Vergelijking van twee oxygenatiedoelen met twee verschillende oximeters - Invloed op zuurstofstroomsnelheden en oxygenatie (Oxygap pong)

3 november 2026 bijgewerkt door: François Lellouche, Laval University

The investigators recently evaluated 4 different oximeters among the most commonly used with arterial catheter in place and compared SpO2 with SaO2 obtained on arterial gas. Correlations between SaO2 and SpO2 were poor for all oximeters, as previously known, and SpO2-SaO2 bias were different between oximeters. Some oximeters (Masimo, Nellcor) had lower biases but they detected less well hypoxemia. Some oximeters underestimated SaO2 (Nonin) but detected very well hypoxemia, and some overestimated SaO2 (Philips). The investigators concluded that oximeters provide different informations to clinicians, and oxygenation targets should take into account for these differences.

The assumption is that the SpO2 target AND oximeter used will both have an impact on oxygen flows and that these effects will add up. With a high SpO2 target, oxygen flows will be significantly greater and with the Nonin oximeter, the required flows will be greater than with the Philips oximeter.

NB: the results obtained were in a population with light skin pigmentation (96% of the patients were Fitzpatrick 1-2, reflecting the local hospitalized population).

Studie Overzicht

Gedetailleerde beschrijving

The oximeters to be evaluated will be the Nonin (Plymouth, MN), and the Philips (Eindhoven, Netherlands). The sensors used will be digital reusable sensors for all oximeters. Because of the impact of the fingers used for SpO2 measurement, each oximeter will be randomly assigned a different finger to each patient (digit 2 and 3 on the hand opposite to the arterial catheter).

Two SpO2 targets will be assessed: 90% and 94%. A total of four 10-minute periods will be performed in randomized order: Nonin 90, Nonin 94, Philips 90, and Philips 94.

At the end of each 10-minute period, arterial gas will be collected through the arterial catheter.

The adjustment of the oxygen flow rates will be performed manually. The average oxygen flow of the last 2 minutes of the period will be considered.

Studietype

Ingrijpend

Inschrijving (Geschat)

40

Fase

  • Niet toepasbaar

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studiecontact

Studie Contact Back-up

Studie Locaties

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

18 jaar en ouder (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Nee

Geslachten die in aanmerking komen voor studie

Allemaal

Beschrijving

Inclusion Criteria:

  • ≥ 18 years of age
  • Patients admitted to postoperative cardiac surgery intensive care unit
  • Presence of an arterial catheter Specific for the extubate patient
  • Ongoing on conventional oxygen therapy at moderate flow (up to a maximum of 5L/min) with SpO2 between 88 and 100% with the usual oximeter and nasal cannula.
  • SpO2 < 92% ambient air Specific for the intubate patient FiO2 <= 0.60 with SpO2 between 88 and 100% with usual oximeter SpO2 < 92% with FiO2 0.21

Exclusion Criteria:

  • No SpO2 signal with oximeter in use
  • False nails or nail polish
  • Methemoglobinemia >0.015 on last available arterial gas
  • Patient in isolation (multi-resistant bacteria, C-Difficile, SARS-CoV-2 ...) Specific for the extubate patient
  • Expected to use another respiratory support within one hour of inclusion (NIV or nasal high flow oxygen) Planned extubation within one hour of inclusion

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Primair doel: Behandeling
  • Toewijzing: Gerandomiseerd
  • Interventioneel model: Crossover-opdracht
  • Masker: Enkel

Wapens en interventies

Deelnemersgroep / Arm
Interventie / Behandeling
Experimenteel: Nonin oximeter with a 90% SpO2 target

During this periods , oxygen will be administered in an manual titration with the FreeO2 device (fixed flow mode) to reach 90% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Nonin oximeter

Experimenteel: Philips oximeter with a 90% SpO2 target

During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 90% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Philips oximeter

Experimenteel: Nonin oximeter with a 94% SpO2 target

During this periods , oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode) to reach 94% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter

Experimenteel: Philips oximeter with a 94% SpO2 target

During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 94% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Oxygen flow (or FiO2)
Tijdsspanne: Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)

Oxygen flows (or FiO2) at the end of each study period (last minute average) to reach SpO2 target to reach SpO2 target with specific oximeter

Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)

Secundaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Difference between SpO2 and SaO2
Tijdsspanne: assessed up to 10 minutes - at the same of the arterial blood gases punction

Difference (%) between SpO2 from each oximeter and SaO2 (% of SaO2 overestimation and underestimation for each oximeter)

assessed up to 10 minutes - at the same of the arterial blood gases punction

Andere uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Mean difference on oxymeter
Tijdsspanne: assessed up to 10 minutes - at the same of the arterial blood gases punction

Mean (±SD) of the gaps between the different oximeters tested

assessed up to 10 minutes - at the same of the arterial blood gases punction
Occult hypoxemia
Tijdsspanne: assessed up to 10 minutes - at the same of the arterial blood gases punction

Percentage of occult hypoxemia (SaO2<90% and SpO2>=90%)

assessed up to 10 minutes - at the same of the arterial blood gases punction
Occult hyperoxemia
Tijdsspanne: assessed up to 10 minutes - at the same of the arterial blood gases punction

Perrcentage of occult hyperoxemia (SaO2>96% and SpO2<=96%)

assessed up to 10 minutes - at the same of the arterial blood gases punction
Partial oxygen weaning
Tijdsspanne: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Rate of partial oxygen weaning (Oxygen flow <=0.5 L/min)

Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
Complete oxygen weaning
Tijdsspanne: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Rate of complete oxygen weaning (Oxygen flow =0 L/min or FiO2 0.21)

Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
Respiratory support escalation
Tijdsspanne: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Indication of initiation of respiratory support escalation (Oxygen flow >6 L/min or FiO2 > 0.60)

Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Werkelijk)

16 november 2022

Primaire voltooiing (Geschat)

1 juni 2027

Studie voltooiing (Geschat)

1 december 2027

Studieregistratiedata

Eerst ingediend

14 oktober 2022

Eerst ingediend dat voldeed aan de QC-criteria

18 oktober 2022

Eerst geplaatst (Werkelijk)

21 oktober 2022

Updates van studierecords

Laatste update geplaatst (Werkelijk)

5 november 2026

Laatste update ingediend die voldeed aan QC-criteria

3 november 2026

Laatst geverifieerd

1 november 2026

Meer informatie

Termen gerelateerd aan deze studie

Andere studie-ID-nummers

  • 22266

Plan Individuele Deelnemersgegevens (IPD)

Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?

NEE

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Nee

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .

Abonneren