Pressure-Volume Measurements on the AMS 800TM Cuff (AUS)
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
The current AMS 800 is not personalized to the patient's needs. Most men only need 20 cm H2O of urethral compression to remain dry when they are sedentary, yet need 120 cm H2O of urethral compression when they are active. The current device compromises with an in between pressure - as high as it can be without exceeding safety tolerance thresholds. As such many men still leak when they are active. In fact, this compromise is so central to the device that a new term has evolved to describe men's suboptimal level of continence after the AMS 800. We call it "social continence", meaning that they are dry enough to only have to wear 1-2 pads per day for protection.
The market is hungry for a device that can adapt to the patient's level of activity, reducing the pressure most of the day to protect the urethra and then briefly increasing the pressure when he is more active.
The experimental procedure is explained below and care will be taken to make sure the pressure sensor is sterile. The experiment involves minimal risk and will be conducted by the PI who has 10+ years of experience in AUS implantations.
The procedure for implanting the AMS 800 is mentioned in the Operating Manual attached with this protocol. The treatment procedure is linked with the experimental setup designed by the PI and the research student in the following manner:
- Follow surgical preparation steps mentioned on pages 14-15 of the AMS 800 Operating Manual. Once the cuff sizer is used to determine the cuff length (usually 4.0cm or 4.5cm), move on to next step. For graphical representation, follow steps 1 to 9b on pages 18- 20 of the AMS 800 Operating Manual.
- The AMS 800 cuff is filled with deionized water such that pressure is 20cmH2O. Follow steps on pages 13-14 of the AMS 800 Operating Manual for filling the cuff.
- The AMS 800 cuff around the urethra has a kink-free tubing which goes into a female connector
- The female connector connects to the male connector of the IV tubing
- The IV tubing is connected to a T-Connector
- One end of the T-Connector goes to the TruWave Pressure Transducer
- Other end goes to a syringe
- The system is pumped with deionized water and is therefore sterile.
- Water is pumped into the cuff with the syringe and as the volume in the cuff increases, corresponding cuff pressure changes. This is done until we have a volume for a pressure reading of 120cmH2O on the pressure transducer.
- Once the reading is obtained, the PI will flush out the excess water from the cuff, and fill it with an optimal volume as required by the treatment procedure which he has performed for over 10 years.
- The treatment procedure will continue to follow the steps hereafter as has been designed for AUS placement.
For all other details including device description and post-operative care, refer to the AMS 800 operating Manual. All devices including the AMS 800 (PMA-P000053) and the Edward LifeSciences Pressure Transducer (501K- K142749) are FDA approved.
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Minnesota
-
Minneapolis, Minnesota, United States, 55455
- University of Minnesota Medical Center
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Men between 18 and 100 years of age who are to undergo AUS placement for treatment of stress urinary incontinence.
Exclusion Criteria:
- Under the age of 18 years old English non-speaking Unable to consent for themselves If a trans-corporal technique to AUS implantation is planned as this may affect the pressure measurements (<10% of the AUSs we implant are transcorporal).
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Device Feasibility
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: AMS 800 Artificial Urinary Sphincter
The AMS 800™ Urinary Control System is an implantable, fluid-filled, solid silicone elastomer prosthesis used to treat urinary incontinence due to reduced outlet resistance (intrinsic sphincter deficiency) following prostate surgery. The AMS 800 Urinary Control System simulates normal sphincter function by opening and closing the urethra, under patient control. When the cuff is closed, urine stays in the bladder. When the patient wishes to void, he simply squeezes and releases the pump several times. This causes the fluid in the cuff to move into the pressure-regulating balloon . The cuff opens and urine passes through the urethra. The balloon then automatically re-pressurizes the cuff through the pump, within several minutes, the cuff again closes the urethra. The control pump, which in implanted in the scrotum, is also designed to allow the clinician or patient to deactivate and activate the system without additional surgery. |
AMS 800 Artificial Urinary Sphincter
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Pressure-Volume Measurements on the AMS 800TM Cuff
Time Frame: 6 months
|
Measure volume and corresponding pressure in AMS 800TM occlusion Cuff
|
6 months
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Sean Elliot, MD, University of Minnesota
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- 1703M09441
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Stress Urinary Incontinence
-
NCT03672461CompletedUrinary Incontinence, Stress | Urge Incontinence | Urinary Stress Incontinence | Stress Incontinence, Urinary | Stress Incontinence | Stress Incontinence, Female | Urgency Urinary
-
NCT02418299CompletedFemale Stress Urinary Incontinence | Mixed Incontinence, Urge and Stress
-
NCT05464316RecruitingUrinary Incontinence | Urinary Stress Incontinence | Post-Prostatectomy Incontinence | Stress Incontinence, Male
-
NCT05677295RecruitingWomen With Stress Urinary Incontinence
-
NCT02288455CompletedMale Stress Urinary Incontinence
-
NCT01903590CompletedFemale Urinary Stress Incontinence
-
NCT00744198Unknown
-
NCT01123135WithdrawnStress Urinary Incontinence
-
NCT00190567Completed
Clinical Trials on AMS 800
-
NCT04088331CompletedStress Urinary Incontinence
-
NCT02901392RecruitingUrinary Incontinence, Stress
-
NCT02490917Unknown
-
NCT03323554Active, not recruitingUrinary Incontinence
-
NCT01383018CompletedErectile Dysfunction | Impotence
-
NCT00638235CompletedPelvic Organ Prolapse
-
NCT00636506Completed
-
NCT01168830Completedde Novo Lesions in Native Coronary Arteries
-
NCT03561922TerminatedRetinal Degeneration | Inherited Retinal Dystrophy Primarily Involving Sensory Retina
-
NCT06110689CompletedCardiac Anomaly | Pectus Excavatum