Femur Fractures
|
Que mesure l'étude ?
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Guérison des fractures
Délai: 6, 12 et 24 semaines après la chirurgie
|
La cicatrisation des fractures est définie cliniquement par la capacité de mise en charge sans douleur, et radiographiquement par la formation de callosités et le pontage.
|
6, 12 et 24 semaines après la chirurgie
|
Mesures de résultats secondaires
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Taille du cal périosté
Délai: 6, 12, 24 semaines après la chirurgie
|
La taille du cal périosté est évaluée aux aspects antérieur, postérieur et médial sur des radiographies à l'aide d'un algorithme de calcul validé.
|
6, 12, 24 semaines après la chirurgie
|
|
Pontage des callosités de la tomodensitométrie
Délai: 12 semaines après la chirurgie
|
Analyse d'image en coupe, appuyée par un rendu 3D, pour détecter le pontage osseux au niveau des aspects antérieur, postérieur et latéral du fémur.
|
12 semaines après la chirurgie
|
|
Échec de fixation
Délai: 24 semaines après la chirurgie
|
Évalué en termes de perte d'alignement à 12 et 24 semaines après la chirurgie
|
24 semaines après la chirurgie
|
|
Défaillance matérielle
Délai: 6, 12, 24 semaines après la chirurgie
|
Casse de vis ou de plaque
|
6, 12, 24 semaines après la chirurgie
|
Collaborateurs et enquêteurs
Parrainer
Les enquêteurs
- Chercheur principal: Michael Bottlang, PhD, Legacy Biomechanics Laboratory
- Chaise d'étude: Steven Madey, MD, Legacy Health System
Publications et liens utiles
Dates d'enregistrement des études
Dates principales de l'étude
Début de l'étude
Achèvement primaire (Réel)
Achèvement de l'étude (Réel)
Dates d'inscription aux études
Première soumission
Première soumission répondant aux critères de contrôle qualité
Première publication (Estimation)
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Estimation)
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière vérification
Plus d'information
Termes liés à cette étude
Mots clés
Termes MeSH pertinents supplémentaires
Autres numéros d'identification d'étude
- MotionLoc Screw
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .