Three Intraocular Lenses Designed to Improve Distance, Intermediate and Near Vision Following Lens Extraction
A Three Arm Prospective Clinical Evaluation of Three FDA-approved Intraocular Lenses Designed to Improve Distance, Intermediate and Near Vision Following Lens Extraction
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Phase 4
Contacts and Locations
Study Locations
-
-
-
Makati City, Philippines, 1200
- Asian Eye Institute
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Subjects must be undergoing primary IOL implantation for the correction of aphakia following continuous curvilinear anterior capsulotomy and phacoemulsification cataract extraction.
- Subjects must require a spherical lens power from 10.00 diopters (D) to 33.00 D.
- Subjects must have the potential for corrected distance visual acuity (CDVA) of 20/32 or better in both eyes.
- Subjects must have stopped contact lens wear for at least two weeks for soft lens wearers or three weeks for gas permeable lens wearers prior to biometry and surgery.
- At the time of surgery, subjects must have an intact centered capsulorhexis, intact posterior capsule, and no zonular rupture.
- Subjects must have ≤ 1.25 D of preoperative corneal astigmatism in both eyes.
Exclusion Criteria:
- Subjects with diagnoses of degenerative visual disorders (eg, macular degeneration, or other retinal disorders) that cause potential acuity losses to a level of 20/32 or worse.
- Subjects with conditions with increased risk of zonular rupture, such as pseudoexfoliation syndrome.
- Subjects who have any active inflammation or edema (swelling) of the cornea, including but not limited to the following: keratitis, keratoconjunctivitis, and keratouveitis.
- Subjects with uncontrolled glaucoma.
- Subjects with previous retinal detachment.
- Subjects with visually significant diabetic retinopathy (proliferative or nonproliferative) which reduces potential acuity to 20/32 or worse.
- Subjects with rubella, bilateral congenital, traumatic, complicated or polar cataract.
- Subjects with marked microphthalmos or aniridia.
- Subjects who have had previous corneal surgery.
- Subjects with irregular corneal astigmatism.
- Subjects with amblyopia which reduces potential acuity to 20/32 or worse.
- Subjects with optic atrophy.
- Subjects with iris neovascularization.
- Subjects with clinically significant retinal pigment epithelium/macular changes which reduces potential acuity to 20/32 or worse.
- Subjects with chronic use of systemic steroids or immunosuppressive medications.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Randomized
- Interventional Model: Parallel Assignment
- Masking: Double
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Crystalens AO
Bausch & Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
|
Bausch & Lomb model silicone multi-piece accommodating IOL.
|
|
Active Comparator: ReSTOR 3.0
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
|
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
|
|
Active Comparator: AMO Tecnis Multifocal
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
|
The Tecnis Multifocal foldable hydrophobic acrylic IOL, Model ZMA00 (Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Monocular Mesopic Contrast Sensitivity Without Glare (1.5, 3 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Monocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
In two eye (Binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Monocular Mesopic Contrast Sensitivity With Glare (3 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree)
Time Frame: Postoperative visit 3 (day 30-60)
|
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 3 (day 30-60)
|
|
Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Monocular Mesopic Contrast Sensitivity With Glare(3 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
|
Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree)
Time Frame: Postoperative visit 4 (day 120-180)
|
In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness).
The participant's ability to detect changes in contrast is determined.
Higher mean indicates improved contrast sensitivity.
|
Postoperative visit 4 (day 120-180)
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Study Director: Tes Ignacio, MD, Bausch & Lomb Incorporated
Study record dates
Study Major Dates
Study Start
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Estimate)
First Posted
Study Record Updates
Last Update Posted (Estimate)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- 656 (Other Identifier: Cardiothoracic Surgical Trials Network)
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.