Pilot Clinical Investigation of Adaptative Eyeglasses for the Correction of Presbyopia (eve-1)
6 april 2022 bijgewerkt door: Laclaree
Pilot Clinical Investigation of Adaptive Eyeglasses for the Correction of Presbyopia
This study is a pilot clinical investigation of adaptive eyeglasses for the correction of presbyopia.
The goal of this clinical investigation is to assess IMD safety and to obtain patient feedback on the usefulness of the Laclarée eyeglasses, through subjective clinical measures and qualitative evaluations.
This is an exploratory investigation.
Studie Overzicht
Toestand
Voltooid
Conditie
Interventie / Behandeling
Gedetailleerde beschrijving
Presbyopia is a visual impairment caused by the normal aging of the eye lens and resulting in progressive loss of accommodation.
As a consequence, from the onset of presbyopia occurring around 40-45 years old, the eye has growing trouble focusing at near and intermediate distances.
Millions of presbyopes encounter difficulties with current corrective solutions, leading to vision and postural discomfort.
Laclarée's adaptive eyeglasses are providing an alternative solution, closer to the native crystalline accommodation.
Besides an enhanced field-of-view, these eyeglasses offer a disruptive ergonomics where the glasses automatically adjust to the wearer's need to provide clear vision at all distances and can be electronically recalibrated to adjust the delivered addition throughout presbyopia.
Studietype
Ingrijpend
Inschrijving (Werkelijk)
40
Fase
- Niet toepasbaar
Contacten en locaties
In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.
Studie Locaties
-
-
-
Saint-Étienne, Frankrijk, 42055
- CHU Saint-Etienne Hôpital Nord, Service d'Ophtalmologie 25, boulevard Pasteur
-
-
Deelname Criteria
Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
- Kind
- Volwassen
- Oudere volwassene
Accepteert gezonde vrijwilligers
Nee
Geslachten die in aanmerking komen voor studie
Allemaal
Beschrijving
Inclusion Criteria:
- Patient >18 years of age, male or female who has given explicit informed consent (dated and signed)
- Diagnosed with presbyopia
- A wearer whose current corrective solution is a pair of progressive lenses that provide clear vision at both distance and near vision (worn at least 5 times per week for 4 hours per day), or, a wearer who has rejected progressive glasses that provide clear vision at both distance and near vision and who uses another corrective solution that provides clear vision at both distance and near vision (bifocal glasses, multifocal or monovision lenses, multiple pairs of glasses or one pair for cases of emmetropia)
- Patient with a distance visual acuity of 10/10 at least with current corrective solution
- Patient with a visual acuity of P2 at least in near vision with current corrective solution
- Patient with at least 10/10 distance vision once corrected with contact lenses for the test
- Inter-pupillary distance in distance vision between 55 and 71mm
- Distance from temple-to-temple not exceeding 161mm
- Addition greater than or equal to 1.75D: Add≥1.75D
- Sphere strictly between -6D and +4D: -6D<S<+4D
- Cylinder below 1.75D:C<1.75D
- Patient uses a computer (at least 3 times a week in private and/or work settings).
- Affiliation to a social security scheme
- French-speaking (fluent).
- Patient willing to participate in the study.
Exclusion Criteria:
- Patient under tutorship or guardianship.
- Patient unable to understand the study procedures and therefore to provide free and informed consent
- Patient with reading difficulties
- Patient with multifocal implants
- Patient who has undergone multifocal surgery
- Patient with prism correction
- Patient with a current or past eye condition that may adversely affect vision
- Patient being treated or having undergone surgery likely to adversely affect vision
- Patient with mobility problems that would prevent tests and workshops from being carried out
- Patient intolerant to wearing contact lenses for a few hours (only for presbyopes with ametropia)
- Patient who is pregnant or breastfeeding
- Patient already included once in the study
- Patient placed in an institution based on a judicial or regulatory ruling, held in psychiatric department, in state-run prison, or employed by the study research sites or by the sponsor.
Studie plan
Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Ondersteunende zorg
- Toewijzing: Gerandomiseerd
- Interventioneel model: Crossover-opdracht
- Masker: Geen (open label)
Wapens en interventies
Deelnemersgroep / Arm |
Interventie / Behandeling |
---|---|
Ander: First Arm
|
The tested investigational medical device is a prototype of semi-integrated eyeglasses.
The comparator is either a pair of progressive lenses that provide clear vision at both distance and near vision, either, for wearers who have rejected progressive glasses, an alternative solution providing clear vision at both distance and near vision which can consist of bifocal glasses, multifocal or monovision lenses, multiple pairs of glasses or one pair for cases of emmetropia.
|
Ander: Second Arm
|
The tested investigational medical device is a prototype of semi-integrated eyeglasses.
The comparator is either a pair of progressive lenses that provide clear vision at both distance and near vision, either, for wearers who have rejected progressive glasses, an alternative solution providing clear vision at both distance and near vision which can consist of bifocal glasses, multifocal or monovision lenses, multiple pairs of glasses or one pair for cases of emmetropia.
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
---|---|---|
IMD safety evaluation
Tijdsspanne: 2 hours
|
To evaluate the IMD safety : Collection of adverse events.
|
2 hours
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
---|---|---|
Exploratory evaluation of the IMD - controlled tests assessing visual acuity and performance on visual tasks
Tijdsspanne: 2 hours
|
To make an exploratory evaluation of the IMD performance on patient vision quality, when compared to that obtained with the patient usual personal eyewear (comparator) : Subjective measures on controlled tests assessing visual acuity and performance on visual tasks.
|
2 hours
|
Exploratory evaluation of the IMD - a questionnaire assessing vision quality
Tijdsspanne: 2 hours
|
To make an exploratory evaluation of the IMD performance on patient vision quality, when compared to that obtained with the patient usual personal eyewear (comparator) : Subjective measures on questionnaires.
|
2 hours
|
Exploratory evaluation of the IMD - patient qualitative feedback
Tijdsspanne: 2 hours
|
To make an exploratory evaluation of the IMD performance on patient vision quality, when compared to that obtained with the patient usual personal eyewear (comparator) : Collection of patient feedback.
|
2 hours
|
Medewerkers en onderzoekers
Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.
Sponsor
Medewerkers
Onderzoekers
- Studie stoel: Bruno BERGE, PhD, SPONSOR: LACLAREE
- Studie directeur: Jessica JAROSZ, PhD, SPONSOR: LACLAREE
- Hoofdonderzoeker: Philippe GAIN, Prof., CHU Saint-Etienne
Publicaties en nuttige links
De persoon die verantwoordelijk is voor het invoeren van informatie over het onderzoek stelt deze publicaties vrijwillig ter beschikking. Dit kan gaan over alles wat met het onderzoek te maken heeft.
Nuttige links
- Jessica Jarosz, Norbert Molliex, Guilhem Chenon, and Bruno Berge, "Adaptive eyeglasses for presbyopia correction: an original variable-focus technology," Opt. Express 27, 10533-10552 (2019)
- Jessica Jarosz, Norbert Molliex, Quentin Lavigne, and Bruno Berge "An original low-power opto-fluidic engine for presbyopia-correcting adaptive eyeglasses", Proc. SPIE 10858, Ophthalmic Technologies XXIX, 1085824 (28 February 2019)
Studie record data
Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.
Bestudeer belangrijke data
Studie start (Werkelijk)
9 september 2021
Primaire voltooiing (Werkelijk)
27 januari 2022
Studie voltooiing (Werkelijk)
27 januari 2022
Studieregistratiedata
Eerst ingediend
1 maart 2022
Eerst ingediend dat voldeed aan de QC-criteria
6 april 2022
Eerst geplaatst (Werkelijk)
13 april 2022
Updates van studierecords
Laatste update geplaatst (Werkelijk)
13 april 2022
Laatste update ingediend die voldeed aan QC-criteria
6 april 2022
Laatst geverifieerd
1 april 2022
Meer informatie
Termen gerelateerd aan deze studie
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- 2021-A00521-40
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
NEE
Beschrijving IPD-plan
Not planned
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Nee
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
Nee
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .