重症监护室的连续血糖监测
重症监护病房的连续血糖监测:验证和实施
The aim of the proposed clinical trial is to study the use of the Dexcom G7 continuous glucose monitor (CGM) within the intensive care unit (ICU) environment. The overall goal is to validate the accuracy and practicality of the Dexcom G7 when used to assess glycemia in critically ill patients.
研究概览
详细说明
The specific goals are 1) to work with intensive care unit (ICU) staff, in a respiratory ICU setting, to determine how best to place and maintain Dexcom G7 continuous glucose monitors (CGMs) on patients and how to place recording devices and chargers in that environment, 2) download CGM readings to the Dexcom Dashboard in an analyzable format, and 3) to compare simultaneous CGM readings with glucose measurements obtained in the course of routine care as recorded in the electronic medical record. The primary analytical goal is the correlation of CGM glucose measurements with those obtained in the course of routine care. A secondary goal is to determine the frequency of device malfunctions that occur either as a result of device failure (e.g. it falls off) or adverse events (e.g. removal due to skin irritation).
Study subjects will be patients admitted to the 6-ICU UMass Memorial Medical Center (UMMMC) University Campus medical intensive care unit for any reason who have known diabetes or who are hyperglycemic (blood glucose concentration ≥200 mg/dL) with diabetes that was previously undiagnosed or unrecognized. All enrolled volunteers will have a Dexcom G7 CGM place on the skin according to manufacturer's specifications. The devices will remain in place for 10 days or until discharge from the ICU
研究类型
注册 (实际的)
阶段
- 不适用
联系人和位置
学习地点
-
-
Massachusetts
-
Worcester、Massachusetts、美国、01605
- UMass Memorial Medical Center
-
-
参与标准
资格标准
适合学习的年龄
接受健康志愿者
有资格学习的性别
描述
Inclusion Criteria:
Campus medical intensive care unit for any reason who have known diabetes or who are hyperglycemic (blood glucose concentration ≥200 mg/dL) with diabetes that was previously undiagnosed or unrecognized.
- Persons with either type 1 or type 2 diabetes.
- Persons of either sex.
- Persons of any age over 18.
- Persons able to give informed consent or cognitively impaired adults whose legally authorized representative (LAR) can give informed consent.
- English or Spanish speaking adults.
Exclusion Criteria:
- Persons unable to provide informed consent.
- Cognitively impaired persons whose LAR declines to provide informed consent.
- Patients receiving hydroxyurea, which interferes with sensor function.
- Patients receiving high dose acetaminophen (> 4 gm/day), which interferes with sensor function.
- Anasarca affecting preferred areas for device placement (arm, upper buttocks, or thigh), which may interfere with sensor accuracy.
- Severe cachexia with absence of subcutaneous fat at preferred sensor placement sites, which might interfere with sensor accuracy.
- Persons who are pregnant.
- Prisoners.
学习计划
研究是如何设计的?
设计细节
- 主要用途:其他
- 分配:不适用
- 介入模型:单组作业
- 屏蔽:无(打开标签)
武器和干预
参与者组/臂 |
干预/治疗 |
|---|---|
|
实验性的:Single Arm
All members of the arm will have a CGM place and glucose concentration downloaded. |
Continuous glucose monitor |
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Mean Absolute Relative Difference (MARD).
大体时间:6 months
|
Concordance of glucose concentration results obtained by standard test methods with those obtained nearly simultaneously by a continuous glucose monitoring device was assessed by Mean Absolute Relative Difference (MARD). The MARD is the average percentage difference between the continuous glucose monitor readings measured over 6 months and glucose concentration results obtained by near simultaneous (+/- 5 minutes) standard test methods was used to gauge the accuracy of the CGM in the ICU setting. Typical MARD values reported for CGMs in the outpatient setting average approximately 8%. Typical MARD values reported for CGMs other than the one used in this study n the ICU setting average approximately 12%. |
6 months
|
|
Clarke Error Grid
大体时间:6 months
|
Concordance of glucose concentration results obtained by standard test methods with those obtained nearly simultaneously by a continuous glucose monitoring device was assessed by Clarke Error Grid. The Clarke Error Grid plots of reference glucose values obtained by point-of-care fingerstick or laboratory determination (x-axis) against CGM values (y-axis). The grid is divided into 5 zones (A-E) that indicate clinical implications of CGM glucose measurements when compared to a "gold standard" Region A: Clinically accurate. Region B: Benign measurement error. Region C: Overcorrection error region CGM values suggest unnecessary treatment. Region D: Serious hypo- or hyperglycemia detection error region. Region E: Erroneous treatment region. CGM fails to detect potentially dangerous hypo- or hyperglycemia Ideally, all values are in Regions A and B. We report the percentage of correlated values that fall in each region. |
6 months
|
次要结果测量
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Unplanned Event Frequency
大体时间:6 months
|
Number of occurrences per patient of device failure or or unforeseen device removal |
6 months
|
合作者和调查者
调查人员
- 首席研究员:John P Mordes, MD、UMass Chan Medical School
研究记录日期
研究主要日期
学习开始 (实际的)
初级完成 (实际的)
研究完成 (实际的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新
最后验证
更多信息
与本研究相关的术语
其他研究编号
- STUDY00001138
计划个人参与者数据 (IPD)
计划共享个人参与者数据 (IPD)?
IPD 计划说明
Study design and deidentified aggregate glucose concentration data will be shared. No individual patient data will be shared.
IPD 共享时间框架
One year
IPD 共享访问标准
Publication
IPD 共享支持信息类型
- 研究方案
- 树液
- 国际碳纤维联合会
药物和器械信息、研究文件
研究美国 FDA 监管的药品
研究美国 FDA 监管的设备产品
在美国制造并从美国出口的产品
此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.