自動圧平眼圧測定 - 更新

2026年11月2日 更新者:Duke University

自動圧平眼圧測定 - 更新版

Goldmann Applanation Tonometry (GAT) is considered the clinical gold standard for eye pressure measurements and yet it is known to be a subjective measurement with limited repeatability and limited portability. Another clinical standard for checking eye pressure is known as the pneumotonometer. This method is more objective but not portable. The purpose of this study is to develop new methods of measuring eye pressure that are more objective, reproducible and portable. In this study, the investigators will be comparing the eye pressure measurements using 2 investigational methods to GAT and pneuumotonometer.

調査の概要

詳細な説明

This is a prospective study. Patients presenting for their scheduled eye appointments will be recruited by verbal communication. Following informed consent, intraocular pressure (IOP) measurements will be made by the following methods:

Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.

Pneumotonometer: This is what is done in routine eye care. The participant will receive a topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure.

Fixed force GAT: From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a complementary metal oxide semiconductor (C-MOS) camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. A similar method using an iPod touch camera has been previously published by this PI.

Upright applanating prototype: With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.

The order of the above 4 measurements will be randomized.

研究の種類

介入

入学 (推定)

200

段階

  • 適用できない

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

研究場所

    • North Carolina
      • Durham、North Carolina、アメリカ、27710
        • 募集
        • Duke Eye Center
        • 主任研究者:
          • Joanne Wen, MD
        • コンタクト:

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年歳以上 (大人、高齢者)

健康ボランティアの受け入れ

いいえ

受講資格のある性別

全て

説明

Criteria:

Inclusion Criteria:

  • Presenting for a routine eye exam
  • ≥ 18 years of age
  • Able and willing to give consent

Exclusion Criteria:

  • History of corneal scarring
  • Active infection of the eye
  • Ocular surface trauma or infection

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:診断
  • 割り当て:ランダム化
  • 介入モデル:並列代入
  • マスキング:なし(オープンラベル)

武器と介入

参加者グループ / アーム
介入・治療
アクティブコンパレータ:Standard GAT

This is the gold-standard method for IOP measurement in clinical practice

Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice

実験的:Fixed-force GAT

From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter

An investigational device similar to standard GAT

実験的:Upright applanating prototype

With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter

An investigational device, prism used in standard and fixed-force GAT is attached to a portable device

アクティブコンパレータ:Pneumotonometer

This is another standard method for IOP measurement in clinical practice

This is a standard for checking eye pressure. A topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
仰臥位 Pneumotonometer により測定される眼圧
時間枠:ベースライン
ベースライン
Intraocular pressure (IOP) as measured by standard GAT
時間枠:Baseline

this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.

Baseline
Intraocular pressure (IOP) as measured by fixed force GAT
時間枠:Baseline

From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular.

Baseline
Intraocular pressure as measured by upright applanating
時間枠:Baseline

With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.

Baseline

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

捜査官

  • 主任研究者:Joanne Wen, MD、Duke Eye Center

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (実際)

2026年3月2日

一次修了 (推定)

2026年12月29日

研究の完了 (推定)

2026年12月29日

試験登録日

最初に提出

2025年12月19日

QC基準を満たした最初の提出物

2025年12月19日

最初の投稿 (実際)

2025年12月23日

学習記録の更新

投稿された最後の更新 (実際)

2026年11月3日

QC基準を満たした最後の更新が送信されました

2026年11月2日

最終確認日

2026年11月1日

詳しくは

本研究に関する用語

追加の関連 MeSH 用語

その他の研究ID番号

  • Pro00106111_1

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

いいえ

IPD プランの説明

No Currently, there is no plan to share data with other researchers.

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

はい

米国で製造され、米国から輸出された製品。