접촉 없이 심장, 호흡 및 움직임을 모니터링하는 도지(Dozee)의 검증
Dozee 시스템을 이용한 무접촉 심박수, 호흡수, 움직임 및 침대 이탈 모니터링의 임상 검증
The Dozee VS system (K221555) is a contactless device that tracks heart rate and breathing rate using a sensor sheet placed under the mattress. It captures body vibrations and converts them into vital signs, which can be monitored remotely in real time. The system also supports add-ons like a blood pressure cuff, pulse oximeter, and thermometer, offering a complete monitoring solution. This study aims to validate the updated AI-powered heart rate and improved breathing rate algorithms across different sensor sheet types to confirm clinical accuracy
연구 개요
상세 설명
The Dozee VS system (K221555), developed by Turtle Shell Technologies, is a non-invasive, contactless device that continuously monitors heart rate (HR) and breathing rate (RR). A thin sensor sheet placed under the mattress detects tiny body vibrations from the heart and lungs, which are processed by smart algorithms and displayed on a secure online dashboard for real-time monitoring.
Unlike traditional methods such as ECGs, nasal tubes, or thermistors-which are invasive, require skilled placement, and are uncomfortable for long-term use-Dozee is simple, comfortable, and doesn't rely on patient compliance like wearables do. The system can also connect to add-on devices, including a blood pressure cuff, pulse oximeter, and thermometer, giving a full picture of a patient's vital signs and helping spot early signs of deterioration.
Hardware upgrades now include dual-sensor sheets that improve accuracy across different mattresses and patient types. The new HR algorithm uses AI/ML for better accuracy, while the improved RR algorithm performs more reliably even with movement or weak signals.
Study Goal: This study aims to validate the updated algorithms (Gen 2) for HR and RR across the three sensor sheet types (DS, S1, and S1-I) to ensure they meet clinical accuracy standards and to benchmark against the previous versions of the algorithms
연구 유형
등록 (실제)
연락처 및 위치
연구 장소
-
-
Texas
-
Midland, Texas, 미국, 79707
- ERA Health Research
-
-
참여기준
자격 기준
공부할 수 있는 나이
건강한 자원 봉사자를 받아들입니다
연구 대상 성별
샘플링 방법
연구 인구
The study will be conducted on atleast 60 subjects in the age group above 18 years regardless of gender. The patients/volunteers be informed about the study and a written informed consent form will be obtained from them. Dozee will be placed under the mattress of the patient's bed and the data will be collected through the device whenever a patient is on the bed.
설명
Inclusion Criteria:
Subjects will be included if they meet the following inclusion criteria:
- Adults Male or female. (Above 18 years of age)
- Provide written informed consent.
- Weight between 40kg/88.18lbs and 120kg/264.55lbs
- Agree to not eat during the testing period
- Able to move and exit the bed independently and safely.
Exclusion Criteria:
Subjects will be excluded if they have any of the following inclusion criteria:
- Are connected to a device that may interfere with the device monitoring in this study.
- Are receiving any bedside care which may be incompatible with the study procedures.
- A likely need to receive or undergo a procedure during the testing period.
- Cannot accept a nasal cannula, or have an ECG leads placed on the chest.
- Have a significant medical condition in the judgement of the investigator, which may compromise the study testing procedures.
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
- 관찰 모델: COHORT
- 시간 관점: PROSPECTIVE
코호트 및 개입
그룹/코호트 |
개입 / 치료 |
|---|---|
|
Single Arm Study (Cohort)
Group/Cohort Description: All participants will be on standard telemetry monitoring for Heart Rate and Respiration Rate using the gold standard (Electrocardiography and Capnography respectively). In addition, a contactless monitoring devices will be placed on the patient bed to measure data. This device is Dozee VS (Investigational Device). Data from all devices will be measured simultaneously. |
All participants will be on standard telemetry monitoring for Heart Rate, Respiration Rate and movements using the gold standard (Electrocardiography and Capnography respectively). |
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Mean Absolute Error (MAE) of Dozee's Heart Rate a from the Gold Standard ECG
기간: 6 months
|
Heart Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Heart Rate will be calculated. |
6 months
|
|
Accuracy (%) of Dozee in identfying subject movement
기간: 6 months
|
Time stamps of subject movement during the study will be recorded manullay. Time stamps of Dozee's movement will be recorded. Accuracy of the Dozee device in detecting movement compared to manual observation will be computed in terms of percentage. |
6 months
|
|
Accuracy(%) of of Dozee's bed-exits
기간: 6 months
|
Time stamps of bed-exits will be recorded manually during the study will be recorded. Time stamps of Dozee's bed-exit will be computed. Accuracy of the Dozee device in detecting bed-exit compared to manual observation as gold standard will be computed in terms of percentage. |
6 months
|
|
Mean Absolute Error (MAE) of Dozee's Respiratory Rate from the Gold Standard Capnography
기간: 6 Months
|
Respiratory Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Respiratory Rate will be calculated. |
6 Months
|
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Range of values (minimum, maximum, and spread) of Heart Rate in beats per min measured by the gold standard and the Dozee
기간: 6 months
|
The outcome will be assessed by comparing the range of values (minimum, maximum, and spread) of Heart Rate ( beats per minute) measured by the gold standard and the Dozee over the study period. |
6 months
|
|
Range of values (minimum, maximum, and spread) of Respiratory Rate in breaths per min measured by the gold standard and the Dozee
기간: 6 Months
|
The outcome will be assessed by comparing the range of values (minimum, maximum, and spread) of Respiratory Rate ( breaths per minute) measured by the gold standard and the Dozee over the study period. |
6 Months
|
|
Bland- Altman Level of Agreement between Dozee Heart Rate and Respiratory Rate with the Gold Standard
기간: 6 months
|
Level of Agreement between Dozee's HR and RR with reference device computed by Bland Altman. |
6 months
|
|
Detection rate (%) of the Dozee's Heart Rate and Respiratory Rate
기간: 6 months
|
The time-stamps of Dozee's Heart Rate and Respiratory Rate and the gold standard Heart Rate and Respiratory Rate will be recorded. DR is the proportion of gold standard points for which the DS successfully provides a corresponding measured value, relative to the total number of gold standard points. |
6 months
|
|
Level of Agreement between Dozee Heart Rate and Respiratory Rate with the Gold Standard
기간: 6 months
|
Level of Agreement between Dozee Heart Rate and Respiratory Rate with the Gold Standard by using Deming Regression with Pearson Co-efficient |
6 months
|
공동 작업자 및 조사자
수사관
- 수석 연구원: Sapna Chilka, MD, ERA Health Research
연구 기록 날짜
연구 주요 날짜
연구 시작 (실제)
기본 완료 (실제)
연구 완료 (실제)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
기타 연구 ID 번호
- DZ_VS_VAL
개별 참가자 데이터(IPD) 계획
개별 참가자 데이터(IPD)를 공유할 계획입니까?
IPD 계획 설명
Individual Participant Data (IPD) are protected under HIPPA Act of 1996.
약물 및 장치 정보, 연구 문서
미국 FDA 규제 의약품 연구
미국 FDA 규제 기기 제품 연구
미국에서 제조되어 미국에서 수출되는 제품
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .