- ICH GCP
- Registro de ensaios clínicos dos EUA
- Ensaio Clínico NCT05590130
Comparação de duas metas de oxigenação com dois oxímetros diferentes - impacto nas taxas de fluxo de oxigênio e na oxigenação (Oxygap pong)
The investigators recently evaluated 4 different oximeters among the most commonly used with arterial catheter in place and compared SpO2 with SaO2 obtained on arterial gas. Correlations between SaO2 and SpO2 were poor for all oximeters, as previously known, and SpO2-SaO2 bias were different between oximeters. Some oximeters (Masimo, Nellcor) had lower biases but they detected less well hypoxemia. Some oximeters underestimated SaO2 (Nonin) but detected very well hypoxemia, and some overestimated SaO2 (Philips). The investigators concluded that oximeters provide different informations to clinicians, and oxygenation targets should take into account for these differences.
The assumption is that the SpO2 target AND oximeter used will both have an impact on oxygen flows and that these effects will add up. With a high SpO2 target, oxygen flows will be significantly greater and with the Nonin oximeter, the required flows will be greater than with the Philips oximeter.
NB: the results obtained were in a population with light skin pigmentation (96% of the patients were Fitzpatrick 1-2, reflecting the local hospitalized population).
Visão geral do estudo
Status
Condições
Intervenção / Tratamento
- Dispositivo: Manual oxygen titration with Nonin oximeter (SpO2 target 90%)
- Dispositivo: Manual Oxygen titration with Philips oximeter (SpO2 target 90%)
- Dispositivo: Manual oxygen titration with Nonin oximeter (SpO2 target 94%)
- Dispositivo: Manual Oxygen titration with Philips oximeter (SpO2 target 94%)
Descrição detalhada
The oximeters to be evaluated will be the Nonin (Plymouth, MN), and the Philips (Eindhoven, Netherlands). The sensors used will be digital reusable sensors for all oximeters. Because of the impact of the fingers used for SpO2 measurement, each oximeter will be randomly assigned a different finger to each patient (digit 2 and 3 on the hand opposite to the arterial catheter).
Two SpO2 targets will be assessed: 90% and 94%. A total of four 10-minute periods will be performed in randomized order: Nonin 90, Nonin 94, Philips 90, and Philips 94.
At the end of each 10-minute period, arterial gas will be collected through the arterial catheter.
The adjustment of the oxygen flow rates will be performed manually. The average oxygen flow of the last 2 minutes of the period will be considered.
Tipo de estudo
Inscrição (Estimado)
Estágio
- Não aplicável
Contactos e Locais
Contato de estudo
- Nome: Francois Lellouche
- Número de telefone: 3572 418656-8711
- E-mail: francois.lellouche@criucpq.ulaval.ca
Estude backup de contato
- Nome: Pierre Alexandre Bouchard
- Número de telefone: 2712 418656-8711
- E-mail: pierre-alexandre.bouchard@criucpq.ulaval.ca
Locais de estudo
-
-
Quebec
-
Québec, Quebec, Canadá, G1V4G5
- Recrutamento
- Institut Universitaire de Cardiologie et de Pneumologie de Québec
-
Contato:
- Francois Lellouche
- E-mail: francois.lellouche@criucpq.ulaval.ca
-
-
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
Aceita Voluntários Saudáveis
Gêneros Elegíveis para o Estudo
Descrição
Inclusion Criteria:
- ≥ 18 years of age
- Patients admitted to postoperative cardiac surgery intensive care unit
- Presence of an arterial catheter Specific for the extubate patient
- Ongoing on conventional oxygen therapy at moderate flow (up to a maximum of 5L/min) with SpO2 between 88 and 100% with the usual oximeter and nasal cannula.
- SpO2 < 92% ambient air Specific for the intubate patient FiO2 <= 0.60 with SpO2 between 88 and 100% with usual oximeter SpO2 < 92% with FiO2 0.21
Exclusion Criteria:
- No SpO2 signal with oximeter in use
- False nails or nail polish
- Methemoglobinemia >0.015 on last available arterial gas
- Patient in isolation (multi-resistant bacteria, C-Difficile, SARS-CoV-2 ...) Specific for the extubate patient
- Expected to use another respiratory support within one hour of inclusion (NIV or nasal high flow oxygen) Planned extubation within one hour of inclusion
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Finalidade Principal: Tratamento
- Alocação: Randomizado
- Modelo Intervencional: Atribuição cruzada
- Mascaramento: Solteiro
Armas e Intervenções
Grupo de Participantes / Braço |
Intervenção / Tratamento |
|---|---|
|
Experimental: Nonin oximeter with a 90% SpO2 target
During this periods , oxygen will be administered in an manual titration with the FreeO2 device (fixed flow mode) to reach 90% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Nonin oximeter |
|
Experimental: Philips oximeter with a 90% SpO2 target
During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 90% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Philips oximeter |
|
Experimental: Nonin oximeter with a 94% SpO2 target
During this periods , oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode) to reach 94% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter |
|
Experimental: Philips oximeter with a 94% SpO2 target
During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 94% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter |
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Oxygen flow (or FiO2)
Prazo: Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Oxygen flows (or FiO2) at the end of each study period (last minute average) to reach SpO2 target to reach SpO2 target with specific oximeter |
Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Medidas de resultados secundários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Difference between SpO2 and SaO2
Prazo: assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Difference (%) between SpO2 from each oximeter and SaO2 (% of SaO2 overestimation and underestimation for each oximeter) |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Outras medidas de resultado
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Mean difference on oxymeter
Prazo: assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Mean (±SD) of the gaps between the different oximeters tested |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
|
Occult hypoxemia
Prazo: assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Percentage of occult hypoxemia (SaO2<90% and SpO2>=90%) |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
|
Occult hyperoxemia
Prazo: assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Perrcentage of occult hyperoxemia (SaO2>96% and SpO2<=96%) |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
|
Partial oxygen weaning
Prazo: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Rate of partial oxygen weaning (Oxygen flow <=0.5 L/min) |
Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
|
Complete oxygen weaning
Prazo: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Rate of complete oxygen weaning (Oxygen flow =0 L/min or FiO2 0.21) |
Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
|
Respiratory support escalation
Prazo: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Indication of initiation of respiratory support escalation (Oxygen flow >6 L/min or FiO2 > 0.60) |
Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Colaboradores e Investigadores
Patrocinador
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Real)
Conclusão Primária (Estimado)
Conclusão do estudo (Estimado)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Termos MeSH relevantes adicionais
Outros números de identificação do estudo
- 22266
Plano para dados de participantes individuais (IPD)
Planeja compartilhar dados de participantes individuais (IPD)?
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .