Comparison of NeuroEXPLORER PET/CT Versus Standard-of-care PET/CT of the Head and Neck Region.

14 мая 2026 г. обновлено: prof. dr. Koen Van Laere

Prospective Comparison of Ultra-high Resolution NeuroExplorer PET/CT Versus Standard-of-care PET/CT of the Head and Neck Region

The goal of this clinical study is to compare the current standard-of-care PET/CT scanner with the new NeuroEXPLORER PET/CT, with the aim of demonstrating the added diagnostic value of the new scanner. As this is the first device of its kind in Europe, the study will also evaluate the safety and general feasibility of the NeuroEXPLORER system for head and neck PET studies.

Due to to its high resolution, the NeuroEXPLORER enables an accurate investigation of areas up to 20 times smaller than what can be visualized with standard clinical PET scanners, providing images with a very high level of detail. This scanner is specifically designed for imaging the brain, spinal cord, and neck region, to support the diagnosis and monitoring of neurological, psychiatric, and other disorders in the head and neck.

The main questions this study aims to answer are:

  • Does the NeuroEXPLORER provide more detailed information with better diagnostic quality compared to standard clinical PET/CT scans for head and neck disorders?
  • Is the NeuroEXPLORER system safe and effective in performing head and neck studies?

Participants will:

  • receive clear information about the study, including its organization, possible risks, and potential benefits. Based on this, they will be able to give their informed consent by signing an informed consent form before the study begins.
  • visit the clinic (UZ Leuven) once for two consecutive PET/CT scans: first the prescribed clinical scan, followed by a shorter scan with the NeuroEXPLORER.
  • before the start of the regular PET/CT scan, have a small catheter placed in a blood vessel in the forearm. Through this, a tracer will be administered. A tracer is a small amount of slightly radioactive material that helps to produce images of processes inside the body. Depending on the tracer used, the PET/CT scan may start immediately or after a short waiting period.
  • lie on their back on the scanner bed for 10 to 30 minutes during the clinical scan.
  • after the clinical scan, undergo a 15-20 minute scan on the NeuroEXPLORER.
  • leave the hospital after completing both scans.
  • afterwards, be asked whether they experienced any discomfort during the study. In addition, they will receive a one-time follow-up phone call within a week after the scan to check if they experienced any discomfort in the days following the study.

Обзор исследования

Статус

Запись по приглашению

Условия

Вмешательство/лечение

Тип исследования

Интервенционный

Регистрация (Оцененный)

1600

Фаза

  • Непригодный

Контакты и местонахождение

В этом разделе приведены контактные данные лиц, проводящих исследование, и информация о том, где проводится это исследование.

Места учебы

    • Vlaams-Brabant
      • Leuven, Vlaams-Brabant, Бельгия, 3000
        • UZ Leuven

Критерии участия

Исследователи ищут людей, которые соответствуют определенному описанию, называемому критериям приемлемости. Некоторыми примерами этих критериев являются общее состояние здоровья человека или предшествующее лечение.

Критерии приемлемости

Возраст, подходящий для обучения

  • Ребенок
  • Взрослый
  • Пожилой взрослый

Принимает здоровых добровольцев

Да

Описание

Inclusion Criteria:

  • All subjects referred for standard-of-care PET in the indications for the substudies are eligible.

Exclusion Criteria:

  • Subject is unwilling to avoid unusual, unaccustomed, or strenuous physical activity beginning 4 days prior to tracer injection up to 1 day after tracer injection.

    • Subject does not understand the study procedures.
    • Subject is unwilling or unable to perform all of the study procedures or is considered unsuitable in any way by the principal investigator.
    • Subject has had exposure to ionizing radiation (> 1 mSv) in other research studies within the last 12 months.
    • Subject does not agree that incidental findings are communicated to the general practitioner and to the participant him/herself.
    • Incapacitated patients that cannot give consent by themselves (in particular for substudies 1-3 in neurodegeneration).

Учебный план

В этом разделе представлена ​​подробная информация о плане исследования, в том числе о том, как планируется исследование и что оно измеряет.

Как устроено исследование?

Детали дизайна

  • Основная цель: Диагностика
  • Распределение: Н/Д
  • Интервенционная модель: Одногрупповое задание
  • Маскировка: Нет (открытая этикетка)

Оружие и интервенции

Группа участников / Армия
Вмешательство/лечение
Экспериментальный: single arm: PET/CT comparison
All participants undergo a scan on both PET/CT systems. First the Standard-of-care PET/CT and immediately after a scan on the NeuroEXPLORER.
This PET/CT is the routinely used clinical PET/CT scan
Другие имена:
  • GE MI4 PET/CT
  • Siemens Vision X PET/CT
This is the new ultra-high resolution PET/CT scan

Что измеряет исследование?

Первичные показатели результатов

Мера результата
Мера Описание
Временное ограничение
Difference in diagnostic accuracy and/or diagnostic confidence of ultra-high resolution PET/CT compared to standard PET/CT and validated against clinical follow-up, or pathological/surgical outcomes in neurodegenerative disorders.
Временное ограничение: From enrollment until two weeks after the scan
For the neurodegenerative indications (n=1000 subjects), standard-of-care (SOC) and ultra-high-resolution (UHR) PET images will be interpreted by three blinded nuclear medicine physicians using standardized MIM software workflows. Standard clinical criteria by the most recent EANM/SNMMI guidelines will be used for interpretation, supplemented by predefined new features unique to UHR images. Quantitative metrics, including standardized uptake values ratio (SUVR) and z-scores will be recorded in cortical regions and small nuclei of interest by semiquantitative region-based analysis based on a UHR brain atlas and normative database. Standard of truth will be the final clinical diagnosis established by expert memory clinical specialists at least 6 months after PET, based on MRI, SOC PET and fluid biomarkers
From enrollment until two weeks after the scan
Difference in diagnostic accuracy and/or diagnostic confidence of ultra-high resolution PET/CT compared to standard PET/CT and validated against clinical follow-up, or pathological/surgical outcomes in head and neck cancer.
Временное ограничение: From enrollment until two weeks after study visit
Lesion detection/delineation and localization performance of the UHR vs. SOC PET will be compared by three blinded expert readers. Images will be co-registered with MRI where available for anatomical precision. Image quantification (semi-quantitative analysis) will include measurement of target-to-background ratio and lesion SUVmax. Standard of truth will be the combination of surgical/histopathological findings and biochemical follow-up.
From enrollment until two weeks after study visit
Comparison of the sensitivity, specificity and accuracy of ultra-high resolution PET vs. standard-of-care PET for the diagnosis of vasculitis and giant cell arteritis, using clinical and pathological confirmation as the benchmark.
Временное ограничение: From enrollment until two weeks after study visit
UHR and SOC PET scans will be analyzed by three nuclear medicine physicians. Vascular FDG uptake will be assessed by scoring the presence and intensity of uptake in various arterial segments. Semiquantitative metrics like SUVmax or target-to-background ratio will be obtained. Standard of truth will be the combination of clinical criteria and pathological biopsy confirmation.
From enrollment until two weeks after study visit

Вторичные показатели результатов

Мера результата
Мера Описание
Временное ограничение
Safety and feasibility of ultra-high resolution PET/CT (MDR generic masterprotocol EUDAMED CIV-25-06-053398)
Временное ограничение: From enrollment until 1 week after the scan
Any device specific adverse events (ADE or SADE), including performance parameters will be tracked. PROM (patient reported outcome measures) and patient tolerability will be monitored by a questionnaire taken immediately and one week (phone call) after the study investigation.
From enrollment until 1 week after the scan
Quality of life (QOL) questionnaire
Временное ограничение: From enrollment until one year after the study visit
Change in health-related quality of life will be evaluated using the EQ-5D-5L (EuroQol) questionnaire. The questionnaire will be administered at baseline and at one year after the study visit.
From enrollment until one year after the study visit
Impact on decision-making
Временное ограничение: From enrollment until 1 year after study visit
Clinical impact will be quantified by the absolute number and proportion of cases in which UHR PET leads to a change in diagnosis, diagnostic confidence, management recommendation, lesion detection, lesion localization, or reader-rated clinical impact compared with standard-of-care PET.
From enrollment until 1 year after study visit

Соавторы и исследователи

Здесь вы найдете людей и организации, участвующие в этом исследовании.

Спонсор

Соавторы

Публикации и полезные ссылки

Лицо, ответственное за внесение сведений об исследовании, добровольно предоставляет эти публикации. Это может быть что угодно, связанное с исследованием.

Даты записи исследования

Эти даты отслеживают ход отправки отчетов об исследованиях и сводных результатов на сайт ClinicalTrials.gov. Записи исследований и сообщаемые результаты проверяются Национальной медицинской библиотекой (NLM), чтобы убедиться, что они соответствуют определенным стандартам контроля качества, прежде чем публиковать их на общедоступном веб-сайте.

Изучение основных дат

Начало исследования (Действительный)

30 сентября 2025 г.

Первичное завершение (Оцененный)

31 августа 2030 г.

Завершение исследования (Оцененный)

31 августа 2030 г.

Даты регистрации исследования

Первый отправленный

27 апреля 2026 г.

Впервые представлено, что соответствует критериям контроля качества

14 мая 2026 г.

Первый опубликованный (Действительный)

19 мая 2026 г.

Обновления учебных записей

Последнее опубликованное обновление (Действительный)

19 мая 2026 г.

Последнее отправленное обновление, отвечающее критериям контроля качества

14 мая 2026 г.

Последняя проверка

1 мая 2026 г.

Дополнительная информация

Термины, связанные с этим исследованием

Дополнительные соответствующие термины MeSH

Другие идентификационные номера исследования

  • CIV-25-06-053414

Планирование данных отдельных участников (IPD)

Планируете делиться данными об отдельных участниках (IPD)?

ДА

Информация о лекарствах и устройствах, исследовательские документы

Изучает лекарственный продукт, регулируемый FDA США.

Нет

Изучает продукт устройства, регулируемый Управлением по санитарному надзору за качеством пищевых продуктов и медикаментов США.

Нет

Эта информация была получена непосредственно с веб-сайта clinicaltrials.gov без каких-либо изменений. Если у вас есть запросы на изменение, удаление или обновление сведений об исследовании, обращайтесь по адресу register@clinicaltrials.gov. Как только изменение будет реализовано на clinicaltrials.gov, оно будет автоматически обновлено и на нашем веб-сайте. .