Medicontur preloaded hydrophobic monofocal toric intraocular lenses (IOLs) are indicated to improve vision at far distance in adults with or without cataract and correction of pre-existing corneal astigmatism secondarily to removal of the crystalline lens.
The intraocular lens is intended to be surgically implanted into the eye with the purpose of restoring optical function in the aphakic eye to provide an optical system with high predictability of the precalculated dioptric power.
The investigation will be performed as a prospective, non-comparative, single arm study with a single center design. Patients will be enrolled who are assigned to be implanted with the 877PTY IOLs mono- or binocularly between January 2024 - June 2025.
Data from six visits will be collected:
- Visit 1: Screening and Baseline (up to 90 days prior to the surgery)
- Visit 2: IOL implantation (Day 0)
- Visit 3: Day 1 post-operatively (+/- 0 day)
- Visit 4: Day 7 post-operatively (+/- 3 days)
- Visit 5: Month 1 post-operatively (+/- 2 weeks)
- Visit 6: Month 6 post-operatively (+/- 1 month)
調査の概要
研究の種類
入学 (実際)
連絡先と場所
研究場所
-
-
-
Győr、ハンガリー、9024
- Győr-Moson-Sopron Vármegyei Petz Aladár Egyetemi Oktató Kórház
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
サンプリング方法
調査対象母集団
Aphakic adult patients (18 years old and older), with cataract and corneal astigmatism who aspire to have improved uncorrected distance vision and reduction of the residual refractive cylinder and assigned to monocular or binocular implantation with 877PTY IOL.
説明
Inclusion Criteria:
- adult patients (18 years old and older);
- cataract and/or corneal astigmatism diagnosis;
- indication for cataract surgery or refractive lens exchange;
- recommended cylinder correction with toric IOL with cylinder power ≥1.0D;
- clear intraocular media other than cataract;
- signed informed consent form;
Exclusion Criteria:
- irregular astigmatism;
- patients who are not targeted to emmetropia (predicted residual spherical equivalent and/or astigmatism is out of the range of ±0.5D and ±0.38D respectively);
- any retinopathy or maculopathy that affects the vision;
- iris neovascularization;
- congenital eye abnormality affecting visual performance;
- advanced glaucoma;
- pseudoexfoliation syndrome affecting IOL stability;
- amblyopia;
- uveitis;
- retinal detachment;
- prior ocular surgery in personal medical history;
- irregular corneal curvature or corneal diseases affecting visual performance;
- high myopia (axial length ≥ 26,5 mm);
- inadequate visualization of the fundus on preoperative examination;
- dilated pupil size not large enough to visualize toric IOL axis markings postoperatively based on the Investigator's decision;
- eye trauma in medical history;
- instability of keratometry or biometry measurements;
- prior corneal refractive surgery such as LASIK, PRK, or SMILE;
- patients deemed ineligible by the investigator because of any systemic disease or treatment;
- pregnancy or lactation;
- current use of systemic steroids or external ophthalmic drugs;
- concurrent participation in another drug or device investigation.
Intraoperative exclusions: the presence of the following intraoperative exclusion criteria will lead to the exclusion of the subject:
- Any unexpected or serious intraoperative complication that makes IOL implantation impossible or results in implantation of another type of IOL that is outside the scope of this clinical trial.
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 観測モデル:CASE_CONTROL
- 時間の展望:PROSPECTIVE
コホートと介入
グループ/コホート |
介入・治療 |
|---|---|
|
877PTY
100 patients/eyes implanted with 877PTY IOL |
IOL implantation with 6-month follow-up period Monocular UDVA/CDVA Manifest Refraction (SPH, CYL, SEQ) Optical Biometry, Keratometry (K1, K2, AXL, ACD) IOL axis orientation determination Visual function questionnaire Glistening evaluation Tonometry Slit-Lamp examination Fundus visualization |
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Primary Performance Endpoint - CDVA
時間枠:6 months postoperatively
|
Visual acuity: monocular CDVA at 6 months after IOL implantation |
6 months postoperatively
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Secondary Performance Endpoint - UDVA
時間枠:6 months postoperatively
|
Visual acuity: monocular UDVA at 6 months after IOL implantation |
6 months postoperatively
|
|
Secondary Performance Endpoint - Subjective CYL
時間枠:6 months postoperatively
|
Subjective (manifest) cylinder and keratometric cylinder of the eye at 6 months postoperatively |
6 months postoperatively
|
|
Secondary Performance Endpoint - Residual CYL and SEQ
時間枠:6 months postoperatively
|
Residual manifest cylinder and spherical equivalent at 6 months postoperatively; |
6 months postoperatively
|
|
Secondary Performance Endpoint - Rotational stability
時間枠:6 months postoperatively
|
IOL rotational stability at 6 months after IOL implantation; |
6 months postoperatively
|
|
Secondary Performance Endpoint - Injector handling
時間枠:6 months postoperatively
|
Ease of IOL implantation and positioning at surgery; |
6 months postoperatively
|
|
Secondary Performance Endpoint - Patient satisfaction (VFQ-25)
時間枠:6 months postoperatively
|
Patient-reported visual functions: overall global vision rating, difficulty with near and distance vision activities, social functioning and role limitations, dependency on others, mental health symptoms due to vision, driving difficulties, limitations with peripheral and color vision, and ocular pain evaluated by visual function questionnaire at 6 months after IOL implantation. |
6 months postoperatively
|
|
Secondary Performance Endpoint - Glistening
時間枠:6 months postoperatively
|
Glistening rate and severity at 6 months after IOL implantation |
6 months postoperatively
|
その他の成果指標
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Safety Endpoints - Intraoperative complications
時間枠:Up to day 1 postoperatively
|
Intraoperative complications of cataract surgery
|
Up to day 1 postoperatively
|
|
Safety Endpoints - Postoperative complications
時間枠:Up to month 1 postoperatively
|
Postoperative complications of cataract surgery
|
Up to month 1 postoperatively
|
|
Safety Endpoints - Complications of IOL implantation
時間枠:Up to month 6 postoperatively
|
Complications of IOL implantation
|
Up to month 6 postoperatively
|