乳房異常患者に対する乳房生検に関する分子乳房画像診断ガイダンス
分子乳房イメージングを評価するためのパイロット臨床研究 - 乳房異常のサンプリングに関するガイダンス
This clinical trial tests how well molecular breast imaging (MBI) works to guide the collection of a breast biopsy in patients with a breast abnormality. Currently, a biopsy is often guided by either ultrasound or mammography in order to ensure that a sample of the correct part of the breast is taken. Sometimes a lesion or part of the lesion cannot be seen on ultrasound or mammography, and, therefore, a biopsy guided with ultrasound or mammography may not be always be accurate. Studies have shown that high resolution MBI may have potential to improve the detection of some breast tumors. This trial uses a new high-resolution MBI system that may help perform a biopsy using MBI.
調査の概要
状態
条件
詳細な説明
PRIMARY OBJECTIVE:
I. To evaluate a methodology for MBI-guided biopsy of suspicious breast lesions.
OUTLINE: Patients are assigned to 1 of 2 study parts.
PART I: Patients receive technetium Tc-99m sestamibi intravenously (IV) and undergo MBI on study.
PART II: Patients receive technetium Tc-99m sestamibi IV and undergo MBI. Patients whose breast lesions of interest are visualized on MBI then undergo breast biopsy using the Stereo Navigator accessory.
After completion of study intervention, patients may be asked to follow up at 3-6 months.
研究の種類
入学 (推定)
段階
- 適用できない
連絡先と場所
研究連絡先
- 名前:Clinical Trials Referral Office
- 電話番号:855-776-0015
- メール:mayocliniccancerstudies@mayo.edu
研究場所
-
-
Minnesota
-
Rochester、Minnesota、アメリカ、55905
- 募集
- Mayo Clinic in Rochester
-
コンタクト:
- Clinical Trials Referral Office
- 電話番号:855-776-0015
- メール:mayocliniccancerstudies@mayo.edu
-
主任研究者:
- Katie N. Hunt, M.D.
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
-
PART I: Women aged 18 years or older
-
PART I: If able to become pregnant, negative pregnancy test within 48 hours prior to MBI biopsy procedure
-
PART I: Individuals who have previously (within the last 3 years) undergo an MBI study and were found to have at least one breast imaging finding on MBI that was subsequently shown (through additional imaging, biopsy or follow-up) to be benign in nature
-
PART I: Individuals who have agreed to participate in the study and who have signed the study-specific informed consent
-
PART II: Women aged 18 years or older
-
PART II: If able to become pregnant, negative pregnancy test within 48 hours prior to MBI biopsy procedure
-
PART II: Individuals who have previously (within the last 3 years) undergone an MBI study and were found to have at least one breast imaging finding on MBI that was subsequently shown (through additional imaging, biopsy or follow-up) to be benign or probably benign in nature
-
PART II: Individuals who had recent conventional imaging work-up including either x-ray mammography, ultrasound, MBI, or MRI of the breast and are found to have at least one breast imaging finding for which biopsy is required or recommended, specifically:
- Individuals who have a breast abnormality(ies) on imaging with mammography, ultrasound, MBI, or MRI (as per American College of Radiology [ACR] Breast Imaging Reporting and Data System [BIRADS] 3, or higher) and requiring imaging follow-up or biopsy confirmation
-
PART II: Individuals who have agreed to participate in the study and who have signed study-specific informed consent
Exclusion Criteria:
- PART I: Women who are pregnant
- PART I: Women who are currently lactating or discontinued breastfeeding < 2 months prior to the study
- PART I: Age less than 18 years
- PART I: Women with breast implant(s) in the breast containing the lesion of interest
- PART I: Inability to provide informed consent
- PART II: Women who are pregnant
- PART II: Women who are currently lactating or discontinued breastfeeding < 2 months prior to the study
- PART II: Age less than 18 years
- PART II: Women with breast implant(s) in the breast containing the lesion of interest
- PART II: Women who have recently (within the last 3 months) undergone biopsy of the breast finding
- PART II: Women who are scheduled for a sentinel node procedure using radioactive Tc-99m within 24 hours of MBI-guided biopsy
- PART II: Patients with contraindications for core biopsy and other invasive procedures such as blood coagulation disorders, infection, or who are unwilling to discontinue use of anticoagulant medication prior to the procedure
- PART II: Inability to provide informed consent
- PART II: Women who have had surgery on the study breast(s) within the past 12 months
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:診断
- 割り当て:非ランダム化
- 介入モデル:順次割り当て
- マスキング:なし(オープンラベル)
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
実験的:Part I (technetium Tc-99m sestamibi, MBI)
Patients receive technetium Tc-99m sestamibi IV and undergo MBI on study. |
Undergo MBI
他の名前:
Given IV
他の名前:
|
|
実験的:Part II (technetium Tc-99m sestamibi, MBI, biopsy)
Patients receive technetium Tc-99m sestamibi IV and undergo MBI. Patients whose breast lesions of interest are visualized on MBI then undergo breast biopsy using the Stereo Navigator accessory. |
Undergo MBI
他の名前:
Given IV
他の名前:
Undergo biopsy of breast
他の名前:
Utilization of the Stereo Navigator accessory |
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Validation of timeframe
時間枠:Baseline
|
Time from injection of the technetium Tc-99m sestamibi to identification of the appropriate needle guide hole will be reviewed to determine if and where any improvements in workflow can be made to reduce the overall procedure time for patients. |
Baseline
|
|
Change in image of lesion
時間枠:Baseline; Up to 6 months
|
The coordinates of the lesion of interest on the high resolution MBI image will be recorded. The guidance procedure will be considered successful if 1) the pre and post images of the lesion demonstrated one of the following: decreased in size, central or eccentric defect where sampled, or decrease in intensity, or 2) pathology results demonstrate malignancy or a specific benign concordant result. |
Baseline; Up to 6 months
|
|
Accuracy of depth estimation by difference in attenuation from opposing views
時間枠:Baseline; Up to 6 months
|
For each patient in whom pathology confirms successful sampling of the lesion seen on high-resolution molecular breast imaging (MBI), the error in depth estimation will be calculated as the average error for each depth estimation method from all core biopsy samples. This will allow comparison of the accuracy of the 3 different depth estimation methods and potentially provide further information on the strengths and weakness of each of the three methods for depth estimation. Uniformity activity along the entire 20 mm core would indicate failure to biopsy the lesion. Any portion of increased uptake will be noted and the center of activity will be determined relative to the mid-point of the core sample. This distance will be recorded as the difference between estimated depth and true depth. |
Baseline; Up to 6 months
|
|
Accuracy of depth estimation by difference in apparent lesion size
時間枠:Baseline; Up to 6 months
|
For each patient in whom pathology confirms successful sampling of the lesion seen on MBI, the error in depth estimation will be calculated as the average error for each depth estimation method from all core biopsy samples. This will allow comparison of the accuracy of the 3 different depth estimation methods and potentially provide further information on the strengths and weakness of each of the three methods for depth estimation. Uniformity activity along the entire 20 mm core would indicate failure to biopsy the lesion. Any portion of increased uptake will be noted and the center of activity will be determined relative to the mid-point of the core sample. This distance will be recorded as the difference between estimated depth and true depth. |
Baseline; Up to 6 months
|
|
Accuracy of depth estimation by triangulation
時間枠:Baseline; Up to 6 months
|
For each patient in whom pathology confirms successful sampling of the lesion seen on MBI, the error in depth estimation will be calculated as the average error for each depth estimation method from all core biopsy samples. This will allow comparison of the accuracy of the 3 different depth estimation methods and potentially provide further information on the strengths and weakness of each of the three methods for depth estimation. Uniformity activity along the entire 20 mm core would indicate failure to biopsy the lesion. Any portion of increased uptake will be noted and the center of activity will be determined relative to the mid-point of the core sample. This distance will be recorded as the difference between estimated depth and true depth. |
Baseline; Up to 6 months
|
協力者と研究者
スポンサー
捜査官
- 主任研究者:Katie N. Hunt, M.D.、Mayo Clinic in Rochester
出版物と役立つリンク
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (推定)
研究の完了 (推定)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日