- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT03610269
Spot-Check Noninvasive Hemoglobin (SpHb) Clinical Validation of INVSENSOR00026
June 5, 2019 updated by: Masimo Corporation
The primary objective of this clinical investigation is to evaluate the accuracy of using Masimo's INVSENSOR00026 Pulse CO-Oximeter and Sensor to measure hemoglobin, as compared to hemoglobin measurements obtained from various lab analyzer(s) and point of care device(s).
Study Overview
Status
Completed
Conditions
Intervention / Treatment
Study Type
Interventional
Enrollment (Actual)
143
Phase
- Not Applicable
Contacts and Locations
This section provides the contact details for those conducting the study, and information on where this study is being conducted.
Study Locations
-
-
California
-
Lomita, California, United States, 90717
- Torrance Clinical Research
-
-
Participation Criteria
Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.
Eligibility Criteria
Ages Eligible for Study
18 years and older (Adult, Older Adult)
Accepts Healthy Volunteers
Yes
Genders Eligible for Study
All
Description
Inclusion Criteria:
- 18 years of age or older
- Subject has given written informed consent to participate in the study
Exclusion Criteria:
- (*) Pregnant or positive human chorionic gonadotropin (hCG) test
- (*) Any severe coagulopathy, chronic bleeding disorders (i.e., hemophilia) or recent thrombolysis
- (*) Hemoglobinopathies and synthesis disorders (i.e., sickle cell, thalassemias, etc.)
- (*) Subjects who are currently taking anticoagulant medication
- (*) Subjects with allergies to lidocaine, Pain Ease, latex, adhesives, or plastic
- (*) Subjects with skin abnormalities at the planned application sites that may interfere with sensor application, per directions-for-use (DFU) or trans-illumination of the site, such as psoriasis, eczema, angioma, burns, scar tissue, substantial skin breakdown, nail polish, acrylic nails, infections, abnormalities, etc.
- (*) Subjects unlikely to be able to refrain from excessive motion during data collection. Excessive motion includes postural changes, making hand gestures, involuntary muscular movements, etc.
- (*) Subjects with elevated blood pressure, skin or wrist abnormalities that may interfere with an arterial blood draw as determined by investigator or research medical staff.
- (*) Subjects who intend on participating in heavy lifting, repetitive movement of their wrist (including riding a motorcycle) or exercise (working out, riding a bike, etc.), or any activity that will put additional stress on the wrist within 24 hours following a study involving an arterial blood draw
- Subjects deemed not suitable for the study at the discretion of the investigator or research medical staff
Note: (*) May be self-reported by subject
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Design Details
- Primary Purpose: Other
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: INVSENSOR00026
All enrolled subjects receive INVSENSOR00026 Pulse CO-Oximeter and sensor for the noninvasive measurement of hemoglobin (SpHb).
|
Noninvasive Pulse CO-Oximeter and sensor that measures noninvasive hemoglobin (SpHb)
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Accuracy of Noninvasive Hemoglobin (SpHb) Measurement by Arms Calculation
Time Frame: Up to 2 hours per subject
|
Accuracy of the INVSENSOR00026 will be determined by comparing the noninvasive hemoglobin measurement (SpHb) of the INVSENSOR00026 to the hemoglobin value obtained from a reference blood sample and calculating the accuracy root mean square (Arms) error value.
In order to obtain the Arms value, the blood sample hemoglobin value is subtracted from the INVSENSOR00026 hemoglobin value for a number of samples, the average of this difference is computed as the bias.
The standard deviation of the differences is computed as the precision.
The square root of the sum of the squares of bias and precision is computed as the Arms Error value.
|
Up to 2 hours per subject
|
Collaborators and Investigators
This is where you will find people and organizations involved with this study.
Sponsor
Study record dates
These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.
Study Major Dates
Study Start (Actual)
August 1, 2018
Primary Completion (Actual)
August 16, 2018
Study Completion (Actual)
August 16, 2018
Study Registration Dates
First Submitted
July 26, 2018
First Submitted That Met QC Criteria
July 26, 2018
First Posted (Actual)
August 1, 2018
Study Record Updates
Last Update Posted (Actual)
June 25, 2019
Last Update Submitted That Met QC Criteria
June 5, 2019
Last Verified
June 1, 2019
More Information
Terms related to this study
Other Study ID Numbers
- TORR0003
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
NO
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
No
Studies a U.S. FDA-regulated device product
Yes
product manufactured in and exported from the U.S.
No
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.