Wearable Evaluation of Ambulatory Readings for Blood Pressure (WEAR-BP)

August 4, 2025 updated by: Stephen Juraschek
The purpose of this project is to compare the performance and validity of novel wearable technologies that measure blood pressure (BP) and physical activity with a Spacelabs Ambulatory Blood Pressure Monitoring (ABPM) device.

Study Overview

Detailed Description

The investigators will recruit community-dwelling participants to wear novel cuffless BP monitoring devices for comparison with a Spacelabs ABPM device throughout a 24-hour period. The identification of accurate cuffless BP technologies would greatly improve access to ABPM and allow for BP measurements at times when the use of cuff-based devices are not feasible. Knowledge from this project will inform subsequent research protocols among adults and under-represented groups.

Up to 250 participants (at least 100) will be involved in this study each wearing up to 7 devices (6 BP devices and 1 activity monitor). The assignment of devices will be randomized by day and by the arm or wrist on which they are worn. During the first visit, participants will provide their consent and be randomly fitted with 3 BP devices; either the Spacelabs ABPM device or the ABPMpro device, the Aktiia wristband BP monitor, and one of two other cuff-based wrist BP monitors, along with an activity monitor to track body positions. These devices will be worn continuously for at least 24-hours, including overnight, before being returned. On the following day, for Visit 2, participants will be fitted with a different set of 3 BP devices; the ABPMpro device or Spacelabs ABPM device, the Aktiia wristband BP device (on the opposite wrist from the previous day), and the final cuff-based wristband BP device. Additionally, participants will wear an ambulatory BP device as a patch on the chest. These devices will also be worn for at least 24 hours. This study will enable characterization of the feasibility, practicality, and effectiveness of the measurements and the devices' accuracy and precision.

Study Type

Interventional

Enrollment (Estimated)

100

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Contact Backup

Study Locations

    • Massachusetts
      • Boston, Massachusetts, United States, 02215
        • Recruiting
        • Beth Israel Deaconess Medical Center
        • Contact:

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

Yes

Description

Inclusion Criteria:

  • Age 18 to 100 years
  • All neighborhoods in Boston area
  • Be able to walk with two limbs and have two arms
  • Willing and able to complete required measurement procedures
  • Able to provide informed consent

Exclusion Criteria:

  • Failure to receive informed consent
  • Arm circumference of more than 50cm

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Other
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Cuff-based ambulatory blood pressure monitor sequence
Participants will be randomized 1:1 to wear (A) the Spacelabs ambulatory blood pressure monitor on the first day and ABPMPro ambulatory blood pressure monitor on the second day or (B) ABPMPro ambulatory blood pressure monitor on the first day then Spacelabs ambulatory blood pressure monitor on the second day.
This is an upper arm, cuff-based intermittent ambulatory BP device. This device has an internal activity and body position sensor which gathers information about sleep-wake estimation.
This is an upper arm cuff-based device with tubing for continuous ambulatory BP monitoring with a unique activity sensor that correlate with patient movement activity with blood pressure changes.
Experimental: Left versus right upper limb
Participants will be randomized to wear the Aktiia cuffless blood pressure device on either the left upper limb versus the right upper limb on the first day. On the second day, the Aktiia blood pressure device will be switched to the other arm which did not get randomized on the first day. Another wristband (LiveMetric versus Bpro) will be worn on the contralateral arm.
This is a wrist cuff-based device which uses an optical sensor (photoplethysmography sensor) to gather data from the arteries under the skin surface. Aktiia has an upper arm cuff device which is used for calibration once every month to provide a baseline blood pressure reading.
This is a wrist cuff-based applanation tonometry continuous ambulatory BP device. This device has automatic self-calibration and comes in various sizes.
This is a wrist cuff-based wireless oscillometric, continuous ambulatory BP device. It has applanation tonometry at the wrist with Pulse wave analysis. This device needs calibration with a brachial cuff BP measurement.
Experimental: Wristband sequence (first day versus second day)
Participants will be randomized 1:1 to wear (A) the LiveMetric wristband on the first day and Bpro wristband on the second day or (B) Bpro wristband on the first day then LiveMetric wristband on the second day. The arm side (i.e., right or left) will be determined by the Aktiia cuffless blood pressure device randomization on the first day. For the second day, the Aktiia cuffless blood pressure device will be worn on the arm opposite the first day with the other device (LiveMetric versus Bpro) worn on the contralateral arm.
This is a wrist cuff-based device which uses an optical sensor (photoplethysmography sensor) to gather data from the arteries under the skin surface. Aktiia has an upper arm cuff device which is used for calibration once every month to provide a baseline blood pressure reading.
This is a wrist cuff-based applanation tonometry continuous ambulatory BP device. This device has automatic self-calibration and comes in various sizes.
This is a wrist cuff-based wireless oscillometric, continuous ambulatory BP device. It has applanation tonometry at the wrist with Pulse wave analysis. This device needs calibration with a brachial cuff BP measurement.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Systolic Blood Pressure
Time Frame: 24 - 48 hours
Measured from all 5 randomized devices
24 - 48 hours

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Diastolic Blood Pressure
Time Frame: 24 - 48 hours
Measured from all 5 randomized devices
24 - 48 hours
Heart Rate
Time Frame: 24 - 48 hours
Measured from all 5 randomized devices
24 - 48 hours
ActivPaL device (Physical activity monitor)
Time Frame: 24 - 48 hours
Physical activity intensity and type of activity will be characterized throughout the 48hr wear period. While the investigators are interested in the impact of devices on activity, the investigators are also interested in the relationship between activities and blood pressure.
24 - 48 hours

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Systolic blood pressure from the Biobeat device (Chest Patch Blood Pressure device)
Time Frame: 24 hours
This is a continuous ambulatory BP device which is worn as a patch on the chest. This device needs calibration with a brachial cuff BP measurement. This will be typically worn on day 2. This is not a randomized assessment.
24 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

June 6, 2025

Primary Completion (Estimated)

May 3, 2026

Study Completion (Estimated)

June 30, 2026

Study Registration Dates

First Submitted

September 12, 2024

First Submitted That Met QC Criteria

September 20, 2024

First Posted (Actual)

September 24, 2024

Study Record Updates

Last Update Posted (Actual)

August 7, 2025

Last Update Submitted That Met QC Criteria

August 4, 2025

Last Verified

August 1, 2025

More Information

Terms related to this study

Additional Relevant MeSH Terms

Other Study ID Numbers

  • 2024P000441
  • 5R01HL158622 (U.S. NIH Grant/Contract)

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

UNDECIDED

IPD Plan Description

Participants will be consented in the event data will be shared with other investigators.

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.