Does the VTS1000 Infra Red Vein Imager Allow Better Identification of Veins Than Normal Eyesight? (VFvalidation)
Validation of Vue Tek Digital Vein Viewing Device
VueTek has developed a head mounted device that will improve the viewing of superficial, subcutaneous vasculature, differentiating it with higher contrast from surrounding tissue. Video of the vasculature is displayed to the user in a portable headset. The headset and display combination facilitates both visualization in the display and an open field of view directly to the subject's anatomy and practitioner's surroundings. The goal is to provide a non-diagnostic, enhanced visual aide that is used alongside standard visual and tactile methodology. The primary and secondary objectives of this validation are as follows:
Primary Objective:
To demonstrate that the VTS1000 affords visualization of superficial, subcutaneous vascular structures, when compared to the naked eye.
- Secondary Objective:
To demonstrate that the VTS1000 enhances visualization of superficial, subcutaneous vascular structures over the naked eye.
Validation Ancillary Endpoints
1. To gather data to demonstrate that the VTS1000 is portable, affords conventional vascular access methods, and provides flexibility, fit and balance during use.
Study Overview
Status
Status
Conditions
Conditions
Detailed Description
Validation methodology:
Ten (10) or more independent qualified practitioners trained to operate the device in accordance with the Directions For Use (DFU), and after completing the Device Usability aspect of the study, will be asked to perform simulated procedures with and without the device as indicated in the DFU on one-hundred (100) or more subjects using the following subject population and sampling table. A minimum of one hundred (100) or more data collection observations will be recorded, tabulated and reported.
Assumptions: The population and sampling sizes chosen for each subject category will meet the goal to attain sufficient diversity to satisfy representation that supports the efficacy of the device across the population samples. Any failures or results from the acceptance criteria may require an increase in the sampling size to ensure coverage representation of all subject demographics.
Study Type
Study Type
Enrollment (Actual)
Enrollment
Contacts and Locations
Study Locations
-
-
New York
-
New York, New York, United States, 10019
- St. Luke's Roosevelt Hospitals
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
- Inclusion Criteria: All subjects will be automatically enrolled on a walk-in basis.
- Exclusion Criteria: No walk-in subjects will be excluded from enrollment.
Study Plan
How is the study designed?
Design Details
Number of groups / cohorts
Cohorts and Interventions
Group / CohortGroup / Cohort |
|---|
|
Normal vision
This within-subjects design is counter balanced for order.
Each operator sees the subject's arm either initially with the normal eye or with the device, then repeats with the second condition (i.e., device or normal vision).
|
|
Vein Imaging Device
This within-subjects design is counter balanced for order.
Each operator sees the subject's arm either initially with the normal eye or with the device, then repeats with the second condition (i.e., device or normal vision).
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
To demonstrate that the VTS1000 affords visualization of superficial, subcutaneous vascular structures, when compared to the naked eye.
Time Frame: Immediate
|
Ten (10) or more independent qualified practitioners trained to operate the device in accordance with the Directions For Use (DFU), and after completing the Device Usability aspect of the study, will be asked to perform simulated procedures with and without the device as indicated in the DFU on one-hundred (100) or more subjects using the following subject population and sampling table.
A minimum of one hundred (100) or more data collection observations will be recorded, tabulated and reported.
|
Immediate
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
To demonstrate that the VTS1000 enhances visualization of superficial, subcutaneous vascular structures over the naked eye.
Time Frame: Immediate
|
Ten (10) or more independent qualified practitioners trained to operate the device in accordance with the Directions For Use (DFU), and after completing the Device Usability aspect of the study, will be asked to perform simulated procedures with and without the device as indicated in the DFU on one-hundred (100) or more subjects using the following subject population and sampling table.
Number of veins identified by normal vision is compared to number of veins identified with the device in a counter balanced design.
|
Immediate
|
Collaborators and Investigators
Sponsor
Sponsor
Collaborators
Collaborators
Investigators
Investigators
- Principal Investigator: Franco Resta-Flarer, M.D., St. Luke's Roosevelt Hospitals
Study record dates
Study Major Dates
Study Start
Study Start
Primary Completion (ACTUAL)
Primary Completion
Study Completion (ACTUAL)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (ESTIMATE)
First Posted
Study Record Updates
Last Update Posted (ESTIMATE)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
Other Study ID Numbers
- VF Validation
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Vision
-
NCT06602102CompletedLow Vision | Low Vision Aids | Low Vision Blindness | Low Vision Digital Assistance
-
NCT04004507CompletedDecrease in Night Vision | Disturbance; Vision, Loss
-
NCT04919837RecruitingArtificial Intelligence | Ophthalmology | Low Vision Aids
-
NCT06407336Active, not recruiting
-
NCT01703559CompletedNight Vision Complaints | Decrease in Night Vision | Disturbance; Vision, Loss
-
NCT05184036Active, not recruitingLow Vision Digital Assistance
-
NCT06477809Recruiting
-
NCT05917457Completed
-
NCT04892316RecruitingArtificial Intelligence | Ophthalmology | Low Vision Aids