Validation of a Wearable Bioimpedance Sensor for Hydration Monitoring (VAMOS)
Validation of a Wearable Bioimpedance Sensor for Non-Invasive Monitoring of Hydration Status
The purpose of the project is to validate the performance of a wearable bioimpedance sensor (Re:Balans®) for monitoring hydration status. The study includes hospitalized patients aged 65 years or older who are admitted to the hospital with suspected dehydration. Data will be collected continuously throughout the hospital stay using one sensor placed on the back, one on the upper arm, and one on the front of the thigh. Sensor measurements will be compared with multiple reference methods for assessing hydration status, including blood tests, body weight and fluid balance charts. Agreement with these reference measures will form the basis for the performance evaluation.
The anticipated benefit of the project is the validation of a novel, user-friendly, and non-invasive tool for hydration monitoring, with the aim of improving patient care, preventing complications, and reducing the need for hospital readmissions.
Przegląd badań
Status
Warunki
Interwencja / Leczenie
Szczegółowy opis
The study aims to evaluate how well a wearable bioimpedance sensor (Re:Balans®) tracks changes in hydration status in hospitalized patients. The objective is to validate the sensor's ability to provide continuous, non-invasive monitoring of fluid balance, which may improve the management and treatment of patients at risk of dehydration or fluid overload. Such a solution could also help reduce the need for hospital admissions and in-person follow-up visits.
The investigation is designed as a prospective cohort study in which patients are followed throughout their hospital stay, typically lasting 3-7 days. Participants will wear three sensors, placed on the upper back, upper arm, and anterior thigh. The sensors automatically perform measurements every minute during the study period.
At baseline and during daily follow-up visits, several assessments of hydration status will be performed. These include body weight measurements, blood tests, and whole-body bioimpedance measurements. In a subgroup of patients, ultrasound assessment of the inferior vena cava will also be performed. The combined results from these assessments will serve as reference measures to determine whether the patient's hydration status has changed over time.
At hospital discharge, the assessments are repeated, the sensors are removed, and participants complete a short questionnaire regarding their experience with the device. In addition, fluid balance data (fluid intake and output) will be collected retrospectively from medical records to allow comparison between the sensor measurements and routine clinical monitoring methods. Relevant clinical information obtained as part of standard patient care will also be recorded if considered important for the assessment of hydration status.
Typ studiów
Zapisy (Szacowany)
Kontakty i lokalizacje
Kontakt w sprawie studiów
- Nazwa: Sigve N Aas, PhD
- Numer telefonu: 004741499074
- E-mail: sigve.aas@modesensors.com
Kopia zapasowa kontaktu do badania
- Nazwa: Frida Bremnes, MSc
- E-mail: frida.bremnes@modesensors.com
Lokalizacje studiów
-
-
Trøndelag
-
Trondheim, Trøndelag, Norwegia, 7030
- St. Olavs Hospital
-
Kontakt:
- Lars P Bjørnsen, PhD
- Numer telefonu: 004793212617
- E-mail: lars.petter.bache-wiig.bjornsen@stolav.no
-
Główny śledczy:
- Lars P Bjørnsen, PhD
-
-
Kryteria uczestnictwa
Kryteria kwalifikacji
Wiek uprawniający do nauki
- Starszy dorosły
Akceptuje zdrowych ochotników
Metoda próbkowania
Badana populacja
Opis
Inclusion Criteria:
- Age ≥ 65 years
- Planned fluid management due to presumed dehydration
- Expected duration of hospital stay ≥ 3 days
Exclusion Criteria:
- Known allergies or skin sensitivities to electrode hydrogel and/or acrylic adhesives
- Implantable pulse generators such as pacemakers and defibrillators, and/or use of other electrical medical equipment for which an interaction effect with the investigational device cannot be ruled out.
- Scheduled MRI
- Scheduled diathermy
- Breached skin at device application areas (upper back, anterior thigh and upper arm)
- Unconsciousness
- Critically ill (presence of acute organ dysfunction requiring intensive monitoring and management)
- Any medical or psychiatric condition, which in the opinion of the investigator precludes participation
- Incapable of giving informed consent
Plan studiów
Jak projektuje się badanie?
Szczegóły projektu
Kohorty i interwencje
Grupa / Kohorta |
Interwencja / Leczenie |
|---|---|
|
Dehydrated patients
The cohort includes individuals 65 years or older admitted to the hospital with dehydration
|
Subjects will wear three bioimpedance sensors throughout their stay at the hospital
|
Co mierzy badanie?
Podstawowe miary wyniku
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Relative change in device output from admission to discharge
Ramy czasowe: Day 1 and day of discharge (up to 1 week)
|
Relative change in device output from admission to discharge in patients with a confirmed increase in fluid volume, compared with subjects whose hydration status remain stable
|
Day 1 and day of discharge (up to 1 week)
|
Miary wyników drugorzędnych
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Sensitivity/specificity of the device for the detection of significant fluid volume changes
Ramy czasowe: Once daily during hospital stay, assessed up to 7 days
|
Proportion of true positives (changed hydration) and true negatives (unchanged hydration) correctly identified by the device
|
Once daily during hospital stay, assessed up to 7 days
|
|
Sensitivity/specificity at the three placements
Ramy czasowe: Once daily during hospital stay, assessed up to 7 days
|
Proportion of true positives (changed hydration) and true negatives (unchanged hydration) correctly identified by the device
|
Once daily during hospital stay, assessed up to 7 days
|
|
Sensitivity/specificity of different device outputs
Ramy czasowe: Once daily during hospital stay, assessed up to 7 days
|
Proportion of true positives (changed hydration) and true negatives (unchanged hydration) correctly identified by the device
|
Once daily during hospital stay, assessed up to 7 days
|
|
Correlation between device output and reference measurements
Ramy czasowe: Day 1 and day of discharge (up to 1 week)
|
Correlation between changes in device output and reference measurements during the hospital stay
|
Day 1 and day of discharge (up to 1 week)
|
|
User satisfaction among patients
Ramy czasowe: Day 3
|
Subjects will respond to a questionnaire related to user comfort.
The questionnaire uses a 5-point Likert scale ranging from 1 to 5, where lower scores indicate better outcomes and higher scores indicate worse outcomes.
The questionnaire will be filled out on day 3 of the hospital stay.
|
Day 3
|
Współpracownicy i badacze
Sponsor
Współpracownicy
Śledczy
- Dyrektor Studium: Sigve N Aas, PhD, Mode Sensors AS
Daty zapisu na studia
Główne daty studiów
Rozpoczęcie studiów (Szacowany)
Zakończenie podstawowe (Szacowany)
Ukończenie studiów (Szacowany)
Daty rejestracji na studia
Pierwszy przesłany
Pierwszy przesłany, który spełnia kryteria kontroli jakości
Pierwszy wysłany (Rzeczywisty)
Aktualizacje rekordów badań
Ostatnia wysłana aktualizacja (Rzeczywisty)
Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości
Ostatnia weryfikacja
Więcej informacji
Terminy związane z tym badaniem
Dodatkowe istotne warunki MeSH
Inne numery identyfikacyjne badania
- CTR0062A VAMOS
Plan dla danych uczestnika indywidualnego (IPD)
Planujesz udostępniać dane poszczególnych uczestników (IPD)?
Informacje o lekach i urządzeniach, dokumenty badawcze
Bada produkt leczniczy regulowany przez amerykańską FDA
Bada produkt urządzenia regulowany przez amerykańską FDA
produkt wyprodukowany i wyeksportowany z USA
Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .
Badania kliniczne na Bioimpedance sensor
-
Imperial College LondonCW+ CharityZakończonyZastosowanie do zdalnego monitorowania w przypadku podejrzenia Covid-19 (koronawirus) (REMOTE-COVID)KoronawirusZjednoczone Królestwo
-
Léman Micro Devices SANieznanyDokładność kliniczna czujnika ciśnienia funkcji życiowych
-
RWTH Aachen UniversityPhilips Technology Aachen GermanyNieznanyOstra zdekompensowana niewydolność serca (ADHF) | Ostry kardiogenny obrzęk płuc (ACPE)Niemcy
-
National Taiwan University HospitalMinistry of Health and Welfare, TaiwanZakończonyRak jelita grubego | Choroba żołądkaTajwan
-
Universitaire Ziekenhuizen KU LeuvenBytefliesZakończony
-
Prim. Prof. Dr. Oliver Findl, MBARekrutacyjny
-
University Hospital, GhentUniversity Ghent; BytefliesRekrutacyjnyPostępujące stwardnienie rozsianeBelgia
-
Busra Kalkan BalakZakończonyZdrowi Wolontariusze | Sprawność fizyczna (G11.427.410.698)Turcja (Türkiye)
-
Abbott Medical DevicesZakończonyUtrwalone migotanie przedsionkówStany Zjednoczone, Australia
-
Chang Gung Memorial HospitalNieznanyNiewydolność serca