Corneal Power, Astigmatism, and Aberration Changes After LASIK
4 avril 2018 mis à jour par: David Huang, Oregon Health and Science University
Observational Study of Corneal Power, Astigmatism, and Aberration Changes After LASIK
The primary goal of the study is to determine if Optical Coherence Tomography (OCT) measurement of corneal power and astigmatism is more accurate than conventional Placido-ring corneal topography.
A second goal is to evaluate OCT-based epithelial thickness mapping.
The postoperative epithelial thickness variation will be used to calculate refractive effects and calibrate the smoothing parameter in the mathematical model relating corneal mean curvature and epithelial thickness.
A third goal is to assess the repeatability of OCT measurements.
A fourth goal is to develop a method of estimating the focusing power, astigmatism, and HOA of the crystalline lens based on measurements of the lens' anterior and posterior capsular topographies and positions with the ultrahigh-speed MIT OCT prototype.
Aperçu de l'étude
Statut
Complété
Les conditions
Description détaillée
A prospective observational study on refraction and HOA changes after LASIK will be conducted at OHSU and Baylor.
We will enroll 50 subjects undergoing routine LASIK for the correction of myopia or hyperopia.
The Optovue anterior segment OCT prototype will be used to measure corneal power, astigmatism, HOA, and epithelial thickness map.
The ultrahigh-speed MIT OCT prototypes will also be used when they become available.
A comprehensive eye examination will be performed preoperatively.
OCT scan, manifest refraction, visual acuity, Placido-ring corneal topography, slit scanning/Scheimpflug camera, and wavefront measurements will be acquired preoperatively and 3-6 months after the surgery.
The LASIK-induced refraction and wavefront HOA changes will be used as the reference standard to evaluate the accuracy of OCT-measured corneal power, astigmatism, and HOA changes.
The primary goal of the study is to determine if OCT measurement of corneal power and astigmatism is more accurate than conventional Placido-ring corneal topography.
A second goal is to evaluate OCT-based epithelial thickness mapping.
The postoperative epithelial thickness variation will be used to calculate refractive effects and calibrate the smoothing parameter in the mathematical model relating corneal mean curvature and epithelial thickness.
A third goal is to assess the repeatability of OCT measurements.
A fourth goal is to develop a method of estimating the focusing power, astigmatism, and HOA of the crystalline lens based on measurements of the lens' anterior and posterior capsular topographies and positions with the ultrahigh-speed MIT OCT prototype.
The reference standard for lens focusing power and astigmatism will be based on a Gaussian optics model that takes as inputs manifest refraction, crystalline lens position, axial eye length, and OCT corneal power and astigmatism measurements.
The reference standard for lens HOA will be the difference between wavefront HOA and corneal HOA.
For each eye, the best-fit effective index of the lens will be found so that OCT measurements will match the reference data.
Although the lens index is likely to be higher centrally than peripherally, this secondary perturbation will be accounted for by a posterior shift of the posterior lens capsule in OCT images, due to the fact that apparent distances in OCT images is a product of physical distance and optical group index.
Thus HOA due to index variation in the lens would translate to unevenness in the posterior capsule surface that could be measured with OCT.
Type d'étude
Observationnel
Inscription (Réel)
22
Contacts et emplacements
Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.
Lieux d'étude
-
-
Oregon
-
Portland, Oregon, États-Unis, 97239
- Oregon Health & Science University
-
-
Critères de participation
Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.
Critère d'éligibilité
Âges éligibles pour étudier
18 ans et plus (Adulte, Adulte plus âgé)
Accepte les volontaires sains
Non
Sexes éligibles pour l'étude
Tout
Méthode d'échantillonnage
Échantillon non probabiliste
Population étudiée
Subjects undergoing routine LASIK for the correction of myopia or hyperopia with or without astigmatism
La description
Inclusion Criteria:
1. Patients who qualify for primary LASIK for either myopia or hyperopia, with or without astigmatism.
Exclusion Criteria:
- Inability to give informed consent.
- Inability to maintain stable fixation for OCT imaging.
- Inability to commit to required visits to complete the study.
- Eyes with concurrent cataract, retinal diseases, glaucoma, or other eye conditions that may limit the visual outcome after surgery
Plan d'étude
Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.
Comment l'étude est-elle conçue ?
Détails de conception
- Modèles d'observation: Cas uniquement
- Perspectives temporelles: Éventuel
Cohortes et interventions
Groupe / Cohorte |
---|
LASIK
Subjects undergoing routine LASIK for the correction of myopia or hyperopia.
|
Que mesure l'étude ?
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
---|---|---|
OCT measurement of corneal power and astigmatism vs Placido-ring
Délai: 5 year
|
primary goal of the study is to determine if OCT measurement of corneal power and astigmatism is more accurate than conventional Placido-ring corneal topography
|
5 year
|
Mesures de résultats secondaires
Mesure des résultats |
Description de la mesure |
Délai |
---|---|---|
An evaluation of OCT-based epithelial thickness mapping
Délai: 5 year
|
The postoperative epithelial thickness variation will be used to calculate refractive effects and calibrate the smoothing parameter in the mathematical model relating corneal mean curvature and epithelial thickness.
|
5 year
|
OCT measurements repeatability
Délai: 5 year
|
An assessment of the repeatability of OCT measurements
|
5 year
|
The Development of a method of estimating the focusing power, astigmatism, and HOA of the crystalline lens
Délai: 5 year
|
The Development of a method of estimating the focusing power, astigmatism, and HOA of the crystalline lens based on measurements of the lens' anterior and posterior capsular topographies and positions with the ultrahigh-speed MIT OCT prototype.
|
5 year
|
Collaborateurs et enquêteurs
C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.
Parrainer
Collaborateurs
Les enquêteurs
- Chercheur principal: David Huang, MD, PhD, Oregon Health & Science Universtiy
Dates d'enregistrement des études
Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.
Dates principales de l'étude
Début de l'étude
1 janvier 2011
Achèvement primaire (Réel)
31 août 2017
Achèvement de l'étude (Réel)
31 août 2017
Dates d'inscription aux études
Première soumission
2 juillet 2013
Première soumission répondant aux critères de contrôle qualité
2 juillet 2013
Première publication (Estimation)
9 juillet 2013
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Réel)
5 avril 2018
Dernière mise à jour soumise répondant aux critères de contrôle qualité
4 avril 2018
Dernière vérification
1 avril 2018
Plus d'information
Termes liés à cette étude
Termes MeSH pertinents supplémentaires
Autres numéros d'identification d'étude
- OHSU IRB#6612
- R01EY018184 (Subvention/contrat des NIH des États-Unis)
Informations sur les médicaments et les dispositifs, documents d'étude
Étudie un produit pharmaceutique réglementé par la FDA américaine
Non
Étudie un produit d'appareil réglementé par la FDA américaine
Non
produit fabriqué et exporté des États-Unis.
Non
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .