Feasibility and Usability of a Chest-Strap Ambulatory ECG Monitor in Active Individuals in Comparison With a Traditional Ambulatory ECG Patch Monitor (MoMe ARC)
Panoramica dello studio
Stato
Stato
Condizioni
Condizioni
Intervento / Trattamento
Intervento / Trattamento
Tipo di studio
Tipo di studio
Iscrizione (Stimato)
Iscrizione
Contatti e Sedi
Contatto studio
Contatto studio
- Nome: Elizabeth Dineen, DO
- Numero di telefono: 904-953-5230
- Email: dineen.elizabeth@mayo.edu
Luoghi di studio
-
-
Arizona
-
Scottsdale, Arizona, Stati Uniti, 85259
- Reclutamento
- Mayo Clinic Arizona
-
Contatto:
- Satyajit Reddy, MD
-
-
Florida
-
Jacksonville, Florida, Stati Uniti, 32224
- Reclutamento
- Mayo Clinic Florida
-
Contatto:
- Elizabeth Dineen, DO
-
-
Minnesota
-
Rochester, Minnesota, Stati Uniti, 55905
- Reclutamento
- Mayo Clinic Rochester
-
Contatto:
- Kathryn Larson, MD
- Numero di telefono: 507-293-6358
-
-
Criteri di partecipazione
Criteri di ammissibilità
Criteri di ammissibilità
Età idonea allo studio
- Adulto
- Adulto più anziano
Accetta volontari sani
Metodo di campionamento
Popolazione di studio
Descrizione
Inclusion Criteria:
- Adult participants (>22 years; Frontier X is approved for use in this age group) who participate in >4 days a week of >150 minutes total per week of moderate to high intensity exercise who were prescribed a ambulatory ECG monitor for clinical purposes for 24 hours or more.
- Patients must agree to wear the Frontier X monitor for the prescribed duration and use the Frontier X Plus application
Exclusion Criteria:
- Current use of Implanted Cardiac Devices
- Pregnancy
- Known allergies to any of the materials listed in the "List of Patient-Contacting Materials"
- Presence of skin injuries or broken skin in the area where the chest strap is worn
- Inability to use a compatible mobile device required for the operation of the study device
- Any cognitive or physical limitations that, in the opinion of the investigator, limits the participant's ability to fully follow study procedures
- Medical restrictions on exercise for safety including cardiac, orthopedic or other relevant past medical history
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
Numero di gruppi/coorti
Coorti e interventi
Gruppo / CoorteGruppo / Coorte |
Intervento / TrattamentoIntervento / Trattamento |
|---|---|
|
Active Adults Undergoing Ambulatory ECG Monitoring
Participants prescribed clinical ambulatory ECG monitoring will wear the Frontier X Plus chest-strap ECG device simultaneously with the MoMe ARC traditional ambulatory ECG monitor during their prescribed monitoring period and typical daily activities, including exercise.
|
The Frontier X Plus is a chest-strap-based wearable ECG device with embedded dry electrodes and a sensor that captures, stores, analyzes, and transmits ECG data through a mobile application and cloud-based platform for subsequent review.
The MoMe ARC monitor will be used as the traditional ambulatory ECG comparator and processed according to standard clinical workflows.
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Concordance of Symptomatic Triggered Events Between Frontier X Plus and MoMe ARC
Lasso di tempo: 24 hours
|
Automated Holter-style reports generated from the Frontier X Plus device will be compared with MoMe ARC reports.
Concordance will be assessed for symptomatic triggered events.
Results will be reported as the number of symptomatic triggered event findings for which the Frontier X Plus and MoMe ARC reports are in agreement.
|
24 hours
|
|
Concordance of Ventricular Tachycardia Episodes Greater Than 3 Beats Detected by Frontier X Plus and MoMe ARC
Lasso di tempo: 24 hours
|
Automated Holter-style reports generated from the Frontier X Plus device will be compared with MoMe ARC reports.
Concordance will be assessed for ventricular tachycardia episodes greater than 3 beats.
Results will be reported as the number of ventricular tachycardia episode findings for which the Frontier X Plus and MoMe ARC reports are in agreement.
|
24 hours
|
|
Concordance of Supraventricular Tachycardia Episodes Greater Than 3 Beats Between Frontier X Plus and MoMe ARC
Lasso di tempo: 24 hours
|
Automated Holter-style reports generated from the Frontier X Plus device will be compared with MoMe ARC reports.
Concordance will be assessed for supraventricular tachycardia episodes greater than 3 beats.
Results will be reported as the number of supraventricular tachycardia episode findings for which the Frontier X Plus and MoMe ARC reports are in agreement.
|
24 hours
|
|
Total Time in Atrial Fibrillation Detected by Frontier X Plus and MoMe ARC
Lasso di tempo: 24 hours
|
Automated Holter-style reports generated from the Frontier X Plus device will be compared with MoMe ARC reports.
Total time in atrial fibrillation will be assessed for each device and reported as the cumulative duration of atrial fibrillation detected during the concordant monitoring period.
|
24 hours
|
|
Concordance of Premature Atrial Contraction Counts Between Frontier X Plus and MoMe ARC
Lasso di tempo: 24 hours
|
Automated Holter-style reports generated from the Frontier X Plus device will be compared with MoMe ARC reports.
Concordance will be assessed for premature atrial contraction counts.
Results will be reported as the number of premature atrial contraction count findings for which the Frontier X Plus and MoMe ARC reports are in agreement.
|
24 hours
|
|
Concordance of Premature Ventricular Contraction Counts Between Frontier X Plus and MoMe ARC
Lasso di tempo: 24 hours
|
Automated Holter-style reports generated from the Frontier X Plus device will be compared with MoMe ARC reports.
Concordance will be assessed for premature ventricular contraction counts.
Results will be reported as the number of premature ventricular contraction count findings for which the Frontier X Plus and MoMe ARC reports are in agreement.
|
24 hours
|
|
Atrial Fibrillation Burden Detected by Frontier X Plus and MoMe ARC
Lasso di tempo: 24 hours
|
Automated Holter-style reports generated from the Frontier X Plus device will be compared with MoMe ARC reports.
Atrial fibrillation burden will be assessed for each device and reported as the percentage of the concordant monitoring period during which atrial fibrillation was detected.
|
24 hours
|
Collaboratori e investigatori
Sponsor
Sponsor
Investigatori
Investigatori
- Investigatore principale: Elizabeth Dineen, DO, Mayo Clinic
Studiare le date dei record
Studia le date principali
Inizio studio (Stimato)
Inizio studio
Completamento primario (Stimato)
Completamento primario
Completamento dello studio (Stimato)
Completamento dello studio
Date di iscrizione allo studio
Primo inviato
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Primo Inserito
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento pubblicato
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
Altri numeri di identificazione dello studio
- 25-012842
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
prodotto fabbricato ed esportato dagli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .