Corneal Power, Astigmatism, and Aberration Changes After LASIK

4 de abril de 2018 actualizado por: David Huang, Oregon Health and Science University

Observational Study of Corneal Power, Astigmatism, and Aberration Changes After LASIK

The primary goal of the study is to determine if Optical Coherence Tomography (OCT) measurement of corneal power and astigmatism is more accurate than conventional Placido-ring corneal topography. A second goal is to evaluate OCT-based epithelial thickness mapping. The postoperative epithelial thickness variation will be used to calculate refractive effects and calibrate the smoothing parameter in the mathematical model relating corneal mean curvature and epithelial thickness. A third goal is to assess the repeatability of OCT measurements. A fourth goal is to develop a method of estimating the focusing power, astigmatism, and HOA of the crystalline lens based on measurements of the lens' anterior and posterior capsular topographies and positions with the ultrahigh-speed MIT OCT prototype.

Descripción general del estudio

Descripción detallada

A prospective observational study on refraction and HOA changes after LASIK will be conducted at OHSU and Baylor. We will enroll 50 subjects undergoing routine LASIK for the correction of myopia or hyperopia. The Optovue anterior segment OCT prototype will be used to measure corneal power, astigmatism, HOA, and epithelial thickness map. The ultrahigh-speed MIT OCT prototypes will also be used when they become available. A comprehensive eye examination will be performed preoperatively. OCT scan, manifest refraction, visual acuity, Placido-ring corneal topography, slit scanning/Scheimpflug camera, and wavefront measurements will be acquired preoperatively and 3-6 months after the surgery. The LASIK-induced refraction and wavefront HOA changes will be used as the reference standard to evaluate the accuracy of OCT-measured corneal power, astigmatism, and HOA changes. The primary goal of the study is to determine if OCT measurement of corneal power and astigmatism is more accurate than conventional Placido-ring corneal topography. A second goal is to evaluate OCT-based epithelial thickness mapping. The postoperative epithelial thickness variation will be used to calculate refractive effects and calibrate the smoothing parameter in the mathematical model relating corneal mean curvature and epithelial thickness. A third goal is to assess the repeatability of OCT measurements. A fourth goal is to develop a method of estimating the focusing power, astigmatism, and HOA of the crystalline lens based on measurements of the lens' anterior and posterior capsular topographies and positions with the ultrahigh-speed MIT OCT prototype. The reference standard for lens focusing power and astigmatism will be based on a Gaussian optics model that takes as inputs manifest refraction, crystalline lens position, axial eye length, and OCT corneal power and astigmatism measurements. The reference standard for lens HOA will be the difference between wavefront HOA and corneal HOA. For each eye, the best-fit effective index of the lens will be found so that OCT measurements will match the reference data. Although the lens index is likely to be higher centrally than peripherally, this secondary perturbation will be accounted for by a posterior shift of the posterior lens capsule in OCT images, due to the fact that apparent distances in OCT images is a product of physical distance and optical group index. Thus HOA due to index variation in the lens would translate to unevenness in the posterior capsule surface that could be measured with OCT.

Tipo de estudio

De observación

Inscripción (Actual)

22

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Ubicaciones de estudio

    • Oregon
      • Portland, Oregon, Estados Unidos, 97239
        • Oregon Health & Science University

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

18 años y mayores (Adulto, Adulto Mayor)

Acepta Voluntarios Saludables

No

Géneros elegibles para el estudio

Todos

Método de muestreo

Muestra no probabilística

Población de estudio

Subjects undergoing routine LASIK for the correction of myopia or hyperopia with or without astigmatism

Descripción

Inclusion Criteria:

1. Patients who qualify for primary LASIK for either myopia or hyperopia, with or without astigmatism.

Exclusion Criteria:

  1. Inability to give informed consent.
  2. Inability to maintain stable fixation for OCT imaging.
  3. Inability to commit to required visits to complete the study.
  4. Eyes with concurrent cataract, retinal diseases, glaucoma, or other eye conditions that may limit the visual outcome after surgery

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Modelos observacionales: Solo caso
  • Perspectivas temporales: Futuro

Cohortes e Intervenciones

Grupo / Cohorte
LASIK
Subjects undergoing routine LASIK for the correction of myopia or hyperopia.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
OCT measurement of corneal power and astigmatism vs Placido-ring
Periodo de tiempo: 5 year
primary goal of the study is to determine if OCT measurement of corneal power and astigmatism is more accurate than conventional Placido-ring corneal topography
5 year

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
An evaluation of OCT-based epithelial thickness mapping
Periodo de tiempo: 5 year
The postoperative epithelial thickness variation will be used to calculate refractive effects and calibrate the smoothing parameter in the mathematical model relating corneal mean curvature and epithelial thickness.
5 year
OCT measurements repeatability
Periodo de tiempo: 5 year
An assessment of the repeatability of OCT measurements
5 year
The Development of a method of estimating the focusing power, astigmatism, and HOA of the crystalline lens
Periodo de tiempo: 5 year
The Development of a method of estimating the focusing power, astigmatism, and HOA of the crystalline lens based on measurements of the lens' anterior and posterior capsular topographies and positions with the ultrahigh-speed MIT OCT prototype.
5 year

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Investigadores

  • Investigador principal: David Huang, MD, PhD, Oregon Health & Science Universtiy

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio

1 de enero de 2011

Finalización primaria (Actual)

31 de agosto de 2017

Finalización del estudio (Actual)

31 de agosto de 2017

Fechas de registro del estudio

Enviado por primera vez

2 de julio de 2013

Primero enviado que cumplió con los criterios de control de calidad

2 de julio de 2013

Publicado por primera vez (Estimar)

9 de julio de 2013

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

5 de abril de 2018

Última actualización enviada que cumplió con los criterios de control de calidad

4 de abril de 2018

Última verificación

1 de abril de 2018

Más información

Términos relacionados con este estudio

Términos MeSH relevantes adicionales

Otros números de identificación del estudio

  • OHSU IRB#6612
  • R01EY018184 (Subvención/contrato del NIH de EE. UU.)

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

producto fabricado y exportado desde los EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .

3
Suscribir