Heidelberg Edge Perimetry (HEP) Detecting Glaucomatous Visual Field Defects (HEP)
Validation and Reproducibility of the Heidelberg Edge Perimeter in the Detection of Visual Field Defects in Glaucomatous Patients
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Each participant undergo the following tests:
- Standard ophthalmic examination (usual care): Best-corrected visual acuity, biomicroscopy (look at front of eye), intraocular pressure (IOP) using Goldman applanation tonometry, central corneal thickness measurement and fundoscopy (look at back of eye).
- Octopus Visual Field (OVF) Analyzer using G-top strategy recording mean deviation (MD) and pattern standard deviation (PSD).
- Heidelberg Edge Perimeter (HEP) standard automated perimetry (SAP) using advanced staircase thresholding algorithm (ASTA).
- Spectralis optical coherence tomography (OCT) using glaucoma module premium edition (GMPE) software to measure minimum rim width (MRW) and retinal nerve fiber layer (RNFL) thickness in 6 sectors (superior nasal, superior temporal, inferior nasal, inferior temporal, nasal and temporal) in the retina.
Order of HEP SAP III and OVF, and order of the eye tested (right vs. left) if both eyes of patients meet inclusion criteria will be randomized. Patients with glaucoma will be randomly chosen and asked to return in 3 to 6 months for repeat HEP and OVF testing.
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Age > 18 years.
- Best corrected visual acuity of 20/40 or better in the tested eye.
- Spherical refraction within ±5.0 D, and cylinder correction within ±3.0 D.
- History of glaucoma, defined as:
- Characteristic of glaucomatous disc damage: (local narrowing, notching, or absence of the neuroretinal rim in the absence of disc pallor elsewhere).
Exclusion Criteria:
- Any condition preventing adequate examination of the pupil or visual field testing (e.g. ptosis, dense corneal opacities or lens opacities);
- Active infection of the anterior or posterior segments of the eye;
- Any intraocular surgical or laser procedure within the previous 4 weeks;
- Participants taking a medication known to affect visual field sensitivity, a coexisting intraocular disease affecting visual field, or a problem other than glaucoma affecting color vision will be excluded.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: Randomized
- Interventional Model: Parallel Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Glaucoma Subjects
Subjects with glaucoma were recruited based on characteristic glaucomatous disc damage and visual field changes.
They will perform visual field with Heidelberg Edge Perimeter and Octopus visual field.
Optical Coherence Tomography will image the retinal nerve fiber layer.
|
HEP measures peripheral vision
Other Names:
OVF measures peripheral vision
Other Names:
OCT measures retinal nerve fiber layer thickness
Other Names:
|
|
Experimental: Healthy Controls
Subjects that do not have glaucoma and are recruited for testing will perform visual field with Heidelberg Edge Perimeter and Octopus visual field.
Optical Coherence Tomography will image the retinal nerve fiber layer.
|
HEP measures peripheral vision
Other Names:
OVF measures peripheral vision
Other Names:
OCT measures retinal nerve fiber layer thickness
Other Names:
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Correlation Coefficient Between HEP and OVF Mean Deviation (MD)
Time Frame: Baseline visit, 1 hour
|
Pearson's correlation coefficient between Heidelberg Edge Perimeter (HEP) and Octopus Visual Field (OVF) Mean Deviation (MD) for glaucoma patients and controls to determine if HEP can detect glaucoma as well as OVF.
The closer the values of both parameters for both machines, the better comparable the two machines are to each other in detecting glaucoma.
|
Baseline visit, 1 hour
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Repeatability of Optical Coherence Tomography (OCT) Parameters
Time Frame: Month 6 visit, 1 hour
|
A subgroup of participants were randomly selected to return at 3 and 6 months for repeat testing.
Not all eyes were included.
Repeatability of optical coherence tomography (OCT) to consistently detect minimum rim width (MRW) global thickness will be assessed by Pearson's interclass correlation coefficients (ICC).
A larger ICC indicates measurements have greater repeatability.
Greater than 0.75 indicated excellent repeatability; 0.40 to 0.75 indicated fair to good, and less than 0.40 indicated poor reliability.
|
Month 6 visit, 1 hour
|
|
Correlations Between Structure-function Automated Report and Clinical Impression
Time Frame: Baseline visit, 1 hour
|
Heidelberg Edge Perimeter (HEP) Visual Field (VF) and spectral domain optical coherence tomography (SD OCT) printouts from only the glaucoma subgroup were assessed.
Retinal Nerve Fiber Layer (RNFL) and Minimum Rim Width (MRW) from SD OCT and HEP VF automated reports were compared to the clinical interpretations by 3 glaucoma specialists.
|
Baseline visit, 1 hour
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Leslie J Katz, MD, Wills Eye Hospital
Publications and helpful links
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Estimate)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- 13-340
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.