Vergleich von zwei Oxygenierungszielen mit zwei verschiedenen Oximetern – Einfluss auf die Sauerstoffdurchflussraten und die Oxygenierung (Oxygap pong)

3. November 2026 aktualisiert von: François Lellouche, Laval University

The investigators recently evaluated 4 different oximeters among the most commonly used with arterial catheter in place and compared SpO2 with SaO2 obtained on arterial gas. Correlations between SaO2 and SpO2 were poor for all oximeters, as previously known, and SpO2-SaO2 bias were different between oximeters. Some oximeters (Masimo, Nellcor) had lower biases but they detected less well hypoxemia. Some oximeters underestimated SaO2 (Nonin) but detected very well hypoxemia, and some overestimated SaO2 (Philips). The investigators concluded that oximeters provide different informations to clinicians, and oxygenation targets should take into account for these differences.

The assumption is that the SpO2 target AND oximeter used will both have an impact on oxygen flows and that these effects will add up. With a high SpO2 target, oxygen flows will be significantly greater and with the Nonin oximeter, the required flows will be greater than with the Philips oximeter.

NB: the results obtained were in a population with light skin pigmentation (96% of the patients were Fitzpatrick 1-2, reflecting the local hospitalized population).

Studienübersicht

Detaillierte Beschreibung

The oximeters to be evaluated will be the Nonin (Plymouth, MN), and the Philips (Eindhoven, Netherlands). The sensors used will be digital reusable sensors for all oximeters. Because of the impact of the fingers used for SpO2 measurement, each oximeter will be randomly assigned a different finger to each patient (digit 2 and 3 on the hand opposite to the arterial catheter).

Two SpO2 targets will be assessed: 90% and 94%. A total of four 10-minute periods will be performed in randomized order: Nonin 90, Nonin 94, Philips 90, and Philips 94.

At the end of each 10-minute period, arterial gas will be collected through the arterial catheter.

The adjustment of the oxygen flow rates will be performed manually. The average oxygen flow of the last 2 minutes of the period will be considered.

Studientyp

Interventionell

Einschreibung (Geschätzt)

40

Phase

  • Unzutreffend

Kontakte und Standorte

Dieser Abschnitt enthält die Kontaktdaten derjenigen, die die Studie durchführen, und Informationen darüber, wo diese Studie durchgeführt wird.

Studienkontakt

Studieren Sie die Kontaktsicherung

Studienorte

Teilnahmekriterien

Forscher suchen nach Personen, die einer bestimmten Beschreibung entsprechen, die als Auswahlkriterien bezeichnet werden. Einige Beispiele für diese Kriterien sind der allgemeine Gesundheitszustand einer Person oder frühere Behandlungen.

Zulassungskriterien

Studienberechtigtes Alter

18 Jahre und älter (Erwachsene, Älterer Erwachsener)

Akzeptiert gesunde Freiwillige

Nein

Studienberechtigte Geschlechter

Alle

Beschreibung

Inclusion Criteria:

  • ≥ 18 years of age
  • Patients admitted to postoperative cardiac surgery intensive care unit
  • Presence of an arterial catheter Specific for the extubate patient
  • Ongoing on conventional oxygen therapy at moderate flow (up to a maximum of 5L/min) with SpO2 between 88 and 100% with the usual oximeter and nasal cannula.
  • SpO2 < 92% ambient air Specific for the intubate patient FiO2 <= 0.60 with SpO2 between 88 and 100% with usual oximeter SpO2 < 92% with FiO2 0.21

Exclusion Criteria:

  • No SpO2 signal with oximeter in use
  • False nails or nail polish
  • Methemoglobinemia >0.015 on last available arterial gas
  • Patient in isolation (multi-resistant bacteria, C-Difficile, SARS-CoV-2 ...) Specific for the extubate patient
  • Expected to use another respiratory support within one hour of inclusion (NIV or nasal high flow oxygen) Planned extubation within one hour of inclusion

Studienplan

Dieser Abschnitt enthält Einzelheiten zum Studienplan, einschließlich des Studiendesigns und der Messung der Studieninhalte.

Wie ist die Studie aufgebaut?

Designdetails

  • Hauptzweck: Behandlung
  • Zuteilung: Zufällig
  • Interventionsmodell: Crossover-Aufgabe
  • Maskierung: Single

Waffen und Interventionen

Teilnehmergruppe / Arm
Intervention / Behandlung
Experimental: Nonin oximeter with a 90% SpO2 target

During this periods , oxygen will be administered in an manual titration with the FreeO2 device (fixed flow mode) to reach 90% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Nonin oximeter

Experimental: Philips oximeter with a 90% SpO2 target

During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 90% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 90%) with Philips oximeter

Experimental: Nonin oximeter with a 94% SpO2 target

During this periods , oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode) to reach 94% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter

Experimental: Philips oximeter with a 94% SpO2 target

During this periods, oxygen will be administered in a manual titration with the FreeO2 device (fixed flow mode), to reach 94% of SpO2.

Oxygen titration with FreeO2 device (fixed flow mode) to reach the SpO2 target (set at 94%) with Philips oximeter

Was misst die Studie?

Primäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Oxygen flow (or FiO2)
Zeitfenster: Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)

Oxygen flows (or FiO2) at the end of each study period (last minute average) to reach SpO2 target to reach SpO2 target with specific oximeter

Mean Oxygen flow in the two last minute of each study period (between minute 8 and 10)

Sekundäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Difference between SpO2 and SaO2
Zeitfenster: assessed up to 10 minutes - at the same of the arterial blood gases punction

Difference (%) between SpO2 from each oximeter and SaO2 (% of SaO2 overestimation and underestimation for each oximeter)

assessed up to 10 minutes - at the same of the arterial blood gases punction

Andere Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Mean difference on oxymeter
Zeitfenster: assessed up to 10 minutes - at the same of the arterial blood gases punction

Mean (±SD) of the gaps between the different oximeters tested

assessed up to 10 minutes - at the same of the arterial blood gases punction
Occult hypoxemia
Zeitfenster: assessed up to 10 minutes - at the same of the arterial blood gases punction

Percentage of occult hypoxemia (SaO2<90% and SpO2>=90%)

assessed up to 10 minutes - at the same of the arterial blood gases punction
Occult hyperoxemia
Zeitfenster: assessed up to 10 minutes - at the same of the arterial blood gases punction

Perrcentage of occult hyperoxemia (SaO2>96% and SpO2<=96%)

assessed up to 10 minutes - at the same of the arterial blood gases punction
Partial oxygen weaning
Zeitfenster: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Rate of partial oxygen weaning (Oxygen flow <=0.5 L/min)

Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
Complete oxygen weaning
Zeitfenster: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Rate of complete oxygen weaning (Oxygen flow =0 L/min or FiO2 0.21)

Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)
Respiratory support escalation
Zeitfenster: Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Indication of initiation of respiratory support escalation (Oxygen flow >6 L/min or FiO2 > 0.60)

Mean oxygen flow in the two last minute of each study period (between minute 8 and 10)

Mitarbeiter und Ermittler

Hier finden Sie Personen und Organisationen, die an dieser Studie beteiligt sind.

Studienaufzeichnungsdaten

Diese Daten verfolgen den Fortschritt der Übermittlung von Studienaufzeichnungen und zusammenfassenden Ergebnissen an ClinicalTrials.gov. Studienaufzeichnungen und gemeldete Ergebnisse werden von der National Library of Medicine (NLM) überprüft, um sicherzustellen, dass sie bestimmten Qualitätskontrollstandards entsprechen, bevor sie auf der öffentlichen Website veröffentlicht werden.

Haupttermine studieren

Studienbeginn (Tatsächlich)

16. November 2022

Primärer Abschluss (Geschätzt)

1. Juni 2027

Studienabschluss (Geschätzt)

1. Dezember 2027

Studienanmeldedaten

Zuerst eingereicht

14. Oktober 2022

Zuerst eingereicht, das die QC-Kriterien erfüllt hat

18. Oktober 2022

Zuerst gepostet (Tatsächlich)

21. Oktober 2022

Studienaufzeichnungsaktualisierungen

Letztes Update gepostet (Tatsächlich)

5. November 2026

Letztes eingereichtes Update, das die QC-Kriterien erfüllt

3. November 2026

Zuletzt verifiziert

1. November 2026

Mehr Informationen

Begriffe im Zusammenhang mit dieser Studie

Zusätzliche relevante MeSH-Bedingungen

Andere Studien-ID-Nummern

  • 22266

Plan für individuelle Teilnehmerdaten (IPD)

Planen Sie, individuelle Teilnehmerdaten (IPD) zu teilen?

NEIN

Arzneimittel- und Geräteinformationen, Studienunterlagen

Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt

Nein

Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt

Nein

Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .