Efficacy and Safety of AEYE-DS Software Device for Automated Detection of Diabetic Retinopathy From Digital Fundus Images
Pivotal Prospective Clinical Trial to Demonstrate the Efficacy and Safety of AEYE-DS Software Device for Automated Diabetic Retinopathy Detection From Digital Fundoscopic Images
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Kentucky
-
Louisville, Kentucky, United States, 40206
- The Eye Care Institute
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Description
Inclusion Criteria:
- Age ≥22
- Male or female
- Documented diagnosis of diabetes mellitus, meeting the criteria established by the American Diabetes Association (ADA) and World Health Organization (WHO).
- Understand the study and volunteer to sign the informed consent
Exclusion Criteria:
- Uncorrectable vision loss (e.g., with the use of eyeglasses), blurred vision, or floaters.
- Diagnosed with macular edema, severe non-proliferative retinopathy, proliferative retinopathy, radiation retinopathy, or retinal vein occlusion.
- History of laser treatment of the retina or injections into either eye, or any history of retinal surgery.
- Currently participating in another investigational eye study and actively receiving investigational product for Diabetic Retinopathy (DR) or Diabetic Macular Edema (DME).
- Participant has a condition that, in the opinion of the investigator, would preclude participation in the study (e.g., unstable medical status including blood pressure or glycemic control, microphthalmia or previous enucleation).
Participant is contraindicated for imaging by fundus imaging systems used in the study:
- Participant is hypersensitive to light
- Participant recently underwent photodynamic therapy (PDT)
- Participant is taking medication that causes photosensitivity
- Participant has a history of angle-closure glaucoma or narrow anterior chamber angles
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: AEYE Software Device
An AI software device (AEYE-DS) to be used as a diagnostic tool to assist primary care clinicians in screening for diabetic retinopathy using digital funduscopic images.
The device automatically detects more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy.
|
Eligible participants will undergo the following procedures:
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Sensitivity and Specificity, Based on Two Macula-centered Images (One Image From Each Eye of the Patient)
Time Frame: 1 day
|
Sensitivity and specificity of the AEYE-DS device to detect mtmDR on digital funduscopic images, acquired by the Topcon NW400 fundoscopy device, based on two macula-centered images (one image from each eye of the patient).
Sensitivity and specificity were calculated by comparing the output results of the AEYE-DS device to the ground truth (diagnostic determination by an independent, blinded panel of expert ophthalmologists at a reading center).
The worst of two eyes was compared with the AEYE-DS output at the participant level.
|
1 day
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Sensitivity and Specificity, Based on Four Images (One Macula Centered Image and One Optic Disc Centered Image Per Eye)
Time Frame: 1 day
|
Sensitivity and specificity of the AEYE-DS device to detect mtmDR from digital funduscopic images, acquired by the Topcon NW400 fundoscopy device, based on four images (one macula centered image and one optic disc centered image per eye).
Sensitivity and specificity were calculated by comparing the output results of the AEYE-DS device to the ground truth (diagnostic determination by an independent, blinded panel of expert ophthalmologists at a reading center).
The worst of two eyes was compared with the AEYE-DS output at the participant level.
|
1 day
|
|
Imageability, Based on Two Macula-centered Images (One Image From Each Eye of the Patient)
Time Frame: 1 day
|
To assess the imageability of AEYE-DS, defined as the percentage of participants with a diagnostic output (positive for more than mild DR or negative for more than mild DR) from the AEYE-DS device.
Participants with no diagnostic output (i.e., AEYE-DS output of "Insufficient Quality") lowers the imageability of the AEYE-DS.
|
1 day
|
|
Imageability, Based on Four Images (One Macula Centered Image and One Optic Disc Centered Image Per Eye)
Time Frame: 1 day
|
To assess the imageability of AEYE-DS, defined as the percentage of participants with a diagnostic output (positive for more than mild DR or negative for more than mild DR) from the AEYE-DS device.
Participants with no diagnostic output (i.e., AEYE-DS output of "Insufficient Quality") lowers the imageability of the AEYE-DS.
|
1 day
|
Collaborators and Investigators
Sponsor
Sponsor
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- AEYE-DS-001
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.