Product Performance of a Daily Disposable Contact Lens
A Study to Assess the Product Performance of a Daily Disposable Contact Lens
Studie Overzicht
Toestand
Toestand
Conditie
Conditie
Interventie / Behandeling
Interventie / Behandeling
Studietype
Studietype
Inschrijving (Werkelijk)
Inschrijving
Fase
Fase
- Niet toepasbaar
Contacten en locaties
Studie Locaties
-
-
-
Singapore, Singapore, 556741
- Bausch & Lomb
-
-
Deelname Criteria
Geschiktheidscriteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Beschrijving
Inclusion Criteria:
- Subjects must be females of Chinese descent and of legal age and no older than 38. OR subjects that are females of Chinese descent and are at least 15 years of age and less than the legal age must: have a parent/legal guardian provide informed consent.
- Subjects must have clear central corneas and be free of any anterior segment disorders.
- Subjects must be adapted current wearers of marketed opaque tinted soft contact lenses, wear their lenses two days or more per week, and wear a lens in each eye.
Exclusion Criteria:
- Subjects who have worn gas permeable (GP) contact lenses within the last 30 days or have worn polymethylmethacrylate (PMMA) lenses within the last three months.
- Subjects who have any systemic disease affecting ocular health.
- Subjects using any systemic or topical medications that will, in the Investigator's opinion, affect ocular physiology or lens performance.
- Subjects with an active ocular disease or are using any ocular medication.
- Subjects with any scar or neovascularization within the central 4 mm of the cornea.
- Subjects who are aphakic or amblyopic.
- Subjects who have had any corneal surgery (eg, refractive surgery).
- Subjects who currently wear monovision, multifocal, or toric contact lenses.
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Ander
- Toewijzing: NVT
- Interventioneel model: Opdracht voor een enkele groep
- Masker: Geen (open label)
Aantal wapens
Wapens en interventies
Deelnemersgroep / ArmDeelnemersgroep / Arm |
Interventie / BehandelingInterventie / Behandeling |
|---|---|
|
Experimenteel: Prism I
Bausch & Lomb daily disposable cosmetic tint contact lens
|
Bausch & Lomb daily disposable cosmetic tint contact lens worn on a daily disposable basis for 1 week.
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Market Research Survey
Tijdsspanne: 1 week
|
Online subjective assessments that subjects responded to after wearing the study lenses for at least one week.
Subjects were asked to rate their overall opinion as: Excellent, Very Good, Good, Fair, or Poor.
|
1 week
|
Secundaire uitkomstmaten
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Lens Evaluation - Centration
Tijdsspanne: Over all study visits through 1 week
|
To determine proper lens parameters the lens relationship to the eye was observed using a slit lamp.
Lens centration, meaning the lens should provide full corneal coverage, was assessed during the slit lamp evaluation.
Findings were reported as: Excellent, Good, Fair, or Poor.
|
Over all study visits through 1 week
|
|
Lens Evaluation - Movement
Tijdsspanne: Over all study visits through 1 week
|
To determine proper lens parameters the lens relationship to the eye was observed using a slit lamp. Lens movement, meaning the lens should provide discernible movement with:
|
Over all study visits through 1 week
|
|
Visual Acuity - Distance High Contrast logMAR (Logarithm of the Minimum Angle of Resolution) Lens Visual Acuity (VA).
Tijdsspanne: Over all study visits for 1 week
|
For determination of high contrast VA, the subject was to be seated so that the distance from the subject's eyes to the logMAR chart is 6.5 feet (2.0 meters).
The chart should be at eye level for the subject.
The logMAR charts have two alternative letter sequences from 28 letters (0.3 logMAR) to 62 letters (-0.3 logMAR).
The visual acuity should be measured through the phoropter using the distance refractive correction with the addition of +0.50D to compensate for the reduced test distance of 6.5 feet (2.0 meters).
|
Over all study visits for 1 week
|
Medewerkers en onderzoekers
Sponsor
Sponsor
Studie record data
Bestudeer belangrijke data
Studie start
Studie start
Primaire voltooiing (Werkelijk)
Primaire voltooiing
Studie voltooiing (Werkelijk)
Studie voltooiing
Studieregistratiedata
Eerst ingediend
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Schatting)
Eerst geplaatst
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update geplaatst
Laatste update ingediend die voldeed aan QC-criteria
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Trefwoorden
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
Andere studie-ID-nummers
- 654
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .