이 페이지는 자동 번역되었으며 번역의 정확성을 보장하지 않습니다. 참조하십시오 영문판 원본 텍스트의 경우.

Validation of a Miniature SpO2 Sensor in Healthy Subjects and Patients With Long-term Oxygen Therapy. (OxyFrame)

2016년 10월 26일 업데이트: University Hospital Inselspital, Berne

Validation of a Miniature SpO2 Sensor in Healthy Subjects and Patients With Long-term Oxygen Therapy. A Prospective, Method Comparison, Proof of Concept Study.

By integrating a miniaturized pulse oximetry sensor into the frame of oxygen delivery glasses which dissimulates the nasal cannula, the investigators hope to optimize and long term oxygen therapy (LTOT) regarding medical and social aspects. The validation of the novel SpO2 sensor is the first step in the concept of personalized, dynamic delivery of oxygen by eyeglasses using a closed-loop system.

연구 개요

상세 설명

Objectives:

  1. To determine accuracy of SpO2 measurements by the novel sensor compared to 2 reference pulse oximeters in healthy subjects undergoing a normobaric hypoxic challenge provoking SpO2 desaturation.
  2. To determine accuracy of SpO2 measurements by the novel sensor compared to 2 reference pulse oximeters and to arterial blood gas analysis in patients qualifying for LTOT, breathing room air at rest (lying, sitting, standing position) and during exercise (slow walking, 6-MWT, ergometer cycle).

연구 유형

중재적

등록 (실제)

36

단계

  • 해당 없음

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 장소

      • Bern, 스위스, 3010
        • Department of Pulmonary Medicine, University Hospital Bern

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

18년 이상 (성인, 고령자)

건강한 자원 봉사자를 받아들입니다

연구 대상 성별

모두

설명

Healthy volunteers

Inclusion Criteria:

  • Age 18 - 60 years,
  • Non smoker,
  • Good general health and physical fitness
  • Informed Consent

Exclusion Criteria:

  • Current or history of tobacco smoking (> 5 py)
  • Current alcohol or drug abuse
  • Heart disease or untreated arterial hypertension
  • Pulmonary disease
  • Epilepsy or other major neurologic disease
  • Dark skin colour
  • Known anemia
  • Severe metabolic disease (e.g. diabetes)
  • Pregnancy

Patients

Inclusion Criteria:

  • Age ≥ 18 years
  • Long term oxygen therapy at rest and/or on physical effort
  • Hypoxemia in arterial blood gas analysis at rest PaO2 ≤ 55 mmHg or < 60 mmHg with PHT or polycythemia (men Hb > 174 g/L, women Hb > 140 g/L) or:

Oxygen desaturation on effort (SpO2 < 90%)

  • Pulmonary disease including Chronic Obstructive Pulmonary Disease, Interstitial Lung Disease and Sarcoidosis
  • Informed Consent

Exclusion Criteria:

  • Severe hypoxemia (pO2 <40 mmHg)
  • Severe hypercapnia (pCO2 >55 mmHg)
  • Acute respiratory infection or systemic infection,
  • Severe systolic heart disease (LVEF < 25%)
  • Myocardial infarction during the last month before inclusion
  • Unstable angina pectoris
  • Severe aortic stenosis (mean gradient > 40mmHg, valve area <1 cm2)
  • Rhythmogenic heart disease
  • Severe or untreated arterial hypertension (blood pressure > 180mmHg systolic, >100mmHg diastolic)
  • Anemia (Hb < 120 g/l)
  • Known or suspected coagulation disorder (e.g. INR > 4)
  • Severe or untreated metabolic disorder
  • Neurologic disease or dementia
  • Pregnancy
  • History of non-compliance to medical treatment
  • Current alcohol, drug abuse or current tobacco smoking
  • Dark skin color

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

  • 할당: 해당 없음
  • 중재 모델: 단일 그룹 할당
  • 마스킹: 없음(오픈 라벨)

무기와 개입

참가자 그룹 / 팔
개입 / 치료
다른: Healthy Subjects and patients
Validation of SpO2 sensor in healthy subjects as a first step. Validation of SpO2 sensor in patients as a second step.
Validation of a miniaturized pulse oximetry sensor for integration into the frame of oxygen delivery glasses.

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Assessment of accuracy in patients
기간: At baseline
Assessment of the accuracy of the novel pulse oximeter sensor in patients qualifying for LTOT over the entire saturation range by calculating accuracy root mean square (ARMS). Continuous SpO2 measurement of the novel pulse oximeter will be compared to measurements assessed by 2 standard pulse oximeters and arterial blood gas analysis.
At baseline

2차 결과 측정

결과 측정
측정값 설명
기간
Assessment of accuracy in healthy volunteers
기간: At baseline
Assessment of accuracy of the novel pulse oximeter sensor in healthy subjects undergoing a normobaric hypoxic challenge provoking SpO2 desaturation. SpO2 will be measured continuously over the hole procedure and SpO2 range by the novel pulse oximeter sensor and the 2 reference pulse oximeters.
At baseline

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

수사관

  • 수석 연구원: Manuela Funke, MD, Department of Pulmonary Medicine, University Hospital Bern

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작

2016년 3월 1일

기본 완료 (실제)

2016년 4월 1일

연구 완료 (실제)

2016년 5월 1일

연구 등록 날짜

최초 제출

2016년 3월 23일

QC 기준을 충족하는 최초 제출

2016년 3월 23일

처음 게시됨 (추정)

2016년 3월 30일

연구 기록 업데이트

마지막 업데이트 게시됨 (추정)

2016년 10월 27일

QC 기준을 충족하는 마지막 업데이트 제출

2016년 10월 26일

마지막으로 확인됨

2016년 10월 1일

추가 정보

이 연구와 관련된 용어

추가 관련 MeSH 약관

기타 연구 ID 번호

  • OxyFrame

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

3
구독하다