Goldmann Applanation Tonometry (GAT) is considered the clinical gold standard for eye pressure measurements and yet it is known to be a subjective measurement with limited repeatability and limited portability. Another clinical standard for checking eye pressure is known as the pneumotonometer. This method is more objective but not portable. The purpose of this study is to develop new methods of measuring eye pressure that are more objective, reproducible and portable. In this study, the investigators will be comparing the eye pressure measurements using 2 investigational methods to GAT and pneuumotonometer.

Обзор исследования

Подробное описание

This is a prospective study. Patients presenting for their scheduled eye appointments will be recruited by verbal communication. Following informed consent, intraocular pressure (IOP) measurements will be made by the following methods:

Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.

Pneumotonometer: This is what is done in routine eye care. The participant will receive a topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure.

Fixed force GAT: From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a complementary metal oxide semiconductor (C-MOS) camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. A similar method using an iPod touch camera has been previously published by this PI.

Upright applanating prototype: With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.

The order of the above 4 measurements will be randomized.

Тип исследования

Интервенционный

Регистрация (Оцененный)

200

Фаза

  • Непригодный

Контакты и местонахождение

В этом разделе приведены контактные данные лиц, проводящих исследование, и информация о том, где проводится это исследование.

Контакты исследования

  • Имя: Sarah Jones
  • Номер телефона: 919 681 6584
  • Электронная почта: sarah.jones1@duke.edu

Места учебы

    • North Carolina
      • Durham, North Carolina, Соединенные Штаты, 27710
        • Рекрутинг
        • Duke Eye Center
        • Главный следователь:
          • Joanne Wen, MD
        • Контакт:
          • Joanne Wen, MD
          • Номер телефона: +1 919 684 2975
          • Электронная почта: joanne.wen@duke.edu

Критерии участия

Исследователи ищут людей, которые соответствуют определенному описанию, называемому критериям приемлемости. Некоторыми примерами этих критериев являются общее состояние здоровья человека или предшествующее лечение.

Критерии приемлемости

Возраст, подходящий для обучения

18 лет и старше (Взрослый, Пожилой взрослый)

Принимает здоровых добровольцев

Нет

Полы, имеющие право на обучение

Все

Описание

Criteria:

Inclusion Criteria:

  • Presenting for a routine eye exam
  • ≥ 18 years of age
  • Able and willing to give consent

Exclusion Criteria:

  • History of corneal scarring
  • Active infection of the eye
  • Ocular surface trauma or infection

Учебный план

В этом разделе представлена ​​подробная информация о плане исследования, в том числе о том, как планируется исследование и что оно измеряет.

Как устроено исследование?

Детали дизайна

  • Основная цель: Диагностика
  • Распределение: Рандомизированный
  • Интервенционная модель: Параллельное назначение
  • Маскировка: Нет (открытая этикетка)

Оружие и интервенции

Группа участников / Армия
Вмешательство/лечение
Активный компаратор: Standard GAT

This is the gold-standard method for IOP measurement in clinical practice

Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice

Экспериментальный: Fixed-force GAT

From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter

An investigational device similar to standard GAT

Экспериментальный: Upright applanating prototype

With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter

An investigational device, prism used in standard and fixed-force GAT is attached to a portable device

Активный компаратор: Pneumotonometer

This is another standard method for IOP measurement in clinical practice

This is a standard for checking eye pressure. A topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure

Что измеряет исследование?

Первичные показатели результатов

Мера результата
Мера Описание
Временное ограничение
Внутриглазное давление, измеренное в положении лежа пневмотонометром
Временное ограничение: Исходный уровень
Исходный уровень
Intraocular pressure (IOP) as measured by standard GAT
Временное ограничение: Baseline

this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.

Baseline
Intraocular pressure (IOP) as measured by fixed force GAT
Временное ограничение: Baseline

From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular.

Baseline
Intraocular pressure as measured by upright applanating
Временное ограничение: Baseline

With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.

Baseline

Соавторы и исследователи

Здесь вы найдете людей и организации, участвующие в этом исследовании.

Спонсор

Следователи

  • Главный следователь: Joanne Wen, MD, Duke Eye Center

Даты записи исследования

Эти даты отслеживают ход отправки отчетов об исследованиях и сводных результатов на сайт ClinicalTrials.gov. Записи исследований и сообщаемые результаты проверяются Национальной медицинской библиотекой (NLM), чтобы убедиться, что они соответствуют определенным стандартам контроля качества, прежде чем публиковать их на общедоступном веб-сайте.

Изучение основных дат

Начало исследования (Действительный)

2 марта 2026 г.

Первичное завершение (Оцененный)

29 декабря 2026 г.

Завершение исследования (Оцененный)

29 декабря 2026 г.

Даты регистрации исследования

Первый отправленный

19 декабря 2025 г.

Впервые представлено, что соответствует критериям контроля качества

19 декабря 2025 г.

Первый опубликованный (Действительный)

23 декабря 2025 г.

Обновления учебных записей

Последнее опубликованное обновление (Действительный)

3 ноября 2026 г.

Последнее отправленное обновление, отвечающее критериям контроля качества

2 ноября 2026 г.

Последняя проверка

1 ноября 2026 г.

Дополнительная информация

Термины, связанные с этим исследованием

Дополнительные соответствующие термины MeSH

Другие идентификационные номера исследования

  • Pro00106111_1

Планирование данных отдельных участников (IPD)

Планируете делиться данными об отдельных участниках (IPD)?

НЕТ

Описание плана IPD

No Currently, there is no plan to share data with other researchers.

Информация о лекарствах и устройствах, исследовательские документы

Изучает лекарственный продукт, регулируемый FDA США.

Нет

Изучает продукт устройства, регулируемый Управлением по санитарному надзору за качеством пищевых продуктов и медикаментов США.

Да

продукт, произведенный в США и экспортированный из США.

Нет

Эта информация была получена непосредственно с веб-сайта clinicaltrials.gov без каких-либо изменений. Если у вас есть запросы на изменение, удаление или обновление сведений об исследовании, обращайтесь по адресу register@clinicaltrials.gov. Как только изменение будет реализовано на clinicaltrials.gov, оно будет автоматически обновлено и на нашем веб-сайте. .

Подписаться