Automatiserad applanationstonometri - Uppdaterad
Goldmann Applanation Tonometry (GAT) is considered the clinical gold standard for eye pressure measurements and yet it is known to be a subjective measurement with limited repeatability and limited portability. Another clinical standard for checking eye pressure is known as the pneumotonometer. This method is more objective but not portable. The purpose of this study is to develop new methods of measuring eye pressure that are more objective, reproducible and portable. In this study, the investigators will be comparing the eye pressure measurements using 2 investigational methods to GAT and pneuumotonometer.
Studieöversikt
Status
Betingelser
Intervention / Behandling
Detaljerad beskrivning
This is a prospective study. Patients presenting for their scheduled eye appointments will be recruited by verbal communication. Following informed consent, intraocular pressure (IOP) measurements will be made by the following methods:
Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.
Pneumotonometer: This is what is done in routine eye care. The participant will receive a topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure.
Fixed force GAT: From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a complementary metal oxide semiconductor (C-MOS) camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. A similar method using an iPod touch camera has been previously published by this PI.
Upright applanating prototype: With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.
The order of the above 4 measurements will be randomized.
Studietyp
Inskrivning (Beräknad)
Fas
- Inte tillämpbar
Kontakter och platser
Studiekontakt
- Namn: Sarah Jones
- Telefonnummer: 919 681 6584
- E-post: sarah.jones1@duke.edu
Studieorter
-
-
North Carolina
-
Durham, North Carolina, Förenta staterna, 27710
- Rekrytering
- Duke Eye Center
-
Huvudutredare:
- Joanne Wen, MD
-
Kontakt:
- Joanne Wen, MD
- Telefonnummer: +1 919 684 2975
- E-post: joanne.wen@duke.edu
-
-
Deltagandekriterier
Urvalskriterier
Åldrar som är berättigade till studier
Tar emot friska volontärer
Kön som är behöriga för studier
Beskrivning
Criteria:
Inclusion Criteria:
- Presenting for a routine eye exam
- ≥ 18 years of age
- Able and willing to give consent
Exclusion Criteria:
- History of corneal scarring
- Active infection of the eye
- Ocular surface trauma or infection
Studieplan
Hur är studien utformad?
Designdetaljer
- Primärt syfte: Diagnostisk
- Tilldelning: Randomiserad
- Interventionsmodell: Parallellt uppdrag
- Maskning: Ingen (Open Label)
Vapen och interventioner
Deltagargrupp / Arm |
Intervention / Behandling |
|---|---|
|
Aktiv komparator: Standard GAT
This is the gold-standard method for IOP measurement in clinical practice |
Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice |
|
Experimentell: Fixed-force GAT
From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter |
An investigational device similar to standard GAT |
|
Experimentell: Upright applanating prototype
With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter |
An investigational device, prism used in standard and fixed-force GAT is attached to a portable device |
|
Aktiv komparator: Pneumotonometer
This is another standard method for IOP measurement in clinical practice |
This is a standard for checking eye pressure. A topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure |
Vad mäter studien?
Primära resultatmått
Resultatmått |
Åtgärdsbeskrivning |
Tidsram |
|---|---|---|
|
Intraokulärt tryck som mäts med liggande pneumotonometer
Tidsram: Baslinje
|
Baslinje
|
|
|
Intraocular pressure (IOP) as measured by standard GAT
Tidsram: Baseline
|
this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial. |
Baseline
|
|
Intraocular pressure (IOP) as measured by fixed force GAT
Tidsram: Baseline
|
From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. |
Baseline
|
|
Intraocular pressure as measured by upright applanating
Tidsram: Baseline
|
With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. |
Baseline
|
Samarbetspartners och utredare
Sponsor
Utredare
- Huvudutredare: Joanne Wen, MD, Duke Eye Center
Studieavstämningsdatum
Studera stora datum
Studiestart (Faktisk)
Primärt slutförande (Beräknad)
Avslutad studie (Beräknad)
Studieregistreringsdatum
Först inskickad
Först inskickad som uppfyllde QC-kriterierna
Första postat (Faktisk)
Uppdateringar av studier
Senaste uppdatering publicerad (Faktisk)
Senaste inskickade uppdateringen som uppfyllde QC-kriterierna
Senast verifierad
Mer information
Termer relaterade till denna studie
Ytterligare relevanta MeSH-villkor
Andra studie-ID-nummer
- Pro00106111_1
Plan för individuella deltagardata (IPD)
Planerar du att dela individuella deltagardata (IPD)?
IPD-planbeskrivning
No Currently, there is no plan to share data with other researchers.
Läkemedels- och apparatinformation, studiedokument
Studerar en amerikansk FDA-reglerad läkemedelsprodukt
Studerar en amerikansk FDA-reglerad produktprodukt
produkt tillverkad i och exporterad från U.S.A.
Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .