Safety and Clinical Performance of the Next-Generation SAFESYNC Transcatheter Heart Valve in Patients With Severe Aortic Stenosis (SAFESYNC NG)

maanantai 16. marraskuuta 2026 päivittänyt: Braile Biomedica Ind. Com. e Repr. Ltda.

Evaluation of the Safety and Clinical Performance of the Next-Generation SAFESYNC Balloon-Expandable Transcatheter Heart Valve in Patients With Severe Aortic Stenosis: A Prospective, Multicenter, Single-Arm Pivotal Study

This pivotal, prospective, multicenter (Brazil, Chile and Argentina), single-arm study will evaluate the safety and clinical performance of the Inovare® Safesync Next-Generation bioprosthetic heart valve, implanted by transfemoral transcatheter aortic valve implantation (TAVI) with the B-Delivery System, in patients aged 70 years or older with severe aortic stenosis and an indication for TAVI, regardless of surgical risk. The study will enroll 210 participants.

Tutkimuksen yleiskatsaus

Tila

Ei vielä rekrytointia

Ehdot

Interventio / Hoito

Opintotyyppi

Interventio

Ilmoittautuminen (Arvioitu)

210

Vaihe

  • Ei sovellettavissa

Yhteystiedot ja paikat

Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.

Opiskeluyhteys

  • Nimi: Clinical Research Department - Braile Biomédica
  • Puhelinnumero: +5517 21367000
  • Sähköposti: pesquisa@braile.com.br

Osallistumiskriteerit

Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.

Kelpoisuusvaatimukset

Opintokelpoiset iät

70 vuotta ja vanhemmat (Vanhempi Aikuinen)

Hyväksyy terveitä vapaaehtoisia

Ei

Sukupuolet, jotka voivat opiskella

Kaikki

Kuvaus

Inclusion Criteria:

Severe aortic valve stenosis diagnosed according to criteria established in current guidelines; Age ≥70 years, regardless of surgical risk; Indication for transcatheter aortic valve implantation (TAVI); Willing to participate and to sign the written informed consent form.

Exclusion Criteria:

Any condition considered a contraindication to implantation of a bioprosthetic heart valve (e.g., patient with an indication for a mechanical heart valve);

Known hypersensitivity or contraindication to any of the following that cannot be adequately managed with premedication:

  1. Aspirin, clopidogrel, or heparin (heparin-induced thrombocytopenia [HIT] or heparin-induced thrombocytopenia with thrombosis [HITTS]);
  2. Nitinol (titanium or nickel);
  3. Contrast media; Blood dyscrasias: leukopenia (white blood cell count <1,000/mm³), thrombocytopenia (platelet count <100,000 cells/mm³), history of coagulopathy, or hypercoagulable states; Active infection, including endocarditis; Cardiogenic shock manifested by low cardiac output, vasopressor dependence, or need for mechanical hemodynamic support; Recent stroke or transient ischemic attack (within 6 months); History of gastrointestinal bleeding or other bleeding that precludes anticoagulant or antiplatelet therapy; Estimated life expectancy of less than 5 years; Other medical, social, or psychological conditions that, in the investigator's opinion, prevent the patient from providing consent or adequately complying with the protocol; Pre-existing prosthetic heart valve in the aortic position (biological or mechanical); Concomitant severe mitral or tricuspid valve disease; Hypertrophic obstructive cardiomyopathy; Percutaneous coronary intervention performed within 72 hours prior to the study procedure; Femoral artery diameter <5.5 mm (for Safesync 20 to 24 mm), <6.0 mm (for Safesync 26 and 28 mm), or <6.5 mm (for Safesync 30 mm); High risk of annular rupture or significant residual paravalvular leak due to severe calcification of the aortic valve or left ventricular outflow tract; High risk of coronary obstruction (coronary ostial height <10 mm with narrow sinuses of Valsalva [SOV], defined as a mean SOV diameter <3 mm larger than the bioprosthesis size), as determined by computed tomography angiography; Left ventricular dysfunction with left ventricular ejection fraction (LVEF) ≤30%; Severe chronic kidney disease, defined as glomerular filtration rate <30 mL/min/1.73 m².

Opintosuunnitelma

Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.

Miten tutkimus on suunniteltu?

Suunnittelun yksityiskohdat

  • Ensisijainen käyttötarkoitus: Hoito
  • Jako: Ei käytössä
  • Inventiomalli: Yksittäinen ryhmätehtävä
  • Naamiointi: Ei mitään (avoin tarra)

Aseet ja interventiot

Osallistujaryhmä / Arm
Interventio / Hoito
Kokeellinen: INOVARE® SAFESYNC Next-Generation TAVI

Participants will undergo TAVI via transfemoral access using the INOVARE® SAFESYNC Next-Generation Bioprosthetic Heart Valve and the B-Delivery System.

Mitä tutkimuksessa mitataan?

Ensisijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Device Success According to VARC-3 Criteria
Aikaikkuna: 30 days after the procedure

Technical success; Absence of mortality; Absence of surgery or intervention related to the device or to major vascular, access-related, or cardiac structural complications; Intended valve performance, defined as mean transvalvular gradient <20 mmHg, peak velocity <3 m/s, Doppler velocity index ≥0.25, and less than moderate aortic regurgitation.

30 days after the procedure

Toissijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Percentage of Participants Achieving Technical Success According to VARC-3 Criteria
Aikaikkuna: During the procedure

Defined as freedom from mortality; successful vascular access, delivery of the device, and retrieval of the delivery system; correct positioning of a single prosthetic heart valve in the intended anatomical location; and freedom from surgery or intervention related to the device or to a major vascular, access-related, or cardiac structural complication. Results will be reported as the number and percentage of participants achieving technical success.

During the procedure
Incidence and Severity of Paravalvular Leak
Aikaikkuna: 30 days, 6 months, 12 months, and annually through 5 years

Paravalvular leak will be assessed by transthoracic echocardiography and classified according to severity. The number and percentage of participants with paravalvular leak in each severity category will be reported.

30 days, 6 months, 12 months, and annually through 5 years
Mean Transvalvular Aortic Pressure Gradient Assessed
Aikaikkuna: 30 days, 6 months, 12 months, and annually through 5 years

Mean transvalvular aortic pressure gradient will be assessed by transthoracic echocardiography and reported in mmHg.

30 days, 6 months, 12 months, and annually through 5 years
Distribution of Participants by New York Heart Association Functional Class
Aikaikkuna: 30 days, 6 months, 12 months, and annually through 5 years

Functional status will be assessed using the New York Heart Association (NYHA) Functional Classification. The classification ranges from Class I to Class IV, where Class I represents no limitation of physical activity and Class IV represents inability to perform physical activity without symptoms and/or symptoms at rest. Higher NYHA classes indicate worse functional status. Results will be reported as the number and percentage of participants in each NYHA functional class.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of New Permanent Pacemaker Implantation
Aikaikkuna: 30 days, 6 months, 12 months, and annually through 5 years

New permanent pacemaker implantation following the index transcatheter aortic valve implantation procedure will be assessed. Results will be reported as the number and percentage of participants who undergo new permanent pacemaker implantation.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of All-Cause Mortality
Aikaikkuna: 30 days, 6 months, 12 months, and annually through 5 years

All-cause mortality will be assessed throughout follow-up. Results will be reported as the number and percentage of participants who die from any cause.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of Stroke
Aikaikkuna: 30 days, 6 months, 12 months, and annually through 5 years

Stroke events will be assessed based on clinical neurological evaluation and classified according to Valve Academic Research Consortium 3 (VARC-3) definitions. Results will be reported as the number and percentage of participants experiencing stroke.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of Bioprosthetic Valve-Related Reintervention
Aikaikkuna: 30 days, 6 months, 12 months, and annually through 5 years

Bioprosthetic valve-related reintervention will be defined as any surgical or transcatheter intervention performed to treat dysfunction or failure of the implanted study valve. Results will be reported as the number and percentage of participants undergoing valve-related reintervention.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of Bioprosthetic Valve Dysfunction According to VARC-3 Definitions
Aikaikkuna: 30 days, 6 months, 12 months, and annually through 5 years

Bioprosthetic valve dysfunction will be assessed according to Valve Academic Research Consortium 3 (VARC-3) definitions. Bioprosthetic valve dysfunction includes structural valve deterioration, non-structural valve dysfunction, valve thrombosis, and endocarditis. Results will be reported as the number and percentage of participants meeting the criteria for bioprosthetic valve dysfunction and according to the corresponding category.

30 days, 6 months, 12 months, and annually through 5 years
Incidence of Bioprosthetic Valve Failure According to VARC-3 Definitions
Aikaikkuna: 30 days, 6 months, 12 months, 24 months, 36 months, 48 months, and 60 months after the index procedure

Bioprosthetic valve failure will be assessed according to Valve Academic Research Consortium 3 (VARC-3) definitions. Results will be reported as the number and percentage of participants meeting the criteria for bioprosthetic valve failure.

30 days, 6 months, 12 months, 24 months, 36 months, 48 months, and 60 months after the index procedure

Yhteistyökumppanit ja tutkijat

Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.

Sponsori

Opintojen ennätyspäivät

Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan ​​julkisella verkkosivustolla.

Opi tärkeimmät päivämäärät

Opiskelun aloitus (Arvioitu)

Sunnuntai 1. marraskuuta 2026

Ensisijainen valmistuminen (Arvioitu)

Torstai 1. heinäkuuta 2027

Opintojen valmistuminen (Arvioitu)

Tiistai 1. kesäkuuta 2032

Opintoihin ilmoittautumispäivät

Ensimmäinen lähetetty

Keskiviikko 19. elokuuta 2026

Ensimmäinen toimitettu, joka täytti QC-kriteerit

Torstai 3. syyskuuta 2026

Ensimmäinen Lähetetty (Todellinen)

Tiistai 8. syyskuuta 2026

Tutkimustietojen päivitykset

Viimeisin päivitys julkaistu (Todellinen)

Torstai 19. marraskuuta 2026

Viimeisin lähetetty päivitys, joka täytti QC-kriteerit

Maanantai 16. marraskuuta 2026

Viimeksi vahvistettu

Sunnuntai 1. marraskuuta 2026

Lisää tietoa

Tähän tutkimukseen liittyvät termit

Muut tutkimustunnusnumerot

  • SAFESYNC NG

Yksittäisten osallistujien tietojen suunnitelma (IPD)

Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?

EI

Lääke- ja laitetiedot, tutkimusasiakirjat

Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta

Ei

Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta

Ei

Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .