2 つの異なる酸素濃度計を使用した 2 つの酸素化ターゲットの比較 - 酸素流量と酸素化への影響 (Oxygap pong)
The investigators recently evaluated 4 different oximeters among the most commonly used with arterial catheter in place and compared SpO2 with SaO2 obtained on arterial gas. Correlations between SaO2 and SpO2 were poor for all oximeters, as previously known, and SpO2-SaO2 bias were different between oximeters. Some oximeters (Masimo, Nellcor) had lower biases but they detected less well hypoxemia. Some oximeters underestimated SaO2 (Nonin) but detected very well hypoxemia, and some overestimated SaO2 (Philips). The investigators concluded that oximeters provide different informations to clinicians, and oxygenation targets should take into account for these differences.
The assumption is that the SpO2 target AND oximeter used will both have an impact on oxygen flows and that these effects will add up. With a high SpO2 target, oxygen flows will be significantly greater and with the Nonin oximeter, the required flows will be greater than with the Philips oximeter.
NB: the results obtained were in a population with light skin pigmentation (96% of the patients were Fitzpatrick 1-2, reflecting the local hospitalized population).
調査の概要
状態
詳細な説明
The oximeters to be evaluated will be the Nonin (Plymouth, MN), and the Philips (Eindhoven, Netherlands). The sensors used will be digital reusable sensors for all oximeters. Because of the impact of the fingers used for SpO2 measurement, each oximeter will be randomly assigned a different finger to each patient (digit 2 and 3 on the hand opposite to the arterial catheter).
Two SpO2 targets will be assessed: 90% and 94%. A total of four 10-minute periods will be performed in randomized order: Nonin 90, Nonin 94, Philips 90, and Philips 94.
At the end of each 10-minute period, arterial gas will be collected through the arterial catheter.
The adjustment of the oxygen flow rates will be performed manually. The average oxygen flow of the last 2 minutes of the period will be considered.
研究の種類
入学 (推定)
段階
- 適用できない
連絡先と場所
研究連絡先
- 名前:Francois Lellouche
- 電話番号:3572 418656-8711
- メール:francois.lellouche@criucpq.ulaval.ca
研究連絡先のバックアップ
- 名前:Pierre Alexandre Bouchard
- 電話番号:2712 418656-8711
- メール:pierre-alexandre.bouchard@criucpq.ulaval.ca
研究場所
-
-
Quebec
-
Québec、Quebec、カナダ、G1V4G5
- 募集
- Institut Universitaire de Cardiologie et de Pneumologie de Québec
-
コンタクト:
- Francois Lellouche
- メール:francois.lellouche@criucpq.ulaval.ca
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- ≥ 18 years of age
- Patients admitted to postoperative cardiac surgery intensive care unit
- Presence of an arterial catheter Specific for the extubate patient
- Ongoing on conventional oxygen therapy at moderate flow (up to a maximum of 5L/min) with SpO2 between 88 and 100% with the usual oximeter and nasal cannula.
- SpO2 < 92% ambient air Specific for the intubate patient FiO2 <= 0.60 with SpO2 between 88 and 100% with usual oximeter SpO2 < 92% with FiO2 0.21
Exclusion Criteria:
- No SpO2 signal with oximeter in use
- False nails or nail polish
- Methemoglobinemia >0.015 on last available arterial gas
- Patient in isolation (multi-resistant bacteria, C-Difficile, SARS-CoV-2 ...) Specific for the extubate patient
- Expected to use another respiratory support within one hour of inclusion (NIV or nasal high flow oxygen) Planned extubation within one hour of inclusion
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:処理
- 割り当て:ランダム化
- 介入モデル:クロスオーバー割り当て
- マスキング:独身
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
実験的:Nonin oximeter with a 90% SpO2 target
During this periods , oxygen will be administered in an manual titration with the FreeO2 device (fixed flow mode) to reach 90% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Nonin oximeter |
|
実験的:Philips oximeter with a 90% SpO2 target
During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 90% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Philips oximeter |
|
実験的:Nonin oximeter with a 94% SpO2 target
During this periods , oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode) to reach 94% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter |
|
実験的:Philips oximeter with a 94% SpO2 target
During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 94% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter |
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Oxygen flow (or FiO2)
時間枠:Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Oxygen flows (or FiO2) at the end of each study period (last minute average) to reach SpO2 target to reach SpO2 target with specific oximeter |
Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Difference between SpO2 and SaO2
時間枠:assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Difference (%) between SpO2 from each oximeter and SaO2 (% of SaO2 overestimation and underestimation for each oximeter) |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
その他の成果指標
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Mean difference on oxymeter
時間枠:assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Mean (±SD) of the gaps between the different oximeters tested |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
|
Occult hypoxemia
時間枠:assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Percentage of occult hypoxemia (SaO2<90% and SpO2>=90%) |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
|
Occult hyperoxemia
時間枠:assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Perrcentage of occult hyperoxemia (SaO2>96% and SpO2<=96%) |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
|
Partial oxygen weaning
時間枠:Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Rate of partial oxygen weaning (Oxygen flow <=0.5 L/min) |
Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
|
Complete oxygen weaning
時間枠:Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Rate of complete oxygen weaning (Oxygen flow =0 L/min or FiO2 0.21) |
Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
|
Respiratory support escalation
時間枠:Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Indication of initiation of respiratory support escalation (Oxygen flow >6 L/min or FiO2 > 0.60) |
Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
協力者と研究者
スポンサー
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (推定)
研究の完了 (推定)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。