Supine Photon-Counting CT and Contrast-Enhanced Mammography Versus Breast MRI for Pretreatment Tumor Size in Breast Cancer Candidates for Neoadjuvant Chemotherapy (PC CEM MRI)
Retrospective single-centre agreement study in women with biopsy-proven invasive breast cancer who underwent CEM, supine PC-CT and prone breast MRI before neoadjuvant chemotherapy. Prone MRI is the clinical reference.
調査の概要
状態
詳細な説明
Breast MRI is the reference imaging test for locoregional staging but is not available to every patient before NAC. PC-CT performed for systemic staging also images both breasts and axillae. Consecutive women who underwent CEM, supine PC-CT and breast MRI before the first NAC cycle were analyzed. Two breast radiologists reviewed each modality in separate sessions. Because all patients received NAC, surgical pathology could not serve as reference for pretreatment size, so prone dynamic MRI was used as the clinical reference and the study was designed as an agreement study, not a diagnostic accuracy study
研究の種類
入学 (実際)
連絡先と場所
研究場所
-
-
Italy
-
Sassari、Italy、イタリア、07100
- Aou Sassari
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
サンプリング方法
調査対象母集団
Consecutive women with biopsy-proven invasive breast cancer referred for NAC to the Department of Radiological, Interventional and Emergency Sciences, Azienda Ospedaliero-Universitaria di Sassari, Italy, between April 2025 and June 2026.
説明
Inclusion Criteria: (a) invasive carcinoma at core needle or vacuum-assisted biopsy; (b) indication for NAC by the multidisciplinary tumor board; (c) CEM, supine PC-CT and breast MRI all performed before NAC.
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 観測モデル:COHORT
- 時間の展望:RETROSPECTIVE
コホートと介入
グループ/コホート |
介入・治療 |
|---|---|
|
NAC candidates with three imaging examinations
Women with biopsy-proven invasive breast cancer and indication for NAC who underwent all three examinations before the first NAC cycle. |
Clinically indicated contrast-enhanced oncologic staging CT of chest and abdomen, performed in the supine position on a dual-source photon-counting detector CT scanner (NAEOTOM Alpha, Siemens Healthineers). Iodinated contrast medium (370 mg I/mL; 1.2 mL/kg) at 3.5 mL/s, acquisition at a fixed delay of 65-70 s. Breast assessment used the chest portion of the acquisition, with no additional radiation exposure or contrast administration. 40-keV and 60-keV virtual monoenergetic images and iodine maps were reconstructed. Performed in the prone position on a 1.5-T scanner (MAGNETOM Altea, Siemens Healthineers) with a dedicated breast coil, including T2-weighted and multiphase T1-weighted sequences before and after gadolinium-based contrast administration. Used as the clinical reference for tumor size. CEM was performed immediately after the staging PC-CT, using the same contrast bolus, without a second injection; low- and high-energy craniocaudal and mediolateral oblique views were acquired within 6 minutes of the end of the PC-CT acquisition. |
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Agreement in maximum tumor diameter between supine PC-CT and prone MRI
時間枠:before the first NAC cycle (April 2025 to June 2026)
|
Bland-Altman mean bias and 95% limits of agreement, and intraclass correlation coefficient, of the maximum diameter of the index cancer. |
before the first NAC cycle (April 2025 to June 2026)
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
CEM vs MRI e PC-CT vs CEM
時間枠:before the first NAC cycle.
|
Proportion of measurements within ±5 mm and ±20% of MRI size |
before the first NAC cycle.
|
協力者と研究者
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
本研究に関する用語
キーワード
その他の研究ID番号
- T_2_2025
個々の参加者データ (IPD) の計画
個々の参加者データ (IPD) を共有する予定はありますか?
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。