Comparação de duas metas de oxigenação com dois oxímetros diferentes - impacto nas taxas de fluxo de oxigênio e na oxigenação (Oxygap pong)

3 de novembro de 2026 atualizado por: François Lellouche, Laval University

The investigators recently evaluated 4 different oximeters among the most commonly used with arterial catheter in place and compared SpO2 with SaO2 obtained on arterial gas. Correlations between SaO2 and SpO2 were poor for all oximeters, as previously known, and SpO2-SaO2 bias were different between oximeters. Some oximeters (Masimo, Nellcor) had lower biases but they detected less well hypoxemia. Some oximeters underestimated SaO2 (Nonin) but detected very well hypoxemia, and some overestimated SaO2 (Philips). The investigators concluded that oximeters provide different informations to clinicians, and oxygenation targets should take into account for these differences.

The assumption is that the SpO2 target AND oximeter used will both have an impact on oxygen flows and that these effects will add up. With a high SpO2 target, oxygen flows will be significantly greater and with the Nonin oximeter, the required flows will be greater than with the Philips oximeter.

NB: the results obtained were in a population with light skin pigmentation (96% of the patients were Fitzpatrick 1-2, reflecting the local hospitalized population).

Visão geral do estudo

Descrição detalhada

The oximeters to be evaluated will be the Nonin (Plymouth, MN), and the Philips (Eindhoven, Netherlands). The sensors used will be digital reusable sensors for all oximeters. Because of the impact of the fingers used for SpO2 measurement, each oximeter will be randomly assigned a different finger to each patient (digit 2 and 3 on the hand opposite to the arterial catheter).

Two SpO2 targets will be assessed: 90% and 94%. A total of four 10-minute periods will be performed in randomized order: Nonin 90, Nonin 94, Philips 90, and Philips 94.

At the end of each 10-minute period, arterial gas will be collected through the arterial catheter.

The adjustment of the oxygen flow rates will be performed manually. The average oxygen flow of the last 2 minutes of the period will be considered.

Tipo de estudo

Intervencional

Inscrição (Estimado)

40

Estágio

  • Não aplicável

Contactos e Locais

Esta seção fornece os detalhes de contato para aqueles que conduzem o estudo e informações sobre onde este estudo está sendo realizado.

Contato de estudo

Estude backup de contato

Locais de estudo

Critérios de participação

Os pesquisadores procuram pessoas que se encaixem em uma determinada descrição, chamada de critérios de elegibilidade. Alguns exemplos desses critérios são a condição geral de saúde de uma pessoa ou tratamentos anteriores.

Critérios de elegibilidade

Idades elegíveis para estudo

18 anos e mais velhos (Adulto, Adulto mais velho)

Aceita Voluntários Saudáveis

Não

Gêneros Elegíveis para o Estudo

Tudo

Descrição

Inclusion Criteria:

  • ≥ 18 years of age
  • Patients admitted to postoperative cardiac surgery intensive care unit
  • Presence of an arterial catheter Specific for the extubate patient
  • Ongoing on conventional oxygen therapy at moderate flow (up to a maximum of 5L/min) with SpO2 between 88 and 100% with the usual oximeter and nasal cannula.
  • SpO2 < 92% ambient air Specific for the intubate patient FiO2 <= 0.60 with SpO2 between 88 and 100% with usual oximeter SpO2 < 92% with FiO2 0.21

Exclusion Criteria:

  • No SpO2 signal with oximeter in use
  • False nails or nail polish
  • Methemoglobinemia >0.015 on last available arterial gas
  • Patient in isolation (multi-resistant bacteria, C-Difficile, SARS-CoV-2 ...) Specific for the extubate patient
  • Expected to use another respiratory support within one hour of inclusion (NIV or nasal high flow oxygen) Planned extubation within one hour of inclusion

Plano de estudo

Esta seção fornece detalhes do plano de estudo, incluindo como o estudo é projetado e o que o estudo está medindo.

Como o estudo é projetado?

Detalhes do projeto

  • Finalidade Principal: Tratamento
  • Alocação: Randomizado
  • Modelo Intervencional: Atribuição cruzada
  • Mascaramento: Solteiro

Armas e Intervenções

Grupo de Participantes / Braço
Intervenção / Tratamento
Experimental: Nonin oximeter with a 90% SpO2 target

During this periods , oxygen will be administered in an manual titration with the FreeO2 device (fixed flow mode) to reach 90% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Nonin oximeter

Experimental: Philips oximeter with a 90% SpO2 target

During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 90% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Philips oximeter

Experimental: Nonin oximeter with a 94% SpO2 target

During this periods , oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode) to reach 94% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter

Experimental: Philips oximeter with a 94% SpO2 target

During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 94% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter

O que o estudo está medindo?

Medidas de resultados primários

Medida de resultado
Descrição da medida
Prazo
Oxygen flow (or FiO2)
Prazo: Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)

Oxygen flows (or FiO2) at the end of each study period (last minute average) to reach SpO2 target to reach SpO2 target with specific oximeter

Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)

Medidas de resultados secundários

Medida de resultado
Descrição da medida
Prazo
Difference between SpO2 and SaO2
Prazo: assessed up to 10 minutes - at the same of the arterial blood gases punction

Difference (%) between SpO2 from each oximeter and SaO2 (% of SaO2 overestimation and underestimation for each oximeter)

assessed up to 10 minutes - at the same of the arterial blood gases punction

Outras medidas de resultado

Medida de resultado
Descrição da medida
Prazo
Mean difference on oxymeter
Prazo: assessed up to 10 minutes - at the same of the arterial blood gases punction

Mean (±SD) of the gaps between the different oximeters tested

assessed up to 10 minutes - at the same of the arterial blood gases punction
Occult hypoxemia
Prazo: assessed up to 10 minutes - at the same of the arterial blood gases punction

Percentage of occult hypoxemia (SaO2<90% and SpO2>=90%)

assessed up to 10 minutes - at the same of the arterial blood gases punction
Occult hyperoxemia
Prazo: assessed up to 10 minutes - at the same of the arterial blood gases punction

Perrcentage of occult hyperoxemia (SaO2>96% and SpO2<=96%)

assessed up to 10 minutes - at the same of the arterial blood gases punction
Partial oxygen weaning
Prazo: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Rate of partial oxygen weaning (Oxygen flow <=0.5 L/min)

Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
Complete oxygen weaning
Prazo: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Rate of complete oxygen weaning (Oxygen flow =0 L/min or FiO2 0.21)

Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
Respiratory support escalation
Prazo: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Indication of initiation of respiratory support escalation (Oxygen flow >6 L/min or FiO2 > 0.60)

Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Colaboradores e Investigadores

É aqui que você encontrará pessoas e organizações envolvidas com este estudo.

Patrocinador

Datas de registro do estudo

Essas datas acompanham o progresso do registro do estudo e os envios de resumo dos resultados para ClinicalTrials.gov. Os registros do estudo e os resultados relatados são revisados ​​pela National Library of Medicine (NLM) para garantir que atendam aos padrões específicos de controle de qualidade antes de serem publicados no site público.

Datas Principais do Estudo

Início do estudo (Real)

16 de novembro de 2022

Conclusão Primária (Estimado)

1 de junho de 2027

Conclusão do estudo (Estimado)

1 de dezembro de 2027

Datas de inscrição no estudo

Enviado pela primeira vez

14 de outubro de 2022

Enviado pela primeira vez que atendeu aos critérios de CQ

18 de outubro de 2022

Primeira postagem (Real)

21 de outubro de 2022

Atualizações de registro de estudo

Última Atualização Postada (Real)

5 de novembro de 2026

Última atualização enviada que atendeu aos critérios de controle de qualidade

3 de novembro de 2026

Última verificação

1 de novembro de 2026

Mais Informações

Termos relacionados a este estudo

Outros números de identificação do estudo

  • 22266

Plano para dados de participantes individuais (IPD)

Planeja compartilhar dados de participantes individuais (IPD)?

NÃO

Informações sobre medicamentos e dispositivos, documentos de estudo

Estuda um medicamento regulamentado pela FDA dos EUA

Não

Estuda um produto de dispositivo regulamentado pela FDA dos EUA

Não

Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .