Non-inflatable Supraglottic Airway (NI-SGA) vs. Face Mask (FM) as a Primary Interface Device for Neonatal Resuscitation
Non-inflatable Supraglottic Airway (NI-SGA) vs. Face Mask (FM) as a Primary Interface Device for Neonatal Resuscitation: A Pilot Study
Panoramica dello studio
Stato
Stato
Condizioni
Condizioni
Intervento / Trattamento
Intervento / Trattamento
Descrizione dettagliata
Tipo di studio
Tipo di studio
Fase
Fase
- Non applicabile
Contatti e Sedi
Luoghi di studio
-
-
Oklahoma
-
Oklahoma City, Oklahoma, Stati Uniti, 73117
- University of Oklahoma Health Sciences Center
-
-
Criteri di partecipazione
Criteri di ammissibilità
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Descrizione
Inclusion Criteria:
- Newborns > or = 2000gms requiring positive pressure ventilation within the first 2 minutes of life
Exclusion Criteria:
- Newborns with expected fetal weight <2000 grams
- Known congenital malformations
- Multiple gestation (twins and above)
- History of meconium stained fluid
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Scopo principale: Trattamento
- Assegnazione: Randomizzato
- Modello interventistico: Assegnazione parallela
- Mascheramento: Nessuno (etichetta aperta)
Numero di armi
Armi e interventi
Gruppo di partecipanti / ArmGruppo di partecipanti / Arm |
Intervento / TrattamentoIntervento / Trattamento |
|---|---|
|
Comparatore attivo: Face Mask
Standard Face Mask with T piece resuscitator for neonatal resuscitation.
Face mask placement per Neonatal Resuscitation Program resuscitation guideline.
|
at delivery - standard airway management per Neonatal Resuscitation Program utilizing face mask.
T-Piece Resuscitator for providing postivie pressure ventilation via face mask or non-inflatable supraglottic airway
|
|
Comparatore attivo: Non-inflatable supraglottic airway
Use of non-inflating supraglottic airway with T-piece resuscitator instead of Standard Face Mask with T piece resuscitator for neonatal resuscitation, replacing standard of care face mask in Neonatal Resuscitation Program guideline.
|
T-Piece Resuscitator for providing postivie pressure ventilation via face mask or non-inflatable supraglottic airway
at delivery if resuscitation is required infant randomized to NI-SGA or face mask as the interface for positive pressure ventilation.
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Time to Spontaneous Breathing
Lasso di tempo: 30 minutes
|
Record of time from beginning of resuscitation to time of spontaneous breathing
|
30 minutes
|
|
Duration of Positive Pressure Ventilation
Lasso di tempo: 30 minutes
|
Total time of positive pressure ventilation required to adequate spontaneous breathing.
|
30 minutes
|
|
Need for subsequent endotracheal intubation
Lasso di tempo: 30 minutes
|
Failure of non-invasive resuscitative efforts by face mask or NI-SGA requiring intubation per NRP guideline.
|
30 minutes
|
Misure di risultato secondarie
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Time needed for NI-SGA placement
Lasso di tempo: 1 minute
|
Time needed adequate placement of non-inflatable supraglottic airway
|
1 minute
|
|
Need for drugs in resuscitation
Lasso di tempo: 30 minutes
|
Drug administration as part of the standard neonatal resuscitation per NRP
|
30 minutes
|
|
Use of Chest Compressions
Lasso di tempo: 30 minutes
|
Need for chest compressions as part of neonatal resuscitation per NRP guidelines
|
30 minutes
|
|
Use of oxygen
Lasso di tempo: 30 minutes
|
Amount and duration of oxygen use in resuscitation
|
30 minutes
|
|
Rate of complications associated with resuscitation
Lasso di tempo: 30 minutes
|
Presence of complications including bleeding, abdominal distension, laryngeal stridor
|
30 minutes
|
|
Heart rate 2 minutes after initiation of resuscitation
Lasso di tempo: 2 minutes
|
Heart rate assessed at 2 minutes
|
2 minutes
|
|
Maximum inspiratory pressure
Lasso di tempo: 30 minutes
|
Record of highest peak inspiratory pressure used in resuscitation
|
30 minutes
|
Collaboratori e investigatori
Sponsor
Sponsor
Investigatori
Investigatori
- Investigatore principale: Arlen Foulks, DO, University of Oklahoma
Pubblicazioni e link utili
Pubblicazioni generali
- Liu L, Oza S, Hogan D, Perin J, Rudan I, Lawn JE, Cousens S, Mathers C, Black RE. Global, regional, and national causes of child mortality in 2000-13, with projections to inform post-2015 priorities: an updated systematic analysis. Lancet. 2015 Jan 31;385(9966):430-40. doi: 10.1016/S0140-6736(14)61698-6. Epub 2014 Sep 30. Erratum In: Lancet. 2015 Jan 31;385(9966):420. Lancet. 2016 Jun 18;387(10037):2506.
- Kattwinkel J, Perlman JM, Aziz K, Colby C, Fairchild K, Gallagher J, Hazinski MF, Halamek LP, Kumar P, Little G, McGowan JE, Nightengale B, Ramirez MM, Ringer S, Simon WM, Weiner GM, Wyckoff M, Zaichkin J. Part 15: neonatal resuscitation: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010 Nov 2;122(18 Suppl 3):S909-19. doi: 10.1161/CIRCULATIONAHA.110.971119. No abstract available. Erratum In: Circulation. 2011 Oct 11;124(15):e406.
- Ersdal HL, Mduma E, Svensen E, Perlman JM. Early initiation of basic resuscitation interventions including face mask ventilation may reduce birth asphyxia related mortality in low-income countries: a prospective descriptive observational study. Resuscitation. 2012 Jul;83(7):869-73. doi: 10.1016/j.resuscitation.2011.12.011. Epub 2011 Dec 23.
- Zhu XY, Lin BC, Zhang QS, Ye HM, Yu RJ. A prospective evaluation of the efficacy of the laryngeal mask airway during neonatal resuscitation. Resuscitation. 2011 Nov;82(11):1405-9. doi: 10.1016/j.resuscitation.2011.06.010. Epub 2011 Jul 16.
- Schmolzer GM, Kamlin OC, O'Donnell CP, Dawson JA, Morley CJ, Davis PG. Assessment of tidal volume and gas leak during mask ventilation of preterm infants in the delivery room. Arch Dis Child Fetal Neonatal Ed. 2010 Nov;95(6):F393-7. doi: 10.1136/adc.2009.174003. Epub 2010 Jun 14.
- Bernhard M, Benger JR. Airway management during cardiopulmonary resuscitation. Curr Opin Crit Care. 2015 Jun;21(3):183-7. doi: 10.1097/MCC.0000000000000201.
- Ersdal HL, Singhal N. Resuscitation in resource-limited settings. Semin Fetal Neonatal Med. 2013 Dec;18(6):373-8. doi: 10.1016/j.siny.2013.07.001. Epub 2013 Jul 27.
- Wyckoff MH, Perlman JM. Effective ventilation and temperature control are vital to outborn resuscitation. Prehosp Emerg Care. 2004 Apr-Jun;8(2):191-5. doi: 10.1016/j.prehos.2003.12.013.
- Wyckoff MH, Perlman JM, Laptook AR. Use of volume expansion during delivery room resuscitation in near-term and term infants. Pediatrics. 2005 Apr;115(4):950-5. doi: 10.1542/peds.2004-0913.
- Szyld E, Aguilar A, Musante GA, Vain N, Prudent L, Fabres J, Carlo WA; Delivery Room Ventilation Devices Trial Group. Comparison of devices for newborn ventilation in the delivery room. J Pediatr. 2014 Aug;165(2):234-239.e3. doi: 10.1016/j.jpeds.2014.02.035. Epub 2014 Mar 29.
- Kattwinkel J, Perlman JM, Aziz K, Colby C, Fairchild K, Gallagher J, Hazinski MF, Halamek LP, Kumar P, Little G, McGowan JE, Nightengale B, Ramirez MM, Ringer S, Simon WM, Weiner GM, Wyckoff M, Zaichkin J; American Heart Association. Neonatal resuscitation: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Pediatrics. 2010 Nov;126(5):e1400-13. doi: 10.1542/peds.2010-2972E. Epub 2010 Oct 18. No abstract available. Erratum In: Pediatrics. 2011 Jul;128(1):176.
- O'Donnell CP, Davis PG, Morley CJ. Positive pressure ventilation at neonatal resuscitation: review of equipment and international survey of practice. Acta Paediatr. 2004 May;93(5):583-8. doi: 10.1111/j.1651-2227.2004.tb02981.x.
- Hartung JC, te Pas AB, Fischer H, Schmalisch G, Roehr CC. Leak during manual neonatal ventilation and its effect on the delivered pressures and volumes: an in vitro study. Neonatology. 2012;102(3):190-5. doi: 10.1159/000339325. Epub 2012 Jul 10.
- Deindl P, O'Reilly M, Zoller K, Berger A, Pollak A, Schwindt J, Schmolzer GM. Influence of mask type and mask position on the effectiveness of bag-mask ventilation in a neonatal manikin. Eur J Pediatr. 2014 Jan;173(1):75-9. doi: 10.1007/s00431-013-2122-4. Epub 2013 Aug 11.
- Schilleman K, van der Pot CJ, Hooper SB, Lopriore E, Walther FJ, te Pas AB. Evaluating manual inflations and breathing during mask ventilation in preterm infants at birth. J Pediatr. 2013 Mar;162(3):457-63. doi: 10.1016/j.jpeds.2012.09.036. Epub 2012 Oct 25.
- Foglia EE, Owen LS, Thio M, Ratcliffe SJ, Lista G, Te Pas A, Hummler H, Nadkarni V, Ades A, Posencheg M, Keszler M, Davis P, Kirpalani H. Sustained Aeration of Infant Lungs (SAIL) trial: study protocol for a randomized controlled trial. Trials. 2015 Mar 15;16:95. doi: 10.1186/s13063-015-0601-9.
- Richmond S, Wyllie J. European Resuscitation Council Guidelines for Resuscitation 2010 Section 7. Resuscitation of babies at birth. Resuscitation. 2010 Oct;81(10):1389-99. doi: 10.1016/j.resuscitation.2010.08.018. No abstract available.
- Schmolzer GM, Agarwal M, Kamlin CO, Davis PG. Supraglottic airway devices during neonatal resuscitation: an historical perspective, systematic review and meta-analysis of available clinical trials. Resuscitation. 2013 Jun;84(6):722-30. doi: 10.1016/j.resuscitation.2012.11.002. Epub 2012 Nov 9.
- Leone TA, Rich W, Finer NN. A survey of delivery room resuscitation practices in the United States. Pediatrics. 2006 Feb;117(2):e164-75. doi: 10.1542/peds.2005-0936.
- Schreiner MS, Feltman D, Wiswell T, Wootton S, Arnold C, Tyson J, Lantos JD. When is waiver of consent appropriate in a neonatal clinical trial? Pediatrics. 2014 Nov;134(5):1006-12. doi: 10.1542/peds.2014-0207. Epub 2014 Oct 6.
- Nichol G, Huszti E. Design and implementation of resuscitation research: special challenges and potential solutions. Resuscitation. 2007 Jun;73(3):337-46. doi: 10.1016/j.resuscitation.2006.10.021. Epub 2007 Feb 9.
- Paterson SJ, Byrne PJ, Molesky MG, Seal RF, Finucane BT. Neonatal resuscitation using the laryngeal mask airway. Anesthesiology. 1994 Jun;80(6):1248-53; discussion 27A.
- Gandini D, Brimacombe JR. Neonatal resuscitation with the laryngeal mask airway in normal and low birth weight infants. Anesth Analg. 1999 Sep;89(3):642-3. doi: 10.1097/00000539-199909000-00018. No abstract available.
- Zanardo V, Weiner G, Micaglio M, Doglioni N, Buzzacchero R, Trevisanuto D. Delivery room resuscitation of near-term infants: role of the laryngeal mask airway. Resuscitation. 2010 Mar;81(3):327-30. doi: 10.1016/j.resuscitation.2009.11.005. Epub 2009 Dec 22.
- Trevisanuto D, Cavallin F, Nguyen LN, Nguyen TV, Tran LD, Tran CD, Doglioni N, Micaglio M, Moccia L. Supreme Laryngeal Mask Airway versus Face Mask during Neonatal Resuscitation: A Randomized Controlled Trial. J Pediatr. 2015 Aug;167(2):286-91.e1. doi: 10.1016/j.jpeds.2015.04.051. Epub 2015 May 21.
Studiare le date dei record
Studia le date principali
Inizio studio (Anticipato)
Inizio studio
Completamento primario (Anticipato)
Completamento primario
Completamento dello studio (Anticipato)
Completamento dello studio
Date di iscrizione allo studio
Primo inviato
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Stima)
Primo Inserito
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento pubblicato
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Altri numeri di identificazione dello studio
Altri numeri di identificazione dello studio
- 5848
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
prodotto fabbricato ed esportato dagli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .