Tuleva kliininen tutkimus 877PTY Toric IOL:n turvallisuuden ja suorituskyvyn arvioimiseksi sarveiskalvon hajataittoisuuden korjaamiseksi
Medicontur preloaded hydrophobic monofocal toric intraocular lenses (IOLs) are indicated to improve vision at far distance in adults with or without cataract and correction of pre-existing corneal astigmatism secondarily to removal of the crystalline lens.
The intraocular lens is intended to be surgically implanted into the eye with the purpose of restoring optical function in the aphakic eye to provide an optical system with high predictability of the precalculated dioptric power.
The investigation will be performed as a prospective, non-comparative, single arm study with a single center design. Patients will be enrolled who are assigned to be implanted with the 877PTY IOLs mono- or binocularly between January 2024 - June 2025.
Data from six visits will be collected:
- Visit 1: Screening and Baseline (up to 90 days prior to the surgery)
- Visit 2: IOL implantation (Day 0)
- Visit 3: Day 1 post-operatively (+/- 0 day)
- Visit 4: Day 7 post-operatively (+/- 3 days)
- Visit 5: Month 1 post-operatively (+/- 2 weeks)
- Visit 6: Month 6 post-operatively (+/- 1 month)
Tutkimuksen yleiskatsaus
Opintotyyppi
Ilmoittautuminen (Todellinen)
Yhteystiedot ja paikat
Opiskelupaikat
-
-
-
Győr, Unkari, 9024
- Győr-Moson-Sopron Vármegyei Petz Aladár Egyetemi Oktató Kórház
-
-
Osallistumiskriteerit
Kelpoisuusvaatimukset
Opintokelpoiset iät
Hyväksyy terveitä vapaaehtoisia
Sukupuolet, jotka voivat opiskella
Näytteenottomenetelmä
Tutkimusväestö
Aphakic adult patients (18 years old and older), with cataract and corneal astigmatism who aspire to have improved uncorrected distance vision and reduction of the residual refractive cylinder and assigned to monocular or binocular implantation with 877PTY IOL.
Kuvaus
Inclusion Criteria:
- adult patients (18 years old and older);
- cataract and/or corneal astigmatism diagnosis;
- indication for cataract surgery or refractive lens exchange;
- recommended cylinder correction with toric IOL with cylinder power ≥1.0D;
- clear intraocular media other than cataract;
- signed informed consent form;
Exclusion Criteria:
- irregular astigmatism;
- patients who are not targeted to emmetropia (predicted residual spherical equivalent and/or astigmatism is out of the range of ±0.5D and ±0.38D respectively);
- any retinopathy or maculopathy that affects the vision;
- iris neovascularization;
- congenital eye abnormality affecting visual performance;
- advanced glaucoma;
- pseudoexfoliation syndrome affecting IOL stability;
- amblyopia;
- uveitis;
- retinal detachment;
- prior ocular surgery in personal medical history;
- irregular corneal curvature or corneal diseases affecting visual performance;
- high myopia (axial length ≥ 26,5 mm);
- inadequate visualization of the fundus on preoperative examination;
- dilated pupil size not large enough to visualize toric IOL axis markings postoperatively based on the Investigator's decision;
- eye trauma in medical history;
- instability of keratometry or biometry measurements;
- prior corneal refractive surgery such as LASIK, PRK, or SMILE;
- patients deemed ineligible by the investigator because of any systemic disease or treatment;
- pregnancy or lactation;
- current use of systemic steroids or external ophthalmic drugs;
- concurrent participation in another drug or device investigation.
Intraoperative exclusions: the presence of the following intraoperative exclusion criteria will lead to the exclusion of the subject:
- Any unexpected or serious intraoperative complication that makes IOL implantation impossible or results in implantation of another type of IOL that is outside the scope of this clinical trial.
Opintosuunnitelma
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
- Havaintomallit: CASE_CONTROL
- Aikanäkymät: PROSPECTIVE
Kohortit ja interventiot
Ryhmä/Kohortti |
Interventio / Hoito |
|---|---|
|
877PTY
100 patients/eyes implanted with 877PTY IOL |
IOL implantation with 6-month follow-up period Monocular UDVA/CDVA Manifest Refraction (SPH, CYL, SEQ) Optical Biometry, Keratometry (K1, K2, AXL, ACD) IOL axis orientation determination Visual function questionnaire Glistening evaluation Tonometry Slit-Lamp examination Fundus visualization |
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Primary Performance Endpoint - CDVA
Aikaikkuna: 6 months postoperatively
|
Visual acuity: monocular CDVA at 6 months after IOL implantation |
6 months postoperatively
|
Toissijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Secondary Performance Endpoint - UDVA
Aikaikkuna: 6 months postoperatively
|
Visual acuity: monocular UDVA at 6 months after IOL implantation |
6 months postoperatively
|
|
Secondary Performance Endpoint - Subjective CYL
Aikaikkuna: 6 months postoperatively
|
Subjective (manifest) cylinder and keratometric cylinder of the eye at 6 months postoperatively |
6 months postoperatively
|
|
Secondary Performance Endpoint - Residual CYL and SEQ
Aikaikkuna: 6 months postoperatively
|
Residual manifest cylinder and spherical equivalent at 6 months postoperatively; |
6 months postoperatively
|
|
Secondary Performance Endpoint - Rotational stability
Aikaikkuna: 6 months postoperatively
|
IOL rotational stability at 6 months after IOL implantation; |
6 months postoperatively
|
|
Secondary Performance Endpoint - Injector handling
Aikaikkuna: 6 months postoperatively
|
Ease of IOL implantation and positioning at surgery; |
6 months postoperatively
|
|
Secondary Performance Endpoint - Patient satisfaction (VFQ-25)
Aikaikkuna: 6 months postoperatively
|
Patient-reported visual functions: overall global vision rating, difficulty with near and distance vision activities, social functioning and role limitations, dependency on others, mental health symptoms due to vision, driving difficulties, limitations with peripheral and color vision, and ocular pain evaluated by visual function questionnaire at 6 months after IOL implantation. |
6 months postoperatively
|
|
Secondary Performance Endpoint - Glistening
Aikaikkuna: 6 months postoperatively
|
Glistening rate and severity at 6 months after IOL implantation |
6 months postoperatively
|
Muut tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Safety Endpoints - Intraoperative complications
Aikaikkuna: Up to day 1 postoperatively
|
Intraoperative complications of cataract surgery
|
Up to day 1 postoperatively
|
|
Safety Endpoints - Postoperative complications
Aikaikkuna: Up to month 1 postoperatively
|
Postoperative complications of cataract surgery
|
Up to month 1 postoperatively
|
|
Safety Endpoints - Complications of IOL implantation
Aikaikkuna: Up to month 6 postoperatively
|
Complications of IOL implantation
|
Up to month 6 postoperatively
|
Yhteistyökumppanit ja tutkijat
Sponsori
Opintojen ennätyspäivät
Opi tärkeimmät päivämäärät
Opiskelun aloitus (Todellinen)
Ensisijainen valmistuminen (Todellinen)
Opintojen valmistuminen (Todellinen)
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen Lähetetty (Todellinen)
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Todellinen)
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeksi vahvistettu
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Muita asiaankuuluvia MeSH-ehtoja
Muut tutkimustunnusnumerot
- M-877PTY-HU-2301
Lääke- ja laitetiedot, tutkimusasiakirjat
Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta
Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .