Multi-Centre Based Study Recognition of Atrial Fibrillation With the Corsano CardioWatch 287-2 System (COR-AF)

April 30, 2025 updated by: Corsano Health B.V.

This concerns a multi-center, single-arm, diagnostic accuracy study. In this study, a wristband that monitors heart rythm is assessed in clinical and free living conditions. The performance of the wristband in detecting atrial fibrillation and normal sinus rhythms will be compared to a short 12-lead ECG or 24-hour holter.

To ensure that both patients with and without atrial fibrillation are included whilst limiting the burden on patients and hospital staff, four groups of patients are being recruited:

  • Group A: patients receiving an ECG holter at the outpatient clinic for 24-48 hours of remote monitoring
  • Group B: patients diagnosed with (paroxysmal) AF presenting to the clinic for routine follow-up evaluation.
  • Group C: patients diagnosed with (paroxysmal) AF registered in a patient organisation for patients with AF.
  • Group D: patients receiving a 12-lead ECG at the outpatient clinic as standard of care.

The secondary outcomes of the study focus on assessing the primary outcomes in light of demographic subgroups, patient physical activity and the signal quality index.

Study Overview

Status

Not yet recruiting

Intervention / Treatment

Detailed Description

Rationale:

Wearables have the potential to monitor patients remotely. The Corsano CardioWatch 287-2 is such a medical device that can monitor atrial fibrillation. The device has been validated using clinical trials in hospitals, but evaluation in the intended remote setting across various demographic groups is lacking.

Primary objective:

To compare episodes of atrial fibrillation across 30-second intervals detected by the Corsano CardioWatch 287-2 with episodes of atrial fibrillation detected by conventional Holter monitoring across 24-48 hours or short in-clinic 12-lead ECG based on the positive predictive value, negative predictive value, sensitivity and specificity.

Secondary objective: To compare the number of atrial fibrillation episodes greater than 6 minutes, 1 hour, 6 hours, and 12 hours duration detected by the CardioWatch 287-2 and with the number of these specific episodes detected by conventional Holter across 24-48 hours days. As well as to assess primary outcomes in light of demographic subgroups, patient activity and the signal quality index.

Study population:

One group (A) receiving an ECG monitoring holter for the duration of 24-48 hours days as part of standard care for atrial fibrillation screening. Besides, one group (B) diagnosed with (paroxysmal) without indication for holter monitoring according to standard of care AF presenting to the clinic for routine follow-up evaluation. Adding, one group (C) diagnosed with (paroxysmal) and registered in a patient organisation for patients with AF. Finally, one group (D) receiving in-clinic 12-lead ECG monitoring as part of standard of care.

Study design:

This concerns a multi-center, single-arm, diagnostic accuracy study.

Four different groups of patients will be approached for participation in the trial:

  • Group A: patients receiving an ECG holter at the outpatient clinic for 24-48 hours of remote monitoring
  • Group B: patients diagnosed with (paroxysmal) AF presenting to the clinic for routine follow-up evaluation.
  • Group C: patients diagnosed with (paroxysmal) AF registered in a patient organisation for patients with AF.
  • Group D: patients receiving a 12-lead ECG at the outpatient clinic as standard of care.

Groups A, B and C will receive ECG holter monitoring as well as continuous monitoring by the Corsano CardioWatch 287-2 across 24-48 hours to assess whether the amount and duration of AF episodes detected by the investigational device correspond to the gold standard. Detection of atrial fibrillation with the Corsano CardioWatch 287-2 is done by optical photoplethysmography (PPG), after which atrial fibrillation can be confirmed with a 30-second ECG measurement. Additionally, all participants will be asked to perform 10 single-lead ECG's at standard timepoints (approx. every 2 hours when awake) across the 24-48 hours as well as additional ECG measurements in case of AF (at least 30 minutes between alerts). Group B was added to ensure that a considerable amount of the included data includes periods with atrial fibrillation. As the results of a prior study (RECAMO study NL NL83281.000.22) show that the amount of patients in Group A with periods of atrial fibrillation is limited (7.0%). Furthermore, Group C was added to limit the additional workload for hospital staff and ensure inclusion of a diverse patient population from different demographic and ethnical groups. For this group two (virtual) visits will be scheduled before and after the study period to detect any alarming signals before participation in the study and to share and explain the results after study completion. Finally, group D, a group receiving a short in-clinic 12-lead ECG as standard of care was added to ensure the inclusion of a broad and sizable demographic group whilst keeping the burden on the patient and hospital staff limited.

Study Type

Observational

Enrollment (Estimated)

350

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

  • Name: Eelko Ronner, MD, PhD
  • Phone Number: +31 015-2603320
  • Email: eronner@rdgg.nl

Study Locations

    • Zuid Holland
      • Delft, Zuid Holland, Netherlands, 2625AD
        • Reinier de Graaf Gasthuis
        • Contact:
          • Eelko Ronner, MD, PhD
          • Phone Number: +31 015 - 260 33 20
          • Email: eronner@rdgg.nl

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

Yes

Sampling Method

Non-Probability Sample

Study Population

Study population for groups A-B-D includes patients visiting the outpatient clinic of at least 4 different hospitals, of which one is located in the Netherlands (Reinier de Graaf Gasthuis, Delft) and three are located in different locations across the USA.

GROUP A Patients receiving an ECG holter at the outpatient clinic for 24 hours of remote monitoring (standard of care)

GROUP B Patients diagnosed with (paroxysmal) AF presenting to the clinic for routine follow-up evaluation (NOT necessarily standard of care)

GROUP C Patients diagnosed with (paroxysmal) AF registered in a patient organisation for patients with AF (standard of care)

GROUP D Patients receiving a 12-lead ECG at the outpatient clinic as standard of care (NOT standard of care)

Description

Inclusion Criteria:

  • Patients with symptoms indicative of cardiac arrhythmias receiving 24-48 hour ECG holter for home monitoring or in-clinic 12-lead ECG per doctor prescription OR previous diagnosis of arrhythmia presenting to the clinic for routine follow-up evaluation
  • Age ≥ 22 years old
  • Able to provide informed consent.
  • Proficient in written and spoken Dutch or English, defined by self-report of comfort reading, writing, and speaking Dutch or English.

Exclusion Criteria:

  • Clinical conditions that in the opinion of the Investigator would compromise the safety of the patient or ability to complete the protocol;
  • Unable to wear the Corsano CardioWatch 287 due to reasons such as allergic reactions, wounds, amputations etc.;
  • Unable or not willing to receive ECG monitoring;
  • Unable or not willing to sign informed consent;

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Group A: patients receiving 24-48 hour holter monitoring as standard of care
Holter monitoring will be compared to continuous monitoring by the Corsano CardioWatch 287-2 across 24-48 hours to assess whether the amount and duration of AF episodes detected by the investigational device correspond to an ECG-holter reference across various demographic groups. Detection of atrial fibrillation with the Corsano CardioWatch 287-2 is done by optical photoplethysmography (PPG), after which atrial fibrillation can be confirmed with a 30-second ECG measurement. Additionally, all participants will be asked to perform 10 single-lead ECG's with the investigational device at standard timepoints (every 2 hours when awake), and when AF is detected (minimal 30 minutes time difference between alerts) across the 24-48 hours.
Patients receive monitoring of their heart rhythm by the CardioWatch 287-2 next to monitoring of their heart rhythm by either a 24-48 holter or a short in-clinic 12-lead ECG monitoring.
Group B: patients diagnosed with (paroxysmal) AF presenting to the clinic for routine follow-up
Patients with a known diagnosis of (paroxysmal) atrial fibrillation presenting to the clinic for routine follow-up evaluation will be asked to eligibly participate in ECG-holter monitoring as well as continuous monitoring by the Corsano CardioWatch 287-2 across 24-48 hours. Similar to group A, these patients will also be asked to perform manual ECG measurements with the investigational device. Group B was added to ensure that a considerable amount of the included data includes periods with atrial fibrillation. As the results of a prior study (RECAMO study NCT05899959) show that the amount of patients in Group A with periods of atrial fibrillation is limited (7.0%).
Patients receive monitoring of their heart rhythm by the CardioWatch 287-2 next to monitoring of their heart rhythm by either a 24-48 holter or a short in-clinic 12-lead ECG monitoring.
Group C: patients diagnosed with (paroxysmal) AF registered in a patient organisation
Patients with a known diagnosis of (paroxysmal) atrial fibrillation registered in a patient organisation for patients with AF will be asked to eligibly participate in ECG-holter monitoring as well as continuous monitoring by the Corsano CardioWatch 287-2 across 24-48 hours. Similar to group A and B, these patients will also be asked to perform manual ECG measurements with the investigational device. For this group two (virtual) visits will be scheduled before (visit a) and after (visit b) the study period. In visit a) additional information regarding the study will be provided and potential alarm signals will be screened to ensure the patient does not participate in the trial in case direct medical attention is required. If the patient does not show any alarm signals they will be asked to provide informed consent. Once informed consent is provided all materials will be mailed to them via postal services after which they can perform the measurements and return the materials via postal
Patients receive monitoring of their heart rhythm by the CardioWatch 287-2 next to monitoring of their heart rhythm by either a 24-48 holter or a short in-clinic 12-lead ECG monitoring.
Group D: patients receiving a 12-lead ECG at the outpatient clinic as standard of care.
Participants in this group will wear the CardioWatch 287-2 during the standard 12-lead ECG recordings conducted in the outpatient clinic. Patients will be asked to perform a manual ECG spot-check with the investigational device during (if not feasible directly before), and immediately following the 12-lead ECG. Furthermore, the CardioWatch 287-2 will be worn for at least 5 minutes prior to and following the 12-lead ECG. Although these recordings are brief, this group was included to enhance the study population's size and demographic diversity while minimizing the burden on both patients and hospital staff. This group is expected to consist of both non-AF and AF patients, providing valuable insights into the investigational device's performance in a setting similar to its intended real-world application.
Patients receive monitoring of their heart rhythm by the CardioWatch 287-2 next to monitoring of their heart rhythm by either a 24-48 holter or a short in-clinic 12-lead ECG monitoring.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Atrial fibrillation detection across 30-second intervals by CardioWatch PPG algorithm vs. 24-48 hour Holter
Time Frame: From enrollment up to 48 hours
The absolute number of detected AF 30-second intervals by the CardioWatch's PPG algorithm will be determined and compared to the reference ECG holter monitoring, by means of calculating the positive predictive value (PPV), negative predictive value (NPV), specificity, and sensitivity. Performance goals for PPV and NPV are ≥90%, and performance goals for sensitivity and specificity are ≥95%.
From enrollment up to 48 hours
Atrial fibrillation detection across 30-second intervals by CardioWatch PPG algorithm vs. short in-clinic 12-lead ECG
Time Frame: From enrollment up to 30 minutes
The absolute number of detected AF 30-second intervals by the CardioWatch's PPG algorithm will be determined and compared to the reference in-clinic 12-lead ECG, by means of calculating the positive predictive value (PPV), negative predictive value (NPV), specificity, and sensitivity. Performance goals for PPV and NPV are ≥90%, and performance goals for sensitivity and specificity are ≥95%.
From enrollment up to 30 minutes
Atrial fibrillation detection across 30-second intervals by CardioWatch single-lead ECG vs. 24-48 hour Holter
Time Frame: From enrollment up to 48 hours
The absolute number of detected AF by the CardioWatch's single-lead ECG algorithm will be determined and compared to the reference ECG holter monitoring at the most similar moment in time by means of calculating the PPV, NPV, specificity, and sensitivity for timestamps corresponding to the single-lead spot checks.
From enrollment up to 48 hours
Atrial fibrillation detection across 30-second intervals by CardioWatch single-lead ECG vs. short in-clinic 12-lead ECG
Time Frame: From enrollment up to 30 minutes
The absolute number of detected AF by the CardioWatch's single-lead ECG algorithm will be determined and compared to the reference in-clinic ECG at the most similar moment in time by means of calculating the PPV, NPV, specificity, and sensitivity for timestamps corresponding to the single-lead spot checks.
From enrollment up to 30 minutes

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Correspondence between cases with AF of greater than 6 minutes, 1 hour, 6 hours, and 12 hours
Time Frame: From enrollment up to 48 hours
Correspondence between cases with AF of greater than 6 minutes, 1 hour, 6 hours, and 12 hours duration detected by the ECG Holter monitor and the CardioWatch's PPG-algorithm, provided as a percentage of the total cases registered by the ECG Holter monitor for each duration as determined by the Clinical Adjudication Committee (CAC).
From enrollment up to 48 hours
Compliance with ECG alerts
Time Frame: From enrollment up to 48 hours
Compliance with ECG alerts during at home holter-monitoring, together with:
From enrollment up to 48 hours
Amount of ECG attempts before successful ECG
Time Frame: From enrollment up to 48 hours
Amount of ECG attempts before successful ECG during at home holter-monitoring
From enrollment up to 48 hours
Confirmation of AF by single-lead ECG
Time Frame: From enrollment up to 48 hours
The confirmation of AF: single-lead ECG vs PPG-based (ECGs within one hour of alert)
From enrollment up to 48 hours
RMSE of HR measurement and reference
Time Frame: From enrollment up to 48 hours
The root mean square error (RMSE) between the single-lead ECG heart rate (HR) and reference ECG holter HR.
From enrollment up to 48 hours
Atrial fibrillation detection performance for males vs. females
Time Frame: Baseline information gathered during enrollment
Atrial fibrillation detection performance determined for subgroups of males and females individually.
Baseline information gathered during enrollment
Atrial fibrillation detection performance for BMI subgroups
Time Frame: Baseline information gathered during enrollment
Atrial fibrillation detection performance determined for subgroups of BMI: Underweight (<18.5); Normal weight (18.5-24.9); Overweight (25-29.9); Obese (≥30).
Baseline information gathered during enrollment
Atrial fibrillation detection performance for skin colour based subgroups
Time Frame: Baseline information gathered during enrollment
Atrial fibrillation detection performance determined for subgroups of Skin colour: Monk skin tone (MST) scale 1-3; MST 4-6; MST 7-10. (Researcher will compare reference colors in scale as described in Monk et al. (2023) to the skin of participants)
Baseline information gathered during enrollment
Atrial fibrillation detection performance for ethinicity based subgroups
Time Frame: Baseline information gathered during enrollment

Atrial fibrillation detection performance determined for subgroups of Ethnicity: Caucasian; African American; Asian; Hispanic/Latino.

According to FDA recommendations at least 30 patients total, 15 AF patients and 15 NSR patients, need to be included at each of the subgroup levels of Monk skin tone, ethnicity, sex, BMI. After enrolment of 150 patients, accounting of the subgroups will be considered, and further enrolment will be restricted to ensure minimal distributions of each subgroup level are enrolled, specifically focusing on enrollments in study group D.

Baseline information gathered during enrollment
Atrial fibrillation detection performance in light of patient activity
Time Frame: From enrollment up to 48 hours
Atrial fibrillation detection performance for subgroups based on patient activity using ACC: motion; resting; standing
From enrollment up to 48 hours
Atrial fibrillation detection performance in light of signal quality index
Time Frame: From enrollment up to 48 hours
Atrial fibrillation detection performance relative to the signal quality index.
From enrollment up to 48 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Eelko Ronner, MD, PhD, Reinier de Graaf Groep

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Estimated)

May 1, 2025

Primary Completion (Estimated)

January 1, 2026

Study Completion (Estimated)

January 1, 2026

Study Registration Dates

First Submitted

April 9, 2025

First Submitted That Met QC Criteria

April 30, 2025

First Posted (Actual)

May 2, 2025

Study Record Updates

Last Update Posted (Actual)

May 2, 2025

Last Update Submitted That Met QC Criteria

April 30, 2025

Last Verified

April 1, 2025

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Paroxysmal Atrial Fibrillation (PAF)

Clinical Trials on Corsano CardioWatch 287-2

Subscribe