THOPAZ PARA AVALIAÇÃO DE PALS
Utilidade do sistema Thopaz+ na avaliação de vazamentos persistentes de ar
Persistent air leaks (PALs) are a common postoperative complication resulting from tears in the visceral pleura or peripheral lung resections. Although highly prevalent with significant consequences, traditional assessment methods lack the capability to objectively quantify air leaks, which further complicates management, results in inconsistent decision making, and prolongs hospital stays. By incorporating the Thopaz+ system, the investigators can introduce a reliable approach to objectively quantify air leaks, potentially improving clinical outcomes.
The purpose of this protocol is to perform a pilot prospective controlled clinical trial to evaluate the utility of incorporating the Thopaz+ system and its volume quantification ability in evaluating patients with persistent air leaks (PALs) to guide treatment decisions, specifically endobronchial valves (EBVs).
In the targeted cohort, the Thopaz+ will be connected to the chest tube and the data collected will be combined with that of serial balloon occlusion testing with a Fogarty balloon to accurately localize the air leak source, and subsequently consider the correct placement of EBVs. The investigators will thereby determine whether this experimental strategy will yield a more accurate and objective assessment of air leaks, facilitating timely interventions and improved patient outcomes.
Visão geral do estudo
Status
Condições
Intervenção / Tratamento
Descrição detalhada
The investigators will conduct a prospective controlled clinical trial. The investigators' target demographic includes all adult patients with lung resection and/or non-lung resection-related PAL with ipsilateral chest tubes who were not candidates for surgical intervention or patients who refused the intervention. These patients would undergo a multidisciplinary review by thoracic surgery, interventional pulmonary and the primary care team to consider a bronchoscopic approach with endobronchial valves and possibly bronchial blood patch, which is currently considered standard of care at our institution. During this standard-of-care intervention, quantitative assessment of distal occluded airway pressures/flows in various airways, as well as quantitative assessment of air leak volume will be assessed.
In current medical practice, when EBVs are considered in the context of managing PALs, bronchoscopists may opt to either conduct complete lobar occlusion (sealing of all segments within the target lobe) or segmental occlusion (sealing only selected segments within the target lobe). The decision is usually guided by localizing the source of air leak which is usually done by monitoring the reduction of air leak on the pleural drainage device following serial inflation of a Fogarty Catheter at various locations within the endobronchial tree and monitoring change in air leak visually in the pleural drainage system.
In this trial, patients with PALs will have their chest tubes connected to the Thopaz+ system. The Thopaz+ will continuously quantify the volume of air leaks, providing real-time data for analysis.
Data from both the Thopaz+ and standard serial occlusion test with Fogarty balloon will be combined and integrated to provide a comprehensive assessment of the air leak source. The bronchoscopist will perform standard visual assessment of changes in air leak during serial balloon occlusion in order to identify the best region for endobronchial treatments; use of measurements from the Thopaz and serial balloon occlusion test will be allowed as part of clinical reasoning but are not required to be used by the treating clinician.
Continuous monitoring using the Thopaz+ system may continue post-procedure to track changes in air leak volumes and patient recovery, ensuring timely interventions and optimal patient outcomes. This integrated approach aims to provide a more accurate, objective, and efficient method of managing PALs compared to traditional techniques.
Tipo de estudo
Inscrição (Estimado)
Contactos e Locais
Contato de estudo
- Nome: Kai Swenson, MD
- Número de telefone: 617-632-8386
- E-mail: BIDMC-IPResearchTeam@bidmc.harvard.edu
Locais de estudo
-
-
Massachusetts
-
Boston, Massachusetts, Estados Unidos, 02215
- Recrutamento
- Beth Israel Deaconess Medical Center
-
Contato:
- Kai Swenson, MD
- Número de telefone: 617-632-8386
- E-mail: BIDMC-IPResearchTeam@bidmc.harvard.edu
-
-
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
Aceita Voluntários Saudáveis
Gêneros Elegíveis para o Estudo
Método de amostragem
População do estudo
All study subjects will be patients who meet the inclusion criteria including a willingness to read, understand, and sign the informed consent. Recruitment of study subjects will be from the pool of admitted patients from our inpatient service. They will be recruited in a consecutive fashion from the clinical practices of the investigators from BIDMC.
Descrição
Inclusion Criteria:
- Age ≥ 18 years
- Hospitalized with a persistent air leak (PAL) lasting ≥5 days despite continuous chest tube drainage.
- Undergoing bronchoscopy as part of clinical care for PAL management.
- Able and willing to provide written informed consent
Exclusion Criteria:
- Contraindications to bronchoscopy or EBV placement.
- Pregnancy.
- Inability to tolerate general anesthesia.
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Modelos de observação: COHORT
- Perspectivas de Tempo: PROSPECTIVE
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Number of Participants with Successful Intraoperative Quantification of Air Leak Volume Using Thopaz+ Digital Drainage System
Prazo: During procedure
|
Feasibility will be defined by the ability to obtain continuous, interpretable quantitative air leak volume data (in mL/min) during bronchoscopic evaluation of persistent air leaks. Successful quantification is defined as capturing measurable air leak volume data during serial balloon occlusion in at least one airway segment. The outcome will be reported as the number and percentage of participants with successful measurements out of the total enrolled population. |
During procedure
|
Medidas de resultados secundários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Diminuição máxima no volume de vazamento de ar durante a oclusão de balão serial, conforme medido pelo dispositivo THOPAZ+
Prazo: Durante o procedimento
|
Durante o procedimento
|
|
|
Falha no tratamento: nenhuma diminuição no grau de vazamento de ar (classificação de cerfolio) e a necessidade de sucção contínua para o ralo torácico
Prazo: De inscrição até 8 semanas de acompanhamento (conforme o padrão de atendimento ao tratamento broncoscópico de amigos)
|
De inscrição até 8 semanas de acompanhamento (conforme o padrão de atendimento ao tratamento broncoscópico de amigos)
|
|
|
Duração da estadia após a colocação do EBV
Prazo: De inscrição até 8 semanas de acompanhamento (conforme o padrão de atendimento ao tratamento broncoscópico de amigos)
|
De inscrição até 8 semanas de acompanhamento (conforme o padrão de atendimento ao tratamento broncoscópico de amigos)
|
|
|
Duração do tubo torácico após broncoscopia
Prazo: De inscrição até 8 semanas de acompanhamento (conforme o padrão de atendimento ao tratamento broncoscópico de amigos)
|
De inscrição até 8 semanas de acompanhamento (conforme o padrão de atendimento ao tratamento broncoscópico de amigos)
|
|
|
A necessidade de intervenção não cirúrgica adicional
Prazo: De inscrição até 8 semanas de acompanhamento (conforme o padrão de atendimento ao tratamento broncoscópico de amigos)
|
De inscrição até 8 semanas de acompanhamento (conforme o padrão de atendimento ao tratamento broncoscópico de amigos)
|
|
|
A necessidade de intervenção cirúrgica adicional para PAL, como uma cirurgia toracoscópica por vídeo com pleurodese mecânica, blebectomia ou lobectomia
Prazo: De inscrição até 8 semanas de acompanhamento (conforme o padrão de atendimento ao tratamento broncoscópico de amigos)
|
De inscrição até 8 semanas de acompanhamento (conforme o padrão de atendimento ao tratamento broncoscópico de amigos)
|
|
|
Percentage of patients with identifiable lobar or segmental airways responsible for PAL
Prazo: During procedure
|
Percentage of patients with identifiable lobar or segmental airways responsible for PAL |
During procedure
|
|
Treatment success: decrease of one grade in Cerfolio classification for air leaks or removal of continuous suction from chest drain.
Prazo: From enrollment until 8 weeks follow-up (as per standard of care for bronchoscopic treatment of PALs)
|
The Cerfolio classification categorizes air leaks based on their occurrence during the respiratory cycle, with grades 1-4 indicating increasing severity. Grade 1 is a forced expiratory leak, seen during coughing; Grade 2, expiratory; Grade 3, inspiratory; and Grade 4, continuous during both inspiration and expiration |
From enrollment until 8 weeks follow-up (as per standard of care for bronchoscopic treatment of PALs)
|
|
Change in Air Leak Severity Graded by Cerfolio Classification
Prazo: From enrollment until 8 weeks follow-up (as per standard of care for bronchoscopic treatment of PALs)
|
The Cerfolio classification categorizes air leaks based on their occurrence during the respiratory cycle, with grades 1-4 indicating increasing severity. Grade 1 is a forced expiratory leak, seen during coughing; Grade 2, expiratory; Grade 3, inspiratory; and Grade 4, continuous during both inspiration and expiration |
From enrollment until 8 weeks follow-up (as per standard of care for bronchoscopic treatment of PALs)
|
|
Change in Quantitative Air Leak Volume Measured by Thopaz+ Digital Drainage System
Prazo: From enrollment until 8 weeks follow-up (as per standard of care for bronchoscopic treatment of PALs)
|
The Thopaz+ device will be used to continuously measure the volume of air leak (in mL/min) before and after bronchoscopic intervention. The outcome will report the change in air leak volume from baseline (pre-bronchoscopy) to post-bronchoscopy values. |
From enrollment until 8 weeks follow-up (as per standard of care for bronchoscopic treatment of PALs)
|
Colaboradores e Investigadores
Patrocinador
Colaboradores
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Real)
Conclusão Primária (Estimado)
Conclusão do estudo (Estimado)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Palavras-chave
Outros números de identificação do estudo
- 2024P001113
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
produto fabricado e exportado dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .