Pressure-Volume Measurements on the AMS 800TM Cuff (AUS)

June 28, 2019 updated by: University of Minnesota
he idea of the experiment is to calculate the volume of fluid that is required to raise cuff pressure from 20 cmH2O to 120 cmH2O in the American Medical System's AMS 800 Artificial Urinary Sphincter (AUS).

Study Overview

Status

Completed

Intervention / Treatment

Detailed Description

The current AMS 800 is not personalized to the patient's needs. Most men only need 20 cm H2O of urethral compression to remain dry when they are sedentary, yet need 120 cm H2O of urethral compression when they are active. The current device compromises with an in between pressure - as high as it can be without exceeding safety tolerance thresholds. As such many men still leak when they are active. In fact, this compromise is so central to the device that a new term has evolved to describe men's suboptimal level of continence after the AMS 800. We call it "social continence", meaning that they are dry enough to only have to wear 1-2 pads per day for protection.

The market is hungry for a device that can adapt to the patient's level of activity, reducing the pressure most of the day to protect the urethra and then briefly increasing the pressure when he is more active.

The experimental procedure is explained below and care will be taken to make sure the pressure sensor is sterile. The experiment involves minimal risk and will be conducted by the PI who has 10+ years of experience in AUS implantations.

The procedure for implanting the AMS 800 is mentioned in the Operating Manual attached with this protocol. The treatment procedure is linked with the experimental setup designed by the PI and the research student in the following manner:

  1. Follow surgical preparation steps mentioned on pages 14-15 of the AMS 800 Operating Manual. Once the cuff sizer is used to determine the cuff length (usually 4.0cm or 4.5cm), move on to next step. For graphical representation, follow steps 1 to 9b on pages 18- 20 of the AMS 800 Operating Manual.
  2. The AMS 800 cuff is filled with deionized water such that pressure is 20cmH2O. Follow steps on pages 13-14 of the AMS 800 Operating Manual for filling the cuff.
  3. The AMS 800 cuff around the urethra has a kink-free tubing which goes into a female connector
  4. The female connector connects to the male connector of the IV tubing
  5. The IV tubing is connected to a T-Connector
  6. One end of the T-Connector goes to the TruWave Pressure Transducer
  7. Other end goes to a syringe
  8. The system is pumped with deionized water and is therefore sterile.
  9. Water is pumped into the cuff with the syringe and as the volume in the cuff increases, corresponding cuff pressure changes. This is done until we have a volume for a pressure reading of 120cmH2O on the pressure transducer.
  10. Once the reading is obtained, the PI will flush out the excess water from the cuff, and fill it with an optimal volume as required by the treatment procedure which he has performed for over 10 years.
  11. The treatment procedure will continue to follow the steps hereafter as has been designed for AUS placement.

For all other details including device description and post-operative care, refer to the AMS 800 operating Manual. All devices including the AMS 800 (PMA-P000053) and the Edward LifeSciences Pressure Transducer (501K- K142749) are FDA approved.

Study Type

Interventional

Enrollment (Actual)

8

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Minnesota
      • Minneapolis, Minnesota, United States, 55455
        • University of Minnesota Medical Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years to 100 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

Male

Description

Inclusion Criteria:

  • Men between 18 and 100 years of age who are to undergo AUS placement for treatment of stress urinary incontinence.

Exclusion Criteria:

  • Under the age of 18 years old English non-speaking Unable to consent for themselves If a trans-corporal technique to AUS implantation is planned as this may affect the pressure measurements (<10% of the AUSs we implant are transcorporal).

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Device Feasibility
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: AMS 800 Artificial Urinary Sphincter

The AMS 800™ Urinary Control System is an implantable, fluid-filled, solid silicone elastomer prosthesis used to treat urinary incontinence due to reduced outlet resistance (intrinsic sphincter deficiency) following prostate surgery. The AMS 800 Urinary Control System simulates normal sphincter function by opening and closing the urethra, under patient control. When the cuff is closed, urine stays in the bladder.

When the patient wishes to void, he simply squeezes and releases the pump several times. This causes the fluid in the cuff to move into the pressure-regulating balloon . The cuff opens and urine passes through the urethra. The balloon then automatically re-pressurizes the cuff through the pump, within several minutes, the cuff again closes the urethra.

The control pump, which in implanted in the scrotum, is also designed to allow the clinician or patient to deactivate and activate the system without additional surgery.

AMS 800 Artificial Urinary Sphincter

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Pressure-Volume Measurements on the AMS 800TM Cuff
Time Frame: 6 months
Measure volume and corresponding pressure in AMS 800TM occlusion Cuff
6 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Sean Elliot, MD, University of Minnesota

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

June 28, 2017

Primary Completion (Actual)

February 7, 2018

Study Completion (Actual)

March 1, 2018

Study Registration Dates

First Submitted

April 17, 2017

First Submitted That Met QC Criteria

April 17, 2017

First Posted (Actual)

April 20, 2017

Study Record Updates

Last Update Posted (Actual)

July 2, 2019

Last Update Submitted That Met QC Criteria

June 28, 2019

Last Verified

June 1, 2019

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Stress Urinary Incontinence

Clinical Trials on AMS 800

Subscribe