Registry of Patients for MRI With Non-MRI Conditional Pacemakers and ICD Non-MRI Conditional Pacemakers and ICD

August 30, 2021 updated by: Lancaster General Hospital

A Single Institution Registry Trial Evaluating the Safety of MRI for Patients With Non-MRI Conditional Pacemakers and ICD

The purpose of the registry is to accumulate pre-and post-scan device interrogation data for the purpose of determining the risk of MRI for patients with implantable devices.

Study Overview

Detailed Description

Registry Purpose

This registry is a prospective single institution study of MRI examinations in patients with pacemakers or ICD's who will undergo an MRI that is ordered for routine clinical care. The purpose of the registry is to accumulate pre-and post-scan device interrogation data for the purpose of determining the risk of MRI for patients with implantable devices.

Registry data will add to existing data on the safety of MRI scans for patients with pacemakers and ICDs. This safety data may impact physicians' diagnostic choices for this patient population. Patients with conditions for whom MRI is the best diagnostic modality but who have not been able to undergo MRI scanning may be able to do so through registry participation. The diagnostic information obtained from the scan may assist physicians' treatment decisions and may ultimately affect patient outcomes.

In addition, to our knowledge, previous registries of MRI's in non-conditional cardiac devices have been limited to academic centers. This registry would be among the first to demonstrate the feasibility and safety of this approach in a community hospital setting.

Study Methods

The registry research involves acquiring and analyzing the data of device function and adverse events related to the MRI scan.

The need for MRI will be determined on a case-by-case basis based on the patient's medical history and treatment plan/goals, independent of the registry and research purposes. Only patients with implanted devices with orders for clinically indicated MRI scans will be considered for registry inclusion. Clinical review will be performed and the patient considered after determination that no other imaging modality would be appropriate.

The appropriateness of the MRI study must be discussed, in person or via phone call, between a cardiac radiologist, as determined by Paul Leslie, M.D., and the requesting physician. The discussion should include consideration of alternative imaging modalities and diagnostic quality of MRI imaging of requested body region given implanted device(s). The requesting physician is then responsible for discussing with the patient and/or their medical proxy the reasoning for the determination, based on the final decision made by a radiologist.

All patients will be screened prior to enrollment. Any patient with any commercially available pacemaker or ICD system implanted after 2001 who meet all of the inclusion criteria and none of the exclusion criteria, and who require a medically-indicated MRI as described above will be eligible to enroll in the study. The presence/absence of capped/abandoned/epicardial leads/subcutaneous coil will be confirmed and documented by review of the most recent chest x-ray obtained. The x-ray must post-date the most recent device/lead intervention, as determined by a review of the patient's medical record.

If an MRI procedure is deemed absolutely necessary, the MRI will be performed at the Lancaster General Hospital Lime Street MRI facility. MRI scan sequences, field intensity and field(s) of exposure will be selected to minimize risk to the patient while gaining needed diagnostic information for diagnosis or therapy management.

The MRI will be performed with a 1.5 T scanner utilizing institutional protocols and the following settings:

  • Normal operating mode
  • SAR ≤ 2.0 W/kg per patient.

Prior to the procedure, a pacemaker technologist or manufacturer clinical representative will interrogate the pacemaker or ICD and record the following parameters:

  • Patient's underlying rhythm
  • Atrial and right/left ventricular pacing thresholds
  • P-wave and R-wave amplitude
  • Lead impedance (atrial, ventricular, high voltage)
  • Battery voltage
  • Arrhythmia history
  • Tachyarrhythmia therapy

The following pacemaker/ICD parameters will be deactivated for the duration of the MRI:

  • Magnet response (ICD)
  • Rate response
  • PVC response
  • Noise reversion response
  • Ventricular sense response
  • Mode switching
  • Tachycardia detection/therapies (ICD)
  • EGM triggers

Patients who are not pacemaker dependent will have their pacemaker/ICD programmed to AAI/VVI/DDI at 40 beats per minute for the duration of the MRI. Patients who are pacemaker dependent (pacing at programmed lower rate during the pre-scan interrogation) will have their pacemaker/ICD programmed to AOO/VOO/DOO at the programmed lower rate for the duration of the MRI. Emergency equipment, including a MRI compatible defibrillator and transcutaneous pacemaker, will be available throughout the procedure in the event of an adverse clinical event. In addition to radiology staff, a pacemaker technologist or manufacturer clinical representative and an EP tech/RN with current ACLS will be present for the duration of the procedure. An EP physician/advanced practice provider (APP) must be immediately available (i.e., in the hospital) to provide medical intervention, such as placement of a temporary external pacemaker if required. In the event that a code is called by the MRI staff, hospital protocol indicates that ER staff will respond to the code.

Radiology staff and the electrophysiology (EP) technologist or nurse with current ACLS certification will monitor the following patient parameters at least every 5 minutes throughout the MRI scan:

  • Heart rate
  • Heart rhythm
  • Non-invasive blood pressure
  • Oxygen saturation
  • Symptoms

If any of the following adverse symptoms/events occur during the scan, the test will be immediately terminated:

  • Burning/pulling sensation in device pocket/chest during MRI
  • Potentially lethal/previously non-diagnosed arrhythmias during MRI
  • Spontaneous and unanticipated pacing rate change
  • SOB
  • Chest pain
  • Feeling of heat/pulling in the chest or device pocket

The patient will be removed from the scanner to the holding area and the device interrogated and reprogrammed to original parameters or as indicated by interrogation. If needed, emergency treatment will be provided as per ACLS protocols or at the discretion of the monitoring physician. The following parameters will be compared to the pre-MRI results:

  • Patient's underlying rhythm
  • Atrial and right/left ventricular pacing thresholds
  • P-wave and R-wave amplitude
  • Lead impedance (atrial, ventricular, high voltage)
  • Battery voltage

All device-related adverse events/symptoms will be documented on the data collection form and reported to the IRB. The patient will remain in the MRI suite holding area until medically stable. The need for further treatment will be at the discretion of the monitoring physician.

The EP physician/APP will review and compare the pre/post MRI device data prior to the patient's discharge from the MRI suite. Following testing, the pacemaker or ICD will be reprogrammed to the pre-procedure parameters. The need for device reprogramming related to the MRI scan will be based on the clinical judgment of the attending electrophysiologist.

Follow-up

Patients who do not experience any device related adverse events during the MRI will return for one follow-up visit 1-6 weeks after the MRI. The follow-up visit will include pacemaker/ICD interrogation and testing to confirm appropriate function. At a minimum, device interrogation/testing will include:

  • Atrial and right/left ventricular pacing thresholds
  • P-wave and R-wave amplitude
  • Lead impedance (atrial, ventricular, high voltage)
  • Battery voltage

Patients who experience one or more device-related adverse events during the MRI will return 1-7 days post-MRI for follow-up pacemaker/ICD testing to confirm appropriate function. At a minimum, device interrogation/testing will include:

  • Atrial and right/left ventricular pacing thresholds
  • P-wave and R-wave amplitude
  • Lead impedance (atrial, ventricular, high voltage)
  • Battery voltage

The need for further follow-up will be determined by the physician based on testing results and clinical need.

The results of the follow-up device interrogation, including whether MRI-related adverse events have resolved, will be documented and kept in the device chart and recorded in the Registry database.

If the insurance provider does not cover the MRI and device checks, those costs may be the full responsibility of the patient. The patient will be informed of the status of the pre-authorization before scheduling the MRI.

Study Type

Observational

Enrollment (Actual)

262

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Pennsylvania
      • Lancaster, Pennsylvania, United States, 17604
        • Lancaster General Hospital

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Any patient with any commercially available pacemaker or ICD system implanted after 2001 who meet all of the inclusion criteria and none of the exclusion criteria, and who require a non-cardiac, medically-indicated MRI as described above will be eligible to enroll in the study.

Description

  1. Patient has a non-MRI conditional permanently implanted pacemaker or ICD implanted after 2001
  2. Patient has a strong clinical indication for MRI where MRI is the diagnostic modality of choice for specific disease state without acceptable alternative imaging technologies as determined by the ordering physician and a radiologist
  3. Patient is at least 18 years of age
  4. Patient is willing and able to sign study informed consent and HIPAA authorization

Exclusion Criteria:

  1. Non-device related contraindication for MRI (such as implanted metallic objects, claustrophobia, morbid obesity)
  2. Presence of capped/abandoned/epicardial leads or subcutaneous coil
  3. Pregnancy
  4. Device generator at elective replacement interval
  5. Abdominal device implant

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Time Frame
Number of Participants With Burning or Pulling Sensations in Chest/Device Pocket During MRI
Time Frame: Through procedure completion, an average of 1 hour
Through procedure completion, an average of 1 hour
Number of Participants With Potentially Lethal/Previously Non-diagnosed Arrhythmias During MRI
Time Frame: Through procedure completion, an average of 1 hour
Through procedure completion, an average of 1 hour
Number of Participants With Spontaneous and Unanticipated Pacing Rate Change
Time Frame: Through procedure completion, an average of 1 hour
Through procedure completion, an average of 1 hour
Number of Participants With Power on Reset of Pacemaker or ICD
Time Frame: Through procedure completion, an average of 1 hour
Through procedure completion, an average of 1 hour
Number of Participants With MRI Related Device Malfunction/Failure
Time Frame: Through procedure completion, an average of 1 hour
Through procedure completion, an average of 1 hour
Number of Participants With Serious Injury/Death Related to MRI Device Malfunction/Failure
Time Frame: Through procedure completion, an average of 1 hour
Through procedure completion, an average of 1 hour
Number of MRI Scans With Post-MRI Device Parameter Change: Greater Than 50 Ohm Change in Atrial or Ventricular (RV and/or LV) Lead Impedance
Time Frame: Through procedure completion, an average of 1 hour
Through procedure completion, an average of 1 hour
Number of MRI Scans With Post-MRI Device Parameter Changes: Greater Than 5 Ohm Change in High Voltage Lead Impedance
Time Frame: Through procedure completion, an average of 1 hour
Through procedure completion, an average of 1 hour
Number of MRI Scans With Post-MRI Device Parameter Changes: Greater Than 50% Decrease in Atrial or Ventricular (RV and/or LV) Sensing Threshold
Time Frame: Through procedure completion, an average of 1 hour
Through procedure completion, an average of 1 hour
Number of MRI Scans With Post-MRI Device Parameter Changes: Greater Than or Equal to a 1 Volt Increase in Atrial or Ventricular (RV and/or LV) Pacing Threshold
Time Frame: Through procedure completion, an average of 1 hour
Through procedure completion, an average of 1 hour

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Sandeep Bansal, MD, Lancaster General Hospital

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Helpful Links

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

December 1, 2014

Primary Completion (Actual)

June 30, 2018

Study Completion (Actual)

June 30, 2018

Study Registration Dates

First Submitted

December 10, 2014

First Submitted That Met QC Criteria

December 16, 2014

First Posted (Estimate)

December 17, 2014

Study Record Updates

Last Update Posted (Actual)

August 31, 2021

Last Update Submitted That Met QC Criteria

August 30, 2021

Last Verified

August 1, 2021

More Information

Terms related to this study

Other Study ID Numbers

  • 2014-68

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.