Safety and Clinical Performance of the Next-Generation SAFESYNC Transcatheter Heart Valve in Patients With Severe Aortic Stenosis (SAFESYNC NG)

16. november 2026 opdateret af: Braile Biomedica Ind. Com. e Repr. Ltda.

Evaluation of the Safety and Clinical Performance of the Next-Generation SAFESYNC Balloon-Expandable Transcatheter Heart Valve in Patients With Severe Aortic Stenosis: A Prospective, Multicenter, Single-Arm Pivotal Study

This pivotal, prospective, multicenter (Brazil, Chile and Argentina), single-arm study will evaluate the safety and clinical performance of the Inovare® Safesync Next-Generation bioprosthetic heart valve, implanted by transfemoral transcatheter aortic valve implantation (TAVI) with the B-Delivery System, in patients aged 70 years or older with severe aortic stenosis and an indication for TAVI, regardless of surgical risk. The study will enroll 210 participants.

Studieoversigt

Status

Ikke rekrutterer endnu

Betingelser

Intervention / Behandling

Undersøgelsestype

Interventionel

Tilmelding (Anslået)

210

Fase

  • Ikke anvendelig

Kontakter og lokationer

Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.

Studiekontakt

  • Navn: Clinical Research Department - Braile Biomédica
  • Telefonnummer: +5517 21367000
  • E-mail: pesquisa@braile.com.br

Deltagelseskriterier

Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.

Berettigelseskriterier

Aldre berettiget til at studere

70 år og ældre (Ældre voksen)

Tager imod sunde frivillige

Ingen

Køn, der er berettiget til at studere

Alle

Beskrivelse

Inclusion Criteria:

Severe aortic valve stenosis diagnosed according to criteria established in current guidelines; Age ≥70 years, regardless of surgical risk; Indication for transcatheter aortic valve implantation (TAVI); Willing to participate and to sign the written informed consent form.

Exclusion Criteria:

Any condition considered a contraindication to implantation of a bioprosthetic heart valve (e.g., patient with an indication for a mechanical heart valve);

Known hypersensitivity or contraindication to any of the following that cannot be adequately managed with premedication:

  1. Aspirin, clopidogrel, or heparin (heparin-induced thrombocytopenia [HIT] or heparin-induced thrombocytopenia with thrombosis [HITTS]);
  2. Nitinol (titanium or nickel);
  3. Contrast media; Blood dyscrasias: leukopenia (white blood cell count <1,000/mm³), thrombocytopenia (platelet count <100,000 cells/mm³), history of coagulopathy, or hypercoagulable states; Active infection, including endocarditis; Cardiogenic shock manifested by low cardiac output, vasopressor dependence, or need for mechanical hemodynamic support; Recent stroke or transient ischemic attack (within 6 months); History of gastrointestinal bleeding or other bleeding that precludes anticoagulant or antiplatelet therapy; Estimated life expectancy of less than 5 years; Other medical, social, or psychological conditions that, in the investigator's opinion, prevent the patient from providing consent or adequately complying with the protocol; Pre-existing prosthetic heart valve in the aortic position (biological or mechanical); Concomitant severe mitral or tricuspid valve disease; Hypertrophic obstructive cardiomyopathy; Percutaneous coronary intervention performed within 72 hours prior to the study procedure; Femoral artery diameter <5.5 mm (for Safesync 20 to 24 mm), <6.0 mm (for Safesync 26 and 28 mm), or <6.5 mm (for Safesync 30 mm); High risk of annular rupture or significant residual paravalvular leak due to severe calcification of the aortic valve or left ventricular outflow tract; High risk of coronary obstruction (coronary ostial height <10 mm with narrow sinuses of Valsalva [SOV], defined as a mean SOV diameter <3 mm larger than the bioprosthesis size), as determined by computed tomography angiography; Left ventricular dysfunction with left ventricular ejection fraction (LVEF) ≤30%; Severe chronic kidney disease, defined as glomerular filtration rate <30 mL/min/1.73 m².

Studieplan

Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.

Hvordan er undersøgelsen tilrettelagt?

Design detaljer

  • Primært formål: Behandling
  • Tildeling: N/A
  • Interventionel model: Enkelt gruppeopgave
  • Maskning: Ingen (Åben etiket)

Våben og indgreb

Deltagergruppe / Arm
Intervention / Behandling
Eksperimentel: INOVARE® SAFESYNC Next-Generation TAVI

Participants will undergo TAVI via transfemoral access using the INOVARE® SAFESYNC Next-Generation Bioprosthetic Heart Valve and the B-Delivery System.

Hvad måler undersøgelsen?

Primære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Device Success According to VARC-3 Criteria
Tidsramme: 30 days after the procedure

Technical success; Absence of mortality; Absence of surgery or intervention related to the device or to major vascular, access-related, or cardiac structural complications; Intended valve performance, defined as mean transvalvular gradient <20 mmHg, peak velocity <3 m/s, Doppler velocity index ≥0.25, and less than moderate aortic regurgitation.

30 days after the procedure

Sekundære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Percentage of Participants Achieving Technical Success According to VARC-3 Criteria
Tidsramme: During the procedure

Defined as freedom from mortality; successful vascular access, delivery of the device, and retrieval of the delivery system; correct positioning of a single prosthetic heart valve in the intended anatomical location; and freedom from surgery or intervention related to the device or to a major vascular, access-related, or cardiac structural complication. Results will be reported as the number and percentage of participants achieving technical success.

During the procedure
Incidence and Severity of Paravalvular Leak
Tidsramme: 30 days, 6 months, 12 months, and annually through 5 years

Paravalvular leak will be assessed by transthoracic echocardiography and classified according to severity. The number and percentage of participants with paravalvular leak in each severity category will be reported.

30 days, 6 months, 12 months, and annually through 5 years
Mean Transvalvular Aortic Pressure Gradient Assessed
Tidsramme: 30 days, 6 months, 12 months, and annually through 5 years

Mean transvalvular aortic pressure gradient will be assessed by transthoracic echocardiography and reported in mmHg.

30 days, 6 months, 12 months, and annually through 5 years
Distribution of Participants by New York Heart Association Functional Class
Tidsramme: 30 days, 6 months, 12 months, and annually through 5 years

Functional status will be assessed using the New York Heart Association (NYHA) Functional Classification. The classification ranges from Class I to Class IV, where Class I represents no limitation of physical activity and Class IV represents inability to perform physical activity without symptoms and/or symptoms at rest. Higher NYHA classes indicate worse functional status. Results will be reported as the number and percentage of participants in each NYHA functional class.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of New Permanent Pacemaker Implantation
Tidsramme: 30 days, 6 months, 12 months, and annually through 5 years

New permanent pacemaker implantation following the index transcatheter aortic valve implantation procedure will be assessed. Results will be reported as the number and percentage of participants who undergo new permanent pacemaker implantation.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of All-Cause Mortality
Tidsramme: 30 days, 6 months, 12 months, and annually through 5 years

All-cause mortality will be assessed throughout follow-up. Results will be reported as the number and percentage of participants who die from any cause.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of Stroke
Tidsramme: 30 days, 6 months, 12 months, and annually through 5 years

Stroke events will be assessed based on clinical neurological evaluation and classified according to Valve Academic Research Consortium 3 (VARC-3) definitions. Results will be reported as the number and percentage of participants experiencing stroke.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of Bioprosthetic Valve-Related Reintervention
Tidsramme: 30 days, 6 months, 12 months, and annually through 5 years

Bioprosthetic valve-related reintervention will be defined as any surgical or transcatheter intervention performed to treat dysfunction or failure of the implanted study valve. Results will be reported as the number and percentage of participants undergoing valve-related reintervention.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of Bioprosthetic Valve Dysfunction According to VARC-3 Definitions
Tidsramme: 30 days, 6 months, 12 months, and annually through 5 years

Bioprosthetic valve dysfunction will be assessed according to Valve Academic Research Consortium 3 (VARC-3) definitions. Bioprosthetic valve dysfunction includes structural valve deterioration, non-structural valve dysfunction, valve thrombosis, and endocarditis. Results will be reported as the number and percentage of participants meeting the criteria for bioprosthetic valve dysfunction and according to the corresponding category.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of Bioprosthetic Valve Failure According to VARC-3 Definitions
Tidsramme: 30 days, 6 months, 12 months, 24 months, 36 months, 48 months, and 60 months after the index procedure

Bioprosthetic valve failure will be assessed according to Valve Academic Research Consortium 3 (VARC-3) definitions. Results will be reported as the number and percentage of participants meeting the criteria for bioprosthetic valve failure.

30 days, 6 months, 12 months, 24 months, 36 months, 48 months, and 60 months after the index procedure

Samarbejdspartnere og efterforskere

Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.

Sponsor

Datoer for undersøgelser

Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.

Studer store datoer

Studiestart (Anslået)

1. november 2026

Primær færdiggørelse (Anslået)

1. juli 2027

Studieafslutning (Anslået)

1. juni 2032

Datoer for studieregistrering

Først indsendt

19. august 2026

Først indsendt, der opfyldte QC-kriterier

3. september 2026

Først opslået (Faktiske)

8. september 2026

Opdateringer af undersøgelsesjournaler

Sidste opdatering sendt (Faktiske)

19. november 2026

Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier

16. november 2026

Sidst verificeret

1. november 2026

Mere information

Begreber relateret til denne undersøgelse

Andre undersøgelses-id-numre

  • SAFESYNC NG

Plan for individuelle deltagerdata (IPD)

Planlægger du at dele individuelle deltagerdata (IPD)?

INGEN

Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter

Studerer et amerikansk FDA-reguleret lægemiddelprodukt

Ingen

Studerer et amerikansk FDA-reguleret enhedsprodukt

Ingen

Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .