Clinical Validation of SMD-RVECG for Predicting Atrial Fibrillation With Rapid Ventricular Response Within 2 Hours (AFRVR)
A Retrospective, Single Center, Single Arm, Superiority Confirmatory Study to Evaluate the Clinical Validity of SMD-RVECG, Artificial Intelligence Software to Predict the Risk of Developing Atrial Fibrillation With Rapid Ventricular Response (AF With RVR) Within 2 Hours Using Single Lead-Based Electrocardiogram Analysis
The purpose of this retrospective study is to evaluate the clinical performance of SMD-RVECG, an artificial intelligence-based medical device software that analyzes single-lead electrocardiogram data to predict the risk of atrial fibrillation with rapid ventricular response occurring within 2 hours.
A total of 348 eligible electrocardiogram datasets collected through VitalDB at Seoul National University Hospital will be included. Atrial fibrillation with rapid ventricular response is defined in this study as atrial fibrillation accompanied by an average heart rate of 110 beats per minute or greater for at least 30 seconds.
Eligible datasets will be classified as positive or negative for atrial fibrillation with rapid ventricular response. Two qualified physicians blinded to the software results will review the electrocardiogram data and relevant medical records to establish the reference-standard classification. The blinded electrocardiogram datasets will then be analyzed using SMD-RVECG, and the software-generated predictions will be compared with the reference standard to evaluate clinical performance.
Panoramica dello studio
Stato
Intervento / Trattamento
Descrizione dettagliata
This clinical study is designed as a retrospective, single-center, single-arm, superiority confirmatory clinical performance study conducted at Seoul National University Hospital.
Pre-screening Electronic medical records and the VitalDB registry will be reviewed to identify potentially eligible data. The study will use continuous single-lead electrocardiogram data collected through VitalDB between September 1, 2022, and September 30, 2025, from patients who were 19 years of age or older at the time of data collection.
Screening and Data Collection A screening number will be assigned sequentially to each potentially eligible dataset. Eligibility will be determined by reviewing the predefined inclusion and exclusion criteria. For eligible cases, continuous single-lead electrocardiogram data, signal loss, heart rate, the occurrence and timing of atrial fibrillation with rapid ventricular response episodes, the electrocardiogram data collection period, demographic information, previous history of atrial fibrillation, and pregnancy status will be collected. Directly identifying personal information will not be collected.
Test Dataset Creation Eligible electrocardiogram datasets will be classified into positive and negative groups for atrial fibrillation with rapid ventricular response.
The positive group will include electrocardiogram datasets in which atrial fibrillation is followed by rapid ventricular response, defined as an average heart rate of 110 beats per minute or greater for at least 30 seconds. The dataset will include electrocardiogram data preceding the onset of rapid ventricular response and data following the episode, as specified in the study protocol.
The negative group will include electrocardiogram datasets with atrial fibrillation that is not accompanied by rapid ventricular response during the predefined assessment period.
The planned study dataset consists of 348 cases, including 58 positive cases and 290 negative cases.
Reference Standard Establishment Two anesthesiology and pain medicine specialists with at least 5 years of relevant clinical experience in electrocardiogram-based arrhythmia interpretation will independently review the electrocardiogram data and relevant medical records. The reviewers will assess electrocardiogram data quality, confirm the presence or absence and timing of atrial fibrillation with rapid ventricular response, and establish the reference-standard classification.
The reviewers will be blinded to each other's initial assessments and to the results generated by SMD-RVECG. If the two reviewers disagree, the classification will be determined through consensus and the basis for the consensus will be documented. If consensus cannot be reached, the dataset will be excluded from the study.
Application of the Investigational Device The medical device operator will receive the test electrocardiogram datasets identified only by screening numbers and will be blinded to the positive or negative reference-standard classifications. The datasets will be retrospectively analyzed using SMD-RVECG to generate risk estimates for atrial fibrillation with rapid ventricular response occurring within 2 hours. The software-generated peak risk values and corresponding times will be recorded.
Statistical Analysis The software-generated predictions will be compared with the reference-standard classifications. The primary performance measure is the area under the receiver operating characteristic curve for predicting atrial fibrillation with rapid ventricular response within 2 hours. A two-sided 95% confidence interval will be calculated using Newcombe's Wald method. The study will be considered successful if the lower bound of the 95% confidence interval is greater than 0.8.
Secondary performance measures include the area under the precision-recall curve, threshold-specific precision, recall, F1 score, sensitivity, specificity, positive predictive value, negative predictive value, confusion-matrix results, and prediction time horizon.
Because this retrospective study uses previously collected data, there will be no direct participant contact, additional examination, treatment, or change in clinical care.
Tipo di studio
Iscrizione (Stimato)
Contatti e Sedi
Contatto studio
- Nome: JooYeon Yoo
- Numero di telefono: +82-2-2051-3161
- Email: clinical_trial@huinno.com
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
- Adulto
- Adulto più anziano
Accetta volontari sani
Metodo di campionamento
Popolazione di studio
Descrizione
Inclusion Criteria:
- Continuous single-lead electrocardiogram data collected through VitalDB at Seoul National University Hospital between September 1, 2022, and September 30, 2025
- Participant aged 19 years or older at the time of electrocardiogram data collection
Electrocardiogram data meeting either of the following group definitions:
- AF with RVR-positive group: atrial fibrillation followed by rapid ventricular response, defined as an average heart rate of 110 beats per minute or greater for at least 30 seconds, with available electrocardiogram data for up to 2 hours before the onset of rapid ventricular response and after the onset in accordance with the protocol
- AF with RVR-negative group: atrial fibrillation without rapid ventricular response, with 3 hours of electrocardiogram data available before a selected index time corresponding to a similar time during hospitalization as that used for positive cases
- If the relevant episode occurred within 2 hours after the start of electrocardiogram recording, availability of electrocardiogram data from the start of recording through up to 1 hour after the episode, with a minimum electrocardiogram recording duration of 10 minutes
- Other data considered suitable for the study by the principal investigator
Exclusion Criteria:
- Electrocardiogram data considered to be of inadequate quality because of noise, poor equipment handling, unclear signals, artifacts, or similar problems
- Electrocardiogram data with signal loss of 5 percent or greater
- Electrocardiogram data with a heart rate below 30 beats per minute or above 300 beats per minute
- Data collected from a participant who was pregnant or breastfeeding at the time of data collection
- Data previously used for training or validation of the artificial intelligence model
- Electrocardiogram data from a participant whose baseline heart rate remained at 110 beats per minute or greater irrespective of atrial fibrillation
- Data considered unsuitable for this clinical study by the principal investigator
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
Coorti e interventi
Gruppo / Coorte |
Intervento / Trattamento |
|---|---|
|
AF With RVR Positive Group
Electrocardiogram datasets from patients with atrial fibrillation in which rapid ventricular response occurred.
Rapid ventricular response is defined as an average heart rate of 110 beats per minute or greater for at least 30 seconds.
The dataset includes electrocardiogram data preceding the first onset of rapid ventricular response and data following the episode in accordance with the protocol.
These datasets will be retrospectively analyzed using SMD-RVECG.
|
SMD-RVECG is an artificial intelligence-based software as a medical device designed to analyze continuous single-lead electrocardiogram data and estimate the risk of atrial fibrillation with rapid ventricular response occurring within 2 hours. The software generates a risk score ranging from 0 to 100 based on electrocardiogram signal characteristics. In this retrospective study, blinded electrocardiogram datasets identified only by screening numbers will be analyzed using SMD-RVECG. The software-generated risk scores, peak values, and corresponding times will be recorded and compared with the reference-standard classifications. The software will not be used to guide the clinical care of the patients whose data are included. |
|
AF With RVR Negative Group
Electrocardiogram datasets from patients with atrial fibrillation in which rapid ventricular response did not occur during the predefined assessment period.
The dataset includes electrocardiogram data from the 3 hours preceding a selected index time corresponding to a similar time during hospitalization as the onset time used for positive cases.
These datasets will be retrospectively analyzed using SMD-RVECG.
|
SMD-RVECG is an artificial intelligence-based software as a medical device designed to analyze continuous single-lead electrocardiogram data and estimate the risk of atrial fibrillation with rapid ventricular response occurring within 2 hours. The software generates a risk score ranging from 0 to 100 based on electrocardiogram signal characteristics. In this retrospective study, blinded electrocardiogram datasets identified only by screening numbers will be analyzed using SMD-RVECG. The software-generated risk scores, peak values, and corresponding times will be recorded and compared with the reference-standard classifications. The software will not be used to guide the clinical care of the patients whose data are included. |
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
AUROC of SMD-RVECG for Predicting Atrial Fibrillation With Rapid Ventricular Response Within 2 Hours
Lasso di tempo: During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
The area under the receiver operating characteristic curve (AUROC) will be calculated by comparing the maximum risk score generated by SMD-RVECG with the reference-standard classification of the AF with RVR-positive and AF with RVR-negative electrocardiogram datasets.
Sensitivity and the false-positive rate will be calculated across the evaluated decision thresholds to construct the receiver operating characteristic curve.
A two-sided 95% confidence interval will be estimated using Newcombe's Wald method.
The performance criterion will be met if the lower bound of the 95% confidence interval is greater than 0.8.
|
During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Area Under the Precision-Recall Curve of SMD-RVECG
Lasso di tempo: During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
The area under the precision-recall curve (AUPRC) will be calculated using the risk scores generated by SMD-RVECG and the reference-standard classification of the AF with RVR-positive and AF with RVR-negative electrocardiogram datasets.
A higher AUPRC indicates better discrimination of AF with RVR-positive datasets.
|
During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
|
Sensitivity of SMD-RVECG for Predicting Atrial Fibrillation With Rapid Ventricular Response
Lasso di tempo: During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
Sensitivity, also referred to as recall, will be calculated at the evaluated decision thresholds as the proportion of reference-standard AF with RVR-positive datasets that are classified as positive by SMD-RVECG.
|
During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
|
Specificity of SMD-RVECG for Predicting Atrial Fibrillation With Rapid Ventricular Response
Lasso di tempo: During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
Specificity will be calculated at the evaluated decision thresholds as the proportion of reference-standard AF with RVR-negative datasets that are classified as negative by SMD-RVECG.
|
During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
|
Positive Predictive Value of SMD-RVECG
Lasso di tempo: During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
Positive predictive value, also referred to as precision, will be calculated at the evaluated decision thresholds as the proportion of datasets classified as positive by SMD-RVECG that are AF with RVR-positive according to the reference standard.
|
During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
|
Negative Predictive Value of SMD-RVECG
Lasso di tempo: During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
Negative predictive value will be calculated at the evaluated decision thresholds as the proportion of datasets classified as negative by SMD-RVECG that are AF with RVR-negative according to the reference standard.
|
During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
|
F1 Score of SMD-RVECG
Lasso di tempo: During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
The F1 score will be calculated at the evaluated decision thresholds as the harmonic mean of precision and recall.
The F1 score ranges from 0 to 1, with higher values indicating a better balance between precision and recall.
|
During retrospective analysis of the predefined electrocardiogram dataset for each case, up to 3 hours
|
|
Prediction Time Horizon of SMD-RVECG for Atrial Fibrillation With Rapid Ventricular Response
Lasso di tempo: Within the 2-hour period preceding the onset of atrial fibrillation with rapid ventricular response
|
For AF with RVR-positive datasets, the prediction time horizon will be calculated as the time interval between the software-generated prediction or alarm and the reference-standard onset of the AF with RVR episode.
|
Within the 2-hour period preceding the onset of atrial fibrillation with rapid ventricular response
|
Collaboratori e investigatori
Sponsor
Studiare le date dei record
Studia le date principali
Inizio studio (Stimato)
Completamento primario (Stimato)
Completamento dello studio (Stimato)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
- HU-AFRVR-01
- RS-2026-25543193 (Altro numero di sovvenzione/finanziamento: Korea Medical Device Development Foundation)
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Descrizione del piano IPD
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .
Prove cliniche su Fibrillazione atriale (FA)
-
Assiut UniversitySconosciutoAF non valvolare, AFEgitto
-
Oregon Health and Science UniversityRitirato
-
Sun Yat-Sen Memorial Hospital of Sun Yat-Sen UniversityThe First Affiliated Hospital of Nanchang University; Xiangya Hospital of Central... e altri collaboratoriIscrizione su invitoPazienti di ablazione AF con moderato+ FMR e AF persistenteCina
-
TriVirum, Inc.CompletatoFibrillazione atriale | Af | ECGStati Uniti
-
TriVirum, Inc.Completato
-
University Hospital DubravaAttivo, non reclutanteAblazione della fibrillazione atriale | AFCroazia
-
Pulse Biosciences, Inc.AvaniaReclutamentoFibrillazione atriale | Ablazione | Procedura del labirinto | AFStati Uniti
-
Vivek ReddyCompletatoAF persistente di lunga durataStati Uniti
-
TriVirum, Inc.CompletatoFibrillazione atriale | Aritmie, cardiache | Fibrillazione atriale parossistica | Af | Battito cardiaco irregolare | Aritmia atriale | Aritmie parossisticheStati Uniti
-
The Cleveland ClinicUniversity of WashingtonReclutamentoCPAP | Sindrome da apnea notturna, ostruttiva | AfStati Uniti
Prove cliniche su SMD-RVECG
-
Johns Hopkins UniversityNational Institute of Mental Health (NIMH)CompletatoGrave disregolazione dell'umoreStati Uniti
-
IVI BilbaoCompletatoAutovalutazioneSpagna
-
University of Illinois at ChicagoNon ancora reclutamentoEmorragia intracerebrale | ICH - Emorragia intracerebraleStati Uniti
-
Medical University of South CarolinaTerminatoCancro alla prostataStati Uniti
-
Ankara City Hospital BilkentCompletatoObesità | Anestesia | Intubazione; Difficile o fallitoTacchino