Comparaison de deux cibles d'oxygénation avec deux oxymètres différents - Impact sur les débits d'oxygène et l'oxygénation (Oxygap pong)
The investigators recently evaluated 4 different oximeters among the most commonly used with arterial catheter in place and compared SpO2 with SaO2 obtained on arterial gas. Correlations between SaO2 and SpO2 were poor for all oximeters, as previously known, and SpO2-SaO2 bias were different between oximeters. Some oximeters (Masimo, Nellcor) had lower biases but they detected less well hypoxemia. Some oximeters underestimated SaO2 (Nonin) but detected very well hypoxemia, and some overestimated SaO2 (Philips). The investigators concluded that oximeters provide different informations to clinicians, and oxygenation targets should take into account for these differences.
The assumption is that the SpO2 target AND oximeter used will both have an impact on oxygen flows and that these effects will add up. With a high SpO2 target, oxygen flows will be significantly greater and with the Nonin oximeter, the required flows will be greater than with the Philips oximeter.
NB: the results obtained were in a population with light skin pigmentation (96% of the patients were Fitzpatrick 1-2, reflecting the local hospitalized population).
Aperçu de l'étude
Statut
Les conditions
Intervention / Traitement
Description détaillée
The oximeters to be evaluated will be the Nonin (Plymouth, MN), and the Philips (Eindhoven, Netherlands). The sensors used will be digital reusable sensors for all oximeters. Because of the impact of the fingers used for SpO2 measurement, each oximeter will be randomly assigned a different finger to each patient (digit 2 and 3 on the hand opposite to the arterial catheter).
Two SpO2 targets will be assessed: 90% and 94%. A total of four 10-minute periods will be performed in randomized order: Nonin 90, Nonin 94, Philips 90, and Philips 94.
At the end of each 10-minute period, arterial gas will be collected through the arterial catheter.
The adjustment of the oxygen flow rates will be performed manually. The average oxygen flow of the last 2 minutes of the period will be considered.
Type d'étude
Inscription (Estimé)
Phase
- N'est pas applicable
Contacts et emplacements
Coordonnées de l'étude
- Nom: Francois Lellouche
- Numéro de téléphone: 3572 418656-8711
- E-mail: francois.lellouche@criucpq.ulaval.ca
Sauvegarde des contacts de l'étude
- Nom: Pierre Alexandre Bouchard
- Numéro de téléphone: 2712 418656-8711
- E-mail: pierre-alexandre.bouchard@criucpq.ulaval.ca
Lieux d'étude
-
-
Quebec
-
Québec, Quebec, Canada, G1V4G5
- Recrutement
- Institut Universitaire de Cardiologie et de Pneumologie de Québec
-
Contact:
- Francois Lellouche
- E-mail: francois.lellouche@criucpq.ulaval.ca
-
-
Critères de participation
Critère d'éligibilité
Âges éligibles pour étudier
Accepte les volontaires sains
Sexes éligibles pour l'étude
La description
Inclusion Criteria:
- ≥ 18 years of age
- Patients admitted to postoperative cardiac surgery intensive care unit
- Presence of an arterial catheter Specific for the extubate patient
- Ongoing on conventional oxygen therapy at moderate flow (up to a maximum of 5L/min) with SpO2 between 88 and 100% with the usual oximeter and nasal cannula.
- SpO2 < 92% ambient air Specific for the intubate patient FiO2 <= 0.60 with SpO2 between 88 and 100% with usual oximeter SpO2 < 92% with FiO2 0.21
Exclusion Criteria:
- No SpO2 signal with oximeter in use
- False nails or nail polish
- Methemoglobinemia >0.015 on last available arterial gas
- Patient in isolation (multi-resistant bacteria, C-Difficile, SARS-CoV-2 ...) Specific for the extubate patient
- Expected to use another respiratory support within one hour of inclusion (NIV or nasal high flow oxygen) Planned extubation within one hour of inclusion
Plan d'étude
Comment l'étude est-elle conçue ?
Détails de conception
- Objectif principal: Traitement
- Répartition: Randomisé
- Modèle interventionnel: Affectation croisée
- Masquage: Seul
Armes et Interventions
Groupe de participants / Bras |
Intervention / Traitement |
|---|---|
|
Expérimental: Nonin oximeter with a 90% SpO2 target
During this periods , oxygen will be administered in an manual titration with the FreeO2 device (fixed flow mode) to reach 90% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Nonin oximeter |
|
Expérimental: Philips oximeter with a 90% SpO2 target
During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 90% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Philips oximeter |
|
Expérimental: Nonin oximeter with a 94% SpO2 target
During this periods , oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode) to reach 94% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter |
|
Expérimental: Philips oximeter with a 94% SpO2 target
During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 94% of SpO2. |
Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter |
Que mesure l'étude ?
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Oxygen flow (or FiO2)
Délai: Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Oxygen flows (or FiO2) at the end of each study period (last minute average) to reach SpO2 target to reach SpO2 target with specific oximeter |
Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Mesures de résultats secondaires
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Difference between SpO2 and SaO2
Délai: assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Difference (%) between SpO2 from each oximeter and SaO2 (% of SaO2 overestimation and underestimation for each oximeter) |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Autres mesures de résultats
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Mean difference on oxymeter
Délai: assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Mean (±SD) of the gaps between the different oximeters tested |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
|
Occult hypoxemia
Délai: assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Percentage of occult hypoxemia (SaO2<90% and SpO2>=90%) |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
|
Occult hyperoxemia
Délai: assessed up to 10 minutes - at the same of the arterial blood gases punction
|
Perrcentage of occult hyperoxemia (SaO2>96% and SpO2<=96%) |
assessed up to 10 minutes - at the same of the arterial blood gases punction
|
|
Partial oxygen weaning
Délai: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Rate of partial oxygen weaning (Oxygen flow <=0.5 L/min) |
Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
|
Complete oxygen weaning
Délai: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Rate of complete oxygen weaning (Oxygen flow =0 L/min or FiO2 0.21) |
Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
|
Respiratory support escalation
Délai: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Indication of initiation of respiratory support escalation (Oxygen flow >6 L/min or FiO2 > 0.60) |
Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
|
Collaborateurs et enquêteurs
Parrainer
Dates d'enregistrement des études
Dates principales de l'étude
Début de l'étude (Réel)
Achèvement primaire (Estimé)
Achèvement de l'étude (Estimé)
Dates d'inscription aux études
Première soumission
Première soumission répondant aux critères de contrôle qualité
Première publication (Réel)
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Réel)
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière vérification
Plus d'information
Termes liés à cette étude
Termes MeSH pertinents supplémentaires
Autres numéros d'identification d'étude
- 22266
Plan pour les données individuelles des participants (IPD)
Prévoyez-vous de partager les données individuelles des participants (DPI) ?
Informations sur les médicaments et les dispositifs, documents d'étude
Étudie un produit pharmaceutique réglementé par la FDA américaine
Étudie un produit d'appareil réglementé par la FDA américaine
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .