UNITY DX Cornea Study
UNITY DX Cornea Precision and Agreement Study
Panoramica dello studio
Stato
Stato
Condizioni
Condizioni
Intervento / Trattamento
Intervento / Trattamento
Descrizione dettagliata
Tipo di studio
Tipo di studio
Iscrizione (Stimato)
Iscrizione
Contatti e Sedi
Contatto studio
Contatto studio
- Nome: Alcon Call Center
- Numero di telefono: 1-888-451-3937
- Email: alcon.medinfo@alcon.com
Luoghi di studio
-
-
Utah
-
Salt Lake City, Utah, Stati Uniti, 84107
- The Eye Institute of Utah
-
-
Criteri di partecipazione
Criteri di ammissibilità
Criteri di ammissibilità
Età idonea allo studio
- Adulto
- Adulto più anziano
Accetta volontari sani
Metodo di campionamento
Popolazione di studio
Descrizione
Key Inclusion Criteria:
- Able to understand and sign an Institutional Review Board approved informed consent form.
- Willing and able to attend all scheduled study visits as required per protocol.
- Identified by the investigator as having normal corneas or abnormal corneas per protocol-specified criteria.
- Other protocol-specified inclusion criteria may apply.
Key Exclusion Criteria:
- Unable to fixate due to nystagmus or other eye movement abnormality, such as significant strabismus.
- Any ocular disease and/or condition that, in the investigator's clinical judgement, may put the subject at significant risk, may compromise study results, or may interfere significantly with the subject's participation in the study.
- Active ocular infection or inflammation.
- Other protocol-specified exclusion criteria may apply.
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
Numero di gruppi/coorti
Coorti e interventi
Gruppo / CoorteGruppo / Coorte |
Intervento / TrattamentoIntervento / Trattamento |
|---|---|
|
Normal Corneas
The eye will be imaged up to seven times with each biometer to achieve three acceptable scans.
This cohort will enroll subjects with normal corneas.
|
Investigational non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and their optical properties
Commercially available non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and their optical properties
|
|
Keratoconus
The eye will be imaged up to seven times with each biometer to achieve three acceptable scans.
This cohort will enroll subjects with keratoconus.
|
Investigational non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and their optical properties
Commercially available non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and their optical properties
|
|
Prior Corneal Laser Vision Correction
The eye will be imaged up to seven times with each biometer to achieve three acceptable scans.
This cohort will enroll subjects with prior corneal laser vision correction.
|
Investigational non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and their optical properties
Commercially available non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and their optical properties
|
|
Current and Habitual Contact Lens Wear
The eye will be imaged up to seven times with each biometer to achieve three acceptable scans.
This cohort will enroll subjects with current and habitual contact lens wear.
|
Investigational non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and their optical properties
Commercially available non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and their optical properties
|
|
Dry Eye
The eye will be imaged up to seven times with each biometer to achieve three acceptable scans.
This cohort will enroll subjects with dry eye.
|
Investigational non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and their optical properties
Commercially available non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and their optical properties
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Mean corneal thickness
Lasso di tempo: Day 1
|
Corneal thickness will be measured by the device and reported in microns.
This outcome measure will be analyzed using Bland-Altman Limits of Agreement, Deming regression, repeatability and reproducibility.
|
Day 1
|
|
Mean epithelial thickness
Lasso di tempo: Day 1
|
Epithelial thickness will be measured by the device and reported in microns.
This outcome measure will be analyzed using Bland-Altman Limits of Agreement, Deming regression, repeatability and reproducibility.
|
Day 1
|
Collaboratori e investigatori
Sponsor
Sponsor
Investigatori
Investigatori
- Direttore dello studio: Clinical Trial Lead, Surgical, Alcon Research, LLC
Studiare le date dei record
Studia le date principali
Inizio studio (Stimato)
Inizio studio
Completamento primario (Stimato)
Completamento primario
Completamento dello studio (Stimato)
Completamento dello studio
Date di iscrizione allo studio
Primo inviato
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Primo Inserito
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento pubblicato
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
Altri numeri di identificazione dello studio
- CTK378-C003
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .