Esta página foi traduzida automaticamente e a precisão da tradução não é garantida. Por favor, consulte o versão em inglês para um texto fonte.

Estudo de Viabilidade Antecipado da Válvula Autus

15 de outubro de 2026 atualizado por: Edwards Lifesciences

Válvula de tamanho ajustável Autus para substituição cirúrgica da válvula pulmonar Estudo de viabilidade antecipado

Prospective, single-arm, multi-center study to evaluate safety and effectiveness of the Autus Size-Adjustable Valve in pediatric patients aged 18 months to 16 years requiring surgical pulmonary valve replacement. The Autus Valve may be expanded pre-implant to match the subject's body size. Subjects will be evaluated prior to the Autus Valve implant procedure, immediately post-implant, at hospital discharge, 30 days, 6 months, and annually through 10 years.

The Autus Valve may be expanded post-implant via transcatheter balloon dilation to accommodate growth of the subject. In subjects who undergo a post-implant valve expansion, follow-up will continue for a minimum of 1 year after the post-implant valve expansion procedure.

Visão geral do estudo

Status

Ativo, não recrutando

Condições

Intervenção / Tratamento

Tipo de estudo

Intervencional

Inscrição (Real)

62

Estágio

  • Não aplicável

Contactos e Locais

Esta seção fornece os detalhes de contato para aqueles que conduzem o estudo e informações sobre onde este estudo está sendo realizado.

Locais de estudo

    • California
      • Los Angeles, California, Estados Unidos, 90027
        • Children's Hospital Los Angeles
    • Colorado
      • Aurora, Colorado, Estados Unidos, 80045
        • Children's Hospital Colorado
    • Georgia
      • Atlanta, Georgia, Estados Unidos, 30322
        • Children's Healthcare of Atlanta
    • Illinois
      • Chicago, Illinois, Estados Unidos, 60611
        • Ann and Robert H. Lurie Children's Hospital of Chicago
    • Massachusetts
      • Boston, Massachusetts, Estados Unidos, 02115
        • Boston Children's Hospital
    • Michigan
      • Ann Arbor, Michigan, Estados Unidos, 48109
        • University of Michigan - Mott Children's Hospital
    • New York
      • New York, New York, Estados Unidos, 10032
        • Columbia University Irving Medical Center/New York Presbyterian Morgan Stanley Children's Hospital
    • North Carolina
      • Durham, North Carolina, Estados Unidos, 27703
        • Duke University
    • Ohio
      • Columbus, Ohio, Estados Unidos, 43205
        • Nationwide Children's Hospital
    • Pennsylvania
      • Philadelphia, Pennsylvania, Estados Unidos, 19104
        • Children's Hospital of Philadelphia
    • Utah
      • Salt Lake City, Utah, Estados Unidos, 84113
        • University of Utah - Intermountain Primary Children's Hospital
    • Washington
      • Seattle, Washington, Estados Unidos, 98105
        • Seattle Children's Hospital

Critérios de participação

Os pesquisadores procuram pessoas que se encaixem em uma determinada descrição, chamada de critérios de elegibilidade. Alguns exemplos desses critérios são a condição geral de saúde de uma pessoa ou tratamentos anteriores.

Critérios de elegibilidade

Idades elegíveis para estudo

1 ano a 16 anos (Filho)

Aceita Voluntários Saudáveis

Não

Gêneros Elegíveis para o Estudo

Tudo

Descrição

Inclusion Criteria:

Candidates must meet all of the following inclusion criteria to be considered for enrollment in this study.

  1. Age 18 months to 16 years.

  2. Male or female.

  3. Subject has a native or repaired right ventricular outflow tract.

  4. Subject has been recommended for surgical pulmonary valve replacement by treating clinical team (cardiologist and cardiac surgeon).

  5. Subject has at least one of the following echocardiographic findings:

    1. Severe pulmonary stenosis (defined as RV to PA peak instantaneous gradient ≥60 mmHg);
    2. Moderate or greater pulmonary regurgitation;
    3. Moderate or greater pulmonary stenosis plus moderate or greater pulmonary regurgitation.
  6. Subject's body size is suitable for implantation of a study device ranging from 12.7 to 22 mm (internal diameter).

  7. Subject and parent/legal representative, where appropriate, are willing to provide informed written consent.

  8. Subject and parent/legal representative, where appropriate, and treating physician agree that the subject will return for, and comply with, all required study assessments and follow-up visits.

Exclusion Criteria:

Candidates will be excluded from the study if any of the following conditions are met:

  1. Subject requires valve replacement in a non-pulmonary position.
  2. Subject has a prosthetic valve at other valve position or will need a prosthetic valve at other valve position (i.e., anticipate additional valve replacements needed within 3 years).
  3. Subject has pulmonary arterial hypertension (defined as mean PA pressure ≥25 mmHg).
  4. Subject has pulmonary atresia and major aortopulmonary collaterals.
  5. Subject has significant peripheral pulmonary artery stenosis.
  6. Subject has an active infection requiring current systemic antibiotic therapy (if temporary illness, subject may be a candidate 4 weeks after discontinuation of antibiotics).
  7. Subject has active endocarditis or a history of infective endocarditis.
  8. Subject has renal insufficiency as determined by a serum creatinine (S-Cr) level ≥2.5 mg/dL within 60 days prior to the Screening Visit, or has end-stage renal disease.
  9. Subject has leukopenia (defined as a white blood cell (WBC) count <3.5 x 103/µL)
  10. Subject has acute or chronic anemia (defined as hemoglobin (Hgb) <10.0 g/dl or 6 mmol/L).
  11. Subject has thrombocytopenia (defined as platelet count <50 x 103/µL.
  12. Subject has a known hypersensitivity to anticoagulants and antiplatelet drugs.
  13. Subject has a known autoimmune disease or receives immunosuppressant and/or immunostimulant drugs that the Investigator or Eligibility Screening Committee (ESC) believes may negatively affect study outcomes.
  14. Subject needs emergency cardiac or vascular surgery or intervention.
  15. Positive pregnancy test prior to valve implant procedure in female subjects who have had their first menses.
  16. Subject has, in the opinion of the Investigator, a life expectancy of less than 5 years.
  17. Subject or parent/legal representative refuses blood transfusions.
  18. Subject has medical, social or psychosocial factors that, in the opinion of the Investigator or ESC, could impact safety or compliance with study procedures.
  19. Subject is participating in an investigational study of a new drug, biologic or device at the time of study screening.

Plano de estudo

Esta seção fornece detalhes do plano de estudo, incluindo como o estudo é projetado e o que o estudo está medindo.

Como o estudo é projetado?

Detalhes do projeto

  • Finalidade Principal: Tratamento
  • Alocação: N / D
  • Modelo Intervencional: Atribuição de grupo único
  • Mascaramento: Nenhum (rótulo aberto)

Armas e Intervenções

Grupo de Participantes / Braço
Intervenção / Tratamento
Experimental: Autus Valve Arm

Pulmonary Valve Replacement Surgery with the Autus Valve

Pulmonary valve replacement surgery with the Autus Valve

O que o estudo está medindo?

Medidas de resultados primários

Medida de resultado
Descrição da medida
Prazo
Valve Implantation Primary Safety Endpoint
Prazo: 30 Days post-valve implantation

Composite Endpoint: Freedom from a device-related complication through 30 days post-valve implantation as adjudicated by an independent Clinical Events Committee (CEC), including:

  • Death;
  • Valve thrombosis requiring intervention;
  • Symptomatic thromboembolism.
30 Days post-valve implantation
Valve Implantation Primary Effectiveness Endpoint
Prazo: 6 Months post-valve implantation

Acceptable hemodynamic performance at 6 months post-valve implantation, as assessed by the Echocardiography Core Laboratory or Clinical Events Committee, defined as:

  • A mean RVOT gradient less than or equal to 40 mmHg and
  • Less than moderate pulmonary regurgitation by transthoracic echo and
  • No Autus Valve reintervention except for valve expansion to address prosthesis-patient mismatch.
6 Months post-valve implantation

Outras medidas de resultado

Medida de resultado
Descrição da medida
Prazo
Other Safety Outcome - Composite
Prazo: 6 months and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following other safety outcome will be evaluated at 6 months, and annually through 10 years post-valve implantation by an independent Clinical Events Committee. The composite outcome includes device-related:

  • Death;
  • Valve thrombosis requiring intervention;
  • Symptomatic thromboembolism.
6 months and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Mortality
Prazo: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Mortality, including: all cause, cardiac-related, procedure-related and device-related

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Valve Thrombosis
Prazo: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Valve thrombosis

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Thromboembolism
Prazo: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Thromboembolism

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Cardiac or Valve Intervention
Prazo: 30 days, 6 months, and annually through a minimum of 510 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Other cardiac or valve intervention (surgical or transcatheter)

30 days, 6 months, and annually through a minimum of 510 years and up to 11 years post-valve implantation
Other Safety Outcome - Major Hemorrhage
Prazo: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Major hemorrhage

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Endocarditis
Prazo: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Endocarditis

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Hemolysis
Prazo: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Hemolysis

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Device Explantation
Prazo: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Device explantation

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Adverse Events (AE)
Prazo: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be reported at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Adverse events

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Valve/Stent Fracture
Prazo: pre-discharge, 12 months, and annually through a minimum of 5 years and up to 6 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-discharge and at 12 months post-valve expansion.

An independent Medical Reviewer will review post-implant chest x-rays to assess for valve/stent fracture. Chest x-rays will be performed at the following timepoints: post-valve implantation at hospital discharge, and annually through 5 years post-valve implantation. If a post-implant valve expansion procedure is performed, chest x-rays will be performed at hospital discharge and 12 months post-procedure.

pre-discharge, 12 months, and annually through a minimum of 5 years and up to 6 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-discharge and at 12 months post-valve expansion.
Other Effectiveness Outcome - Valve Implant Acute Procedural Success
Prazo: Immediately Post-Valve Implant Procedure

To meet the definition of success, all the parameters listed below must be achieved, as assessed by Sites using direct pressure measurements and/or transesophageal echo (TEE):

  • RV to PA peak systolic pressure gradient <40 mmHg (direct pressure measurement);
  • None or trivial pulmonary regurgitation (assessed via TEE);
  • No paravalvular leak (assessed via TEE).
Immediately Post-Valve Implant Procedure
Other Effectiveness Outcome - Pulmonary Regurgitation Grade
Prazo: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Pulmonary regurgitation grade categorized as none, trivial, mild, moderate, severe as evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Pulmonary Stenosis Grade
Prazo: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Pulmonary stenosis grade categorized as none/trivial, mild, moderate, severe, as evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Right Ventricular Measures
Prazo: Baseline, 30 days, 6 months, and annually through 5 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Right ventricular (RV) measures, including RV dimension and strain will be evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 5 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Left Ventricular Measures
Prazo: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Left ventricular (LV) measures, including LV ejection fraction and strain will be evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Modified Ross Functional Classification
Prazo: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Modified Ross Functional Class as assessed by the Site Investigators

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - PedsQL - Cardiac Module
Prazo: Baseline, 6 months and annually through 10 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-valve expansion, and at 6 and 12 months post-valve expansion.

The Pediatric Qualify of Life Inventory (PedsQL) Cardiac Module will be administered by Site Personnel.

Baseline, 6 months and annually through 10 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-valve expansion, and at 6 and 12 months post-valve expansion.
Other Post-Implant Valve Expansion Safety Outcome - Composite
Prazo: 30 days after each post-implant valve expansion procedure

Composite Endpoint: Freedom from device-related complication through 30 days post-valve expansion as adjudicated by an independent CEC:

  • Death;
  • Extracorporeal membrane oxygenation (ECMO) support;
  • Unplanned intervention (surgical or transcatheter);
  • Major hemorrhage
30 days after each post-implant valve expansion procedure
Other Post-Implant Valve Expansion Effectiveness Outcome - Acute Procedural Success
Prazo: Immediately after each post-implant valve expansion procedure

To meet the definition of success, at least one of the parameters listed below must be achieved after each post-implant valve expansion is performed, assessed during the same catheterization procedure:

  • RV to aortic systolic pressure ratio: ≤0.5.
  • Peak-to-peak RV to PA systolic pressure gradient: ≤15 mmHg;
  • Peak-to-peak RV to PA systolic pressure gradient: ≥50% reduction (pre vs post valve expansion).
Immediately after each post-implant valve expansion procedure
Other Post-Implant Valve Expansion Effectiveness Outcome - Hemodynamic Performance
Prazo: 6 months after each post-implant valve expansion procedure

Clinically acceptable hemodynamic performance at 6 months post-valve expansion procedure, as assessed by the Echo Core Laboratory via TTE, defined as:

  • Less than moderate pulmonary regurgitation (PR) by Doppler echocardiography;
  • Transvalvular gradient (assessed by CW Doppler) at 6-months is lower than the pre-expansion gradient.
6 months after each post-implant valve expansion procedure

Colaboradores e Investigadores

É aqui que você encontrará pessoas e organizações envolvidas com este estudo.

Patrocinador

Publicações e links úteis

A pessoa responsável por inserir informações sobre o estudo fornece voluntariamente essas publicações. Estes podem ser sobre qualquer coisa relacionada ao estudo.

Datas de registro do estudo

Essas datas acompanham o progresso do registro do estudo e os envios de resumo dos resultados para ClinicalTrials.gov. Os registros do estudo e os resultados relatados são revisados ​​pela National Library of Medicine (NLM) para garantir que atendam aos padrões específicos de controle de qualidade antes de serem publicados no site público.

Datas Principais do Estudo

Início do estudo (Real)

5 de fevereiro de 2024

Conclusão Primária (Real)

24 de agosto de 2026

Conclusão do estudo (Estimado)

5 de fevereiro de 2036

Datas de inscrição no estudo

Enviado pela primeira vez

23 de julho de 2021

Enviado pela primeira vez que atendeu aos critérios de CQ

12 de agosto de 2021

Primeira postagem (Real)

16 de agosto de 2021

Atualizações de registro de estudo

Última Atualização Postada (Real)

20 de outubro de 2026

Última atualização enviada que atendeu aos critérios de controle de qualidade

15 de outubro de 2026

Última verificação

1 de outubro de 2026

Mais Informações

Termos relacionados a este estudo

Outros números de identificação do estudo

  • AUT-CP-001

Plano para dados de participantes individuais (IPD)

Planeja compartilhar dados de participantes individuais (IPD)?

NÃO

Informações sobre medicamentos e dispositivos, documentos de estudo

Estuda um medicamento regulamentado pela FDA dos EUA

Não

Estuda um produto de dispositivo regulamentado pela FDA dos EUA

Sim

Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .